Design optimization of a contact-aided continuum robot for endobronchial interventions based on anatomical constraints

PurposeA laser-profiled continuum robot (CR) with a series of interlocking joints has been developed in our center to reach deeper areas of the airways. However, it deflects with constant curvature, which thus increases the difficulty of entering specific bronchi without relying on the tissue reaction forces. This paper aims to propose an optimization framework to find the best design parameters for nonconstant curvature CRs to reach distal targets while attempting to avoid the collision with the surrounding tissue.MethodsFirst, the contact-aided compliant mechanisms (CCMs) are integrated with the continuum robot to achieve the nonconstant curvature. Second, forward kinematics considering CCMs is built. Third, inverse kinematics is implemented to steer the robot tip toward the desired targets within the confined anatomy. Finally, an optimization framework is proposed to find the best robot design to reach the target with the least collision to the bronchi walls.ResultsExperiments are carried out to verify the feasibility of CCMs to enable the nonconstant curvature deflection, and simulations demonstrate a lower cost function value to reach a target for the nonconstant curvature optimized design with respect to the standard constant curvature robot (0.11 vs. 2.66). In addition, the higher capacity of the optimized design to complete the task is validated by interventional experiments using fluoroscopy.ConclusionResults demonstrate the effectiveness of the proposed framework to find an optimized CR with nonconstant curvature to perform safer interventions to reach distal targets.

[1]  G. K. Ananthasuresh,et al.  Contact Aided Compliant Mechanisms: Concept and Preliminaries , 2002 .

[2]  Guang-Zhong Yang,et al.  Design and analysis of a wire-driven flexible manipulator for bronchoscopic interventions , 2016, 2016 IEEE International Conference on Robotics and Automation (ICRA).

[3]  Yun Zou,et al.  A Contact-Aided Asymmetric Steerable Catheter for Atrial Fibrillation Ablation , 2017, IEEE Robotics and Automation Letters.

[4]  D. Feller-Kopman,et al.  Interventional bronchoscopy in adults , 2018, Expert review of respiratory medicine.

[5]  Pierre E. Dupont,et al.  Concentric Tube Robot Design and Optimization Based on Task and Anatomical Constraints , 2015, IEEE Transactions on Robotics.

[6]  Robert J. Webster,et al.  Task-oriented design of concentric tube robots using mechanics-based models , 2012, 2012 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[7]  Tomer Anor,et al.  Algorithms for design of continuum robots using the concentric tubes approach: A neurosurgical example , 2011, 2011 IEEE International Conference on Robotics and Automation.

[8]  Robert J. Webster,et al.  Motion planning for a three-stage multilumen transoral lung access system , 2015, 2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS).

[9]  Robert J. Webster,et al.  A Telerobotic System for Transnasal Surgery , 2014, IEEE/ASME Transactions on Mechatronics.

[10]  Pierre E. Dupont,et al.  Design optimization of concentric tube robots based on task and anatomical constraints , 2011, 2011 IEEE International Conference on Robotics and Automation.

[11]  Sungchul Kang,et al.  Design parameter optimization of a novel serial manipulator for microsurgery , 2017, 2017 14th International Conference on Ubiquitous Robots and Ambient Intelligence (URAI).