Safety Assessment of Advanced Imaging Sequences I: Measurements

A method for rapid measurement of intensities (Ispta), mechanical index (MI), and probe surface temperature for any ultrasound scanning sequence is presented. It uses the scanner's sampling capability to give an accurate measurement of the whole imaging sequence for all emissions to yield the true distributions. The method is several orders of magnitude faster than approaches using an oscilloscope, and it also facilitates validating the emitted pressure field and the scanner's emission sequence software. It has been implemented using the experimental synthetic aperture real-time ultrasound system (SARUS) scanner and the Onda AIMS III intensity measurement system (Onda Corporation, Sunnyvale, CA, USA). Four different sequences have been measured: a fixed focus emission, a duplex sequence containing B-mode and flow emissions, a vector flow sequence with B-mode and flow emissions in 17 directions, and finally a SA duplex flow sequence. A BK8820e (BK Medical, Herlev, Denmark) convex array probe is used for the first three sequences and a BK8670 linear array probe for the SA sequence. The method is shown to give the same intensity values within 0.24% of the AIMS III Soniq 5.0 (Onda Corporation, Sunnyvale, CA, USA) commercial intensity measurement program. The approach can measure and store data for a full imaging sequence in 3.8-8.2 s per spatial position. Based on Ispta, MI, and probe surface temperature, the method gives the ability to determine whether a sequence is within U.S. FDA limits, or alternatively indicate how to scale it to be within limits.

[1]  Mickael Tanter,et al.  Ultrafast imaging in biomedical ultrasound , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[2]  J. Jensen Estimation of Blood Velocities Using Ultrasound: A Signal Processing Approach , 1996 .

[3]  B. Tomov,et al.  Simulation and efficient measurements of intensities for complex imaging sequences , 2014, 2014 IEEE International Ultrasonics Symposium.

[4]  J. Jensen,et al.  Multielement synthetic transmit aperture imaging using temporal encoding , 2003, IEEE Transactions on Medical Imaging.

[5]  Jørgen Arendt Jensen,et al.  Rapid measurements of intensities for safety assessment of advanced imaging sequences , 2014, Medical Imaging.

[6]  Yongmin Kim,et al.  Research interface on a programmable ultrasound scanner. , 2008, Ultrasonics.

[7]  Jørgen Arendt Jensen,et al.  Convex array vector velocity imaging using transverse oscillation and its optimization , 2015, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[8]  Jørgen Arendt Jensen,et al.  Implementation of a versatile research data acquisition system using a commercially available medical ultrasound scanner , 2012, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[9]  S. I. Nikolov,et al.  SARUS: A synthetic aperture real-time ultrasound system , 2013, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[10]  Steffen Tretbar,et al.  A highly customizable ultrasound research platform for clinical use with a software architecture for 2d-/3d-reconstruction and processing including closed-loop control , 2009 .

[11]  Jørgen Arendt Jensen,et al.  Synthetic aperture ultrasound imaging. , 2006, Ultrasonics.

[12]  Helmut Ermert,et al.  An ultrasound research interface for a clinical system. , 2006, IEEE transactions on ultrasonics, ferroelectrics, and frequency control.

[13]  M. O'Donnell,et al.  Coded excitation system for improving the penetration of real-time phased-array imaging systems , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[14]  J.A. Jensen,et al.  Use of modulated excitation signals in medical ultrasound. Part I: basic concepts and expected benefits , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[15]  Jørgen Arendt Jensen Safety Assessment of Advanced Imaging Sequences II: Simulations , 2016, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[16]  A. Dallai,et al.  ULA-OP: an advanced open platform for ultrasound research , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[17]  O. K. Bates,et al.  Thermal Conductivity of Liquid Silicones , 1949 .

[18]  J. Jensen,et al.  Adaptive multi-lag for synthetic aperture vector flow imaging , 2014, 2014 IEEE International Ultrasonics Symposium.

[19]  W. McDicken,et al.  Doppler Ultrasound: Physics, Instrumentation and Signal Processing , 2000 .

[20]  J.A. Jensen,et al.  Ultrasound research scanner for real-time synthetic aperture data acquisition , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[21]  R. Phillips Guidance for Industry and FDA Staff Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers , 2008 .

[22]  H. Ermert,et al.  An ultrasound research interface for a clinical system , 2006, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.