Specific absorption rate benefits of including measured electric field interactions in parallel excitation pulse design

Specific absorption rate management and excitation fidelity are key aspects of radiofrequency pulse design for parallel transmission at ultra–high magnetic field strength. The design of radiofrequency pulses for multiple channels is often based on the solution of regularized least‐squares optimization problems for which a regularization term is typically selected to control the integrated or peak pulse waveform amplitude. Unlike single‐channel transmission, the specific absorption rate of parallel transmission is significantly influenced by interferences between the electric fields associated with the individual transmission elements, which a conventional regularization term does not take into account. This work explores the effects upon specific absorption rate of incorporating experimentally measurable electric field interactions into parallel transmission pulse design. Results of numerical simulations and phantom experiments show that the global specific absorption rate during parallel transmission decreases when electric field interactions are incorporated into pulse design optimization. The results also show that knowledge of electric field interactions enables robust prediction of the net power delivered to the sample or subject by parallel radiofrequency pulses before they are played out on a scanner. Magn Reson Med, 2011. © 2011 Wiley‐Liss, Inc.

[1]  Douglas C Noll,et al.  Additive angle method for fast large‐tip‐angle RF pulse design in parallel excitation , 2008, Magnetic resonance in medicine.

[2]  Vivek K Goyal,et al.  Specific absorption rate studies of the parallel transmission of inner‐volume excitations at 7T , 2008, Journal of magnetic resonance imaging : JMRI.

[3]  D. Sodickson,et al.  Electrodynamic constraints on homogeneity and radiofrequency power deposition in multiple coil excitations , 2009, Magnetic resonance in medicine.

[4]  P. Börnert,et al.  Transmit SENSE , 2003, Magnetic resonance in medicine.

[5]  Douglas C Noll,et al.  Iterative RF pulse design for multidimensional, small‐tip‐angle selective excitation , 2005, Magnetic resonance in medicine.

[6]  T. Ibrahim,et al.  Effect of RF coil excitation on field inhomogeneity at ultra high fields: a field optimized TEM resonator. , 2001, Magnetic resonance imaging.

[7]  Jens Frahm,et al.  Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imaging , 2002, Magnetic resonance in medicine.

[8]  Ke Feng,et al.  Reducing SAR in parallel excitation using variable-density spirals: a simulation-based study. , 2008, Magnetic resonance imaging.

[9]  Zhi-Pei Liang,et al.  A noniterative method to design large‐tip‐angle multidimensional spatially‐selective radio frequency pulses for parallel transmission , 2007, Magnetic resonance in medicine.

[10]  Albert Macovski,et al.  A linear class of large-tip-angle selective excitation pulses , 1989 .

[11]  Yudong Zhu,et al.  Parallel excitation with an array of transmit coils , 2004, Magnetic resonance in medicine.

[12]  Y. Zhu In Vivo RF Power and SAR Calibration for Multi-Port RF Transmission , 2008 .

[13]  Douglas C Noll,et al.  Fast‐kz three‐dimensional tailored radiofrequency pulse for reduced B1 inhomogeneity , 2006, Magnetic resonance in medicine.

[14]  E. Adalsteinsson,et al.  Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channels , 2008, Magnetic resonance in medicine.

[15]  Y. Zhu,et al.  Deposition and RF Power Requirements in Parallel Transmission with Increasing Distance from the Coil to the Sample , 2008 .

[16]  D. Hoult The principle of reciprocity in signal strength calculations—a mathematical guide , 2000 .

[17]  Douglas C Noll,et al.  Spatial domain method for the design of RF pulses in multicoil parallel excitation , 2006, Magnetic resonance in medicine.

[18]  Johnpauly A k-Space Analysis of Small-Tip-Angle Excitation , 2012 .

[19]  Elfar Adalsteinsson,et al.  Simple analytic variable density spiral design , 2003, Magnetic resonance in medicine.

[20]  M. W. Vogel,et al.  Sparse Parallel Transmit Pulse Design Using Orthogonal Matching Pursuit Method , 2008 .

[21]  Klaas P Pruessmann,et al.  Optimal design of multiple‐channel RF pulses under strict power and SAR constraints , 2010, Magnetic resonance in medicine.

[22]  D. Hoult Sensitivity and Power Deposition in a High‐Field Imaging Experiment , 2000, Journal of magnetic resonance imaging : JMRI.

[23]  G. Mckinnon,et al.  Designing multichannel, multidimensional, arbitrary flip angle RF pulses using an optimal control approach , 2008, Magnetic resonance in medicine.

[24]  P. Börnert,et al.  Basic considerations on the impact of the coil array on the performance of Transmit SENSE , 2005, Magnetic Resonance Materials in Physics, Biology and Medicine.

[25]  J. Pauly,et al.  Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging]. , 1991, IEEE transactions on medical imaging.

[26]  V A Stenger,et al.  Three‐dimensional tailored RF pulses for the reduction of susceptibility artifacts in T*2‐weighted functional MRI , 2000, Magnetic resonance in medicine.