Bone vibration measurement using ultrasound: application to detection of hip prosthesis loosening.

Hip prosthesis loosening can be determined in vivo using a vibration-based technique called vibrometry. In this technique, a low frequency (<1000Hz) sinusoidal vibration is applied to the femoral condyles and the resulting vibration is measured at the greater trochanter. If the prosthesis is securely fixed, the output vibration signal matches that of the input vibration, whereas if the prosthesis is loose, the output vibration signal is distorted and shows the marked presence of harmonics of the input frequency. One of the main problems with this application of this technique is in measuring the output vibration where significant amounts of soft tissue cover the measurement site. In order to circumvent this problem, an ultrasound probe, normally used for the measurement of blood flow, has been used to measure the output vibration. This has been evaluated by comparing the results obtained from the ultrasound probe with those from a conventional accelerometer in models representing a tight and loose hip prosthesis under simulated clinical conditions. The ultrasound probe was able to consistently detect the output vibration, for both the loose and secure prostheses. Under the test conditions used (which attempted to simulate a large thickness of soft tissue), the ultrasound probe was able to produce a greatly enhanced output vibration signal compared to the accelerometer. This suggests that the use of an ultrasound probe to detect mechanically induced vibration through significant amounts of soft tissue appears to be viable and could lead to enhanced detection of prosthesis loosening using this technique.

[1]  A. Georgiou,et al.  Accurate diagnosis of hip prosthesis loosening using a vibrational technique. , 2001, Clinical biomechanics.

[2]  N Meredith,et al.  The design of a vibration transducer to monitor the integrity of dental implants , 1998, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[3]  P. Ogrodnik,et al.  Cortex Distraction Forces Caused by the Insertion of External Fixator Pins , 2001, Journal of orthopaedic trauma.

[4]  G F McCoy,et al.  The Differentiation of Loose and Secure Femoral Implants in Total Hip Replacement Using a Vibrational Technique: An Anatomical and Pilot Clinical Study , 1989, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[5]  E. Salvati,et al.  Evaluation of painful hip arthroplasties. Are technetium bone scans necessary? , 1993, The Journal of bone and joint surgery. British volume.

[6]  J. Korst,et al.  Effects of total hip replacement on quality of life in patients with osteoarthritis and in patients with rheumatoid arthritis , 1994, Clinical Rheumatology.

[7]  G. van der Perre,et al.  Vibration, sonic and ultrasonic wave propagation analysis for the detection of osteoporosis. , 1994, Clinical rheumatology.

[8]  Gang Qi,et al.  How much can a vibrational diagnostic tool reveal in total hip arthroplasty loosening? , 2003, Clinical biomechanics.

[9]  Georges Van der Perre,et al.  Analysis of the fixation quality of cementless hip prostheses using a vibrational technique , 2004 .

[10]  J L Cunningham,et al.  Design and testing of external fixator bone screws. , 1990, Journal of biomedical engineering.

[11]  S. G. Roberts,et al.  Efficacy of monitoring long‐bone fracture healing by measurement of either bone stiffness or resonant frequency: Numerical simulation , 2000, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[12]  R. Puers,et al.  Measurement of implant stability and osseointegration: non-linearities and resonance frequencies , 2000 .

[13]  N B Jones,et al.  Loosening of total hip arthroplasty. Diagnosis by vibration analysis. , 1995, The Journal of bone and joint surgery. British volume.

[14]  J. Orr,et al.  An Investigation into the Effect of Cyclic Loading and Frequency on the Temperature of Pmma Bone Cement in Hip Prostheses , 1989, Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine.

[15]  T. Brown,et al.  Reliability of detecting prosthesis/cement interface radiolucencies in total hip arthroplasty , 2002, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[16]  Y. Nakatsuchi,et al.  The vibrational mode of the tibia and assessment of bone union in experimental fracture healing using the impulse response method. , 1996, Medical engineering & physics.

[17]  Robert Puers,et al.  A telemetry system for the detection of hip prosthesis loosening by vibration analysis , 2000 .

[18]  W. Harris,et al.  Failed total hip replacement: assessment by plain radiographs, arthrograms, and aspiration of the hip joint. , 1984, The Journal of bone and joint surgery. American volume.