The influence of operational protocol on the fluid dynamics in the 12 cc Penn state pulsatile pediatric ventricular assist device: the effect of end-diastolic delay.

The success of adult ventricular assist devices (VADs), coupled with the high transplant waiting list mortality of infants (40%) has prompted Penn State to develop a pediatric version of the clinically successful adult device. Although the primary use of this device will be bridge-to-transplant, there has been sufficient clinical data to demonstrate the efficacy of VADs in a bridge-to-recovery setting. However, removing the patient from the device, a process known as weaning, demands operation of the device at a lower beat rate and concomitant increased risk for thromboembolism. Previous studies have shown that the interrelated flow characteristics necessary for the prevention of thrombosis in a pulsatile VAD are a strong inlet jet, a late diastolic recirculating flow, and a wall shear rate greater than 500/s. In an effort to develop a strong inlet jet and rotational flow pattern at a lower beat and flow rate, we have compressed diastole by altering the end-diastolic delay time (EDD). Particle image velocimetry was used to compare the flow fields and wall shear rates in the chamber of the 12 cc Penn State pulsatile pediatric VAD operated at 50 beats per minute using EDDs of 10, 50, and 100 ms. Although we expected the 100 ms EDD to have the best wall shear profiles, we found that the 50 ms EDD condition was superior to both the 10 and 100 EDD conditions, due to a longer sustained inlet jet.

[1]  Steven Deutsch,et al.  Correlation of In Vivo Clot Deposition With the Flow Characteristics in the 50 cc Penn State Artificial Heart: A Preliminary Study , 2004, ASAIO journal.

[2]  Steven Deutsch,et al.  The 12 cc Penn State pulsatile pediatric ventricular assist device: fluid dynamics associated with valve selection. , 2008, Journal of biomechanical engineering.

[3]  A. Feldman,et al.  Downregulation of Matrix Metalloproteinases and Reduction in Collagen Damage in the Failing Human Heart After Support With Left Ventricular Assist Devices , 2001, Circulation.

[4]  M. Slaughter,et al.  A new method of monitoring recovery and weaning the Thoratec left ventricular assist device. , 2001, The Annals of thoracic surgery.

[5]  M. Oz,et al.  Improvement of myocardial mitochondrial function after hemodynamic support with left ventricular assist devices in patients with heart failure. , 1998, The Journal of thoracic and cardiovascular surgery.

[6]  W. Williams,et al.  Pediatric heart transplantation: improving results in high-risk patients. , 2001, The Journal of thoracic and cardiovascular surgery.

[7]  O. Frazier,et al.  Mechanical unloading increases caveolin expression in the failing human heart. , 2003, Cardiovascular research.

[8]  Steven Deutsch,et al.  Viscoelasticity of Pediatric Blood and its Implications for the Testing of a Pulsatile Pediatric Blood Pump , 2005, ASAIO journal.

[9]  C. Mavroudis,et al.  Pediatric Cardiac Surgery , 1994 .

[10]  R. Körfer,et al.  The time course of natriuretic hormones as plasma markers of myocardial recovery in heart transplant candidates during ventricular assist device support reveals differences among device types. , 2001, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[11]  R. Hetzer,et al.  [Temporary mechanical left heart support. Recovery of heart function in patients with end-stage idiopathic dilated cardiomyopathy]. , 1997, Herz.

[12]  J. Müller,et al.  Temporäre mechanische Linksherzentlastung , 1997, Herz.

[13]  T. Myers,et al.  Exercise capacity recovers slowly but fully in patients with a left ventricular assist device. , 1996, ASAIO journal.

[14]  Gábor Janiga,et al.  Turbulent flame visualization using direct numerical simulation , 2007, J. Vis..

[15]  J. Tarbell,et al.  Evaluation of a transparent blood analog fluid: aqueous xanthan gum/glycerin. , 1993, Biorheology.

[16]  C. Zapanta,et al.  Effect of the Diastolic and Systolic Duration on Valve Cavitation in a Pediatric Pulsatile Ventricular Assist Device , 2005, ASAIO journal.

[17]  Steven Deutsch,et al.  Wall shear-rate estimation within the 50cc Penn State artificial heart using particle image velocimetry. , 2004, Journal of biomechanical engineering.

[18]  S. Deutsch,et al.  Flow behavior within the 12-cc Penn State pulsatile pediatric ventricular assist device: an experimental study of the initial design. , 2008, Artificial organs.

[19]  J. Hubbell,et al.  Visualization and analysis of mural thrombogenesis on collagen, polyurethane and nylon. , 1986, Biomaterials.

[20]  G. Noon,et al.  Regression of fibrosis and hypertrophy in failing myocardium following mechanical circulatory support. , 2001, The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation.

[21]  Steven Deutsch,et al.  EXPERIMENTAL FLUID MECHANICS OF PULSATILE ARTIFICIAL BLOOD PUMPS , 2006 .

[22]  S. Deutsch,et al.  The 12 cc Penn State Pulsatile Pediatric Ventricular Assist Device: Flow Field Observations at a Reduced Beat Rate With Application to Weaning , 2008, ASAIO journal.

[23]  O. Frazier,et al.  Mechanical bridging to improvement in severe acute "nonischemic, nonmyocarditis" heart failure. , 2004, Congestive heart failure.

[24]  P. Ferrazzi,et al.  Assisted circulation for myocardial recovery after repair of congenital heart disease. , 1991, European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery.

[25]  P. D. del Nido,et al.  Mechanical circulatory support in children with cardiac disease. , 1999, The Journal of thoracic and cardiovascular surgery.

[26]  O. Frazier,et al.  Improved left ventricular function after chronic left ventricular unloading. , 1996, The Annals of thoracic surgery.