Development of a wireless intra-vaginal transducer for monitoring intra-abdominal pressure in women

Pelvic floor disorders (PFD) affect one of every four women in the United States. Elevated intra-abdominal pressure (IAP) during daily activity or strenuous physical activity has been identified as a risk factor in the prevalence of PFD. However, the relationship between IAP and physical activity is poorly understood and oftentimes activity restrictions are prescribed by physicians without clinical evidence linking various activities to elevated IAP. There are currently no pressure transducers capable of monitoring IAP non-invasively out of a clinical environment. To overcome this shortcoming, a novel intra-vaginal pressure transducer (IVT) was developed to continuously monitor IAP. Improvements were made to the first generation IVT by incorporating wireless capability to enhance the device’s mobility while creating a more robust IAP monitoring system. To ensure the changes maintained the functionality of the original device design, comparison testing with standard clinical pressure transducers in both bench top and clinical settings was conducted. The wireless device was found to have high linearity, robust signal transmission, and dynamic response that outperforms the clinical standard rectal transducer and is similar to the original first generation non-wireless design. The wireless IVT presented here is a mobile wireless device capable of measuring, storing and transmitting IAP data during various physical activities.

[1]  A. McTiernan,et al.  Pelvic organ prolapse in the Women's Health Initiative: gravity and gravidity. , 2002, American journal of obstetrics and gynecology.

[2]  R. Hitchcock,et al.  Development and testing of a vaginal pressure sensor to measure intra‐abdominal pressure in women , 2009, Neurourology and urodynamics.

[3]  Morton B. Brown,et al.  A multicentered comparison of measurements obtained with microtip and external water pressure transducers , 2006, International Urogynecology Journal.

[4]  K. Wise,et al.  A wireless microsystem for the remote sensing of pressure, temperature, and relative humidity , 2005, Journal of Microelectromechanical Systems.

[5]  Timothy McClure,et al.  Development of a fully implantable wireless pressure monitoring system , 2009, Biomedical microdevices.

[6]  R. Minns Intracranial pressure monitoring. , 1984, Archives of disease in childhood.

[7]  Jennifer M Wu,et al.  Prevalence of symptomatic pelvic floor disorders in US women. , 2008, JAMA.

[8]  Mark R Stoker,et al.  Principles of pressure transducers, resonance, damping and frequency response , 2004 .

[9]  Manu L. N. G. Malbrain,et al.  Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal , 2004, Intensive Care Medicine.

[10]  John C. Fang,et al.  Comparison of Air-Coupled Balloon Esophageal and Anorectal Manometry Catheters with Solid-State Esophageal Manometry and Water-Perfused Anorectal Manometry Catheters , 2004, Digestive Diseases and Sciences.

[11]  D. Griffiths,et al.  Standardisation of urethral pressure measurement: Report from the standardisation sub‐committee of the International Continence Society , 2002, Neurourology and urodynamics.

[12]  C. Goodman Association for the Advancement of Medical Instrumentation , 1988 .

[13]  P. Hunziker Accuracy and dynamic response of disposable pressure transducertubing systems , 1987, Canadian journal of anaesthesia = Journal canadien d'anesthesie.

[14]  Monir K. Parikh,et al.  Development of a novel intra-vaginal transducer with improved dynamic response , 2009, Biomedical microdevices.

[15]  C. Gibbs,et al.  Simultaneous urethral pressure profilometry using a microtip transducer catheter in the bitch: Effects of bitch position and transducer orientation , 1990 .