A durable, low-cost electrogoniometer for dynamic measurement of joint trajectories.

This article introduces a method and step-by-step instructions for the design of a low-cost, flexible electrogoniometer, suitable for kinesiology, rehabilitation, and biometric applications. Two unidirectional flexible sensors are placed back-to-back, and a multivariate linear regression model was used to combine measurements from the two sensors. Following a short calibration procedure, the electrogoniometer can be reliably used for measurement of flexion/extension angles of various hinge joints. The performance of the goniometer has been tested on a population of 21 healthy subjects performing flexion/extension of index finger, wrist and elbow. The proposed device achieves the quality of joint angle measurements comparable to that of commercial electrogoniometers, while having a significantly higher durability-to-cost ratio.

[1]  R Maksimovic,et al.  Classification of tetraplegics through automatic movement evaluation. , 1999, Medical engineering & physics.

[2]  Bruce J Sangeorzan,et al.  Assessment of an electronic goniometer designed to measure equinus contracture. , 2003, Journal of rehabilitation research and development.

[3]  A Keys,et al.  Indices of relative weight and obesity. , 1972, Journal of chronic diseases.

[4]  B Zappa,et al.  A model of an electro-goniometer and its calibration for biomechanical applications. , 2000, Medical engineering & physics.

[5]  David Zeltzer,et al.  A survey of glove-based input , 1994, IEEE Computer Graphics and Applications.

[6]  F Franchignoni,et al.  Flexible electrogoniometers: kinesiological advantages with respect to potentiometric goniometers. , 1995, Clinical biomechanics.

[7]  Ac Nicol,et al.  Measurement of joint motion , 1989 .

[8]  P. Holland,et al.  Robust regression using iteratively reweighted least-squares , 1977 .

[9]  Gert-Åke Hansson,et al.  Improving goniometer accuracy by compensating for individual transducer characteristics. , 2009, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[10]  A. John Mallinckrodt,et al.  Data Reduction and Error Analysis for the Physical Sciences , 1993 .