3D versus 2D steering in patient anatomies: A comparison using hyperthermia treatment planning

Purpose: In this study hyperthermia treatment planning is used to investigate whether the target temperature during hyperthermia treatment can be increased using the 3D AMC-8 instead of the 2D AMC-4 system (AMC: Academic Medical Center). Methods and materials: The heating ability of the AMC-4 and AMC-8 system was analysed for five patients with cervix uteri carcinoma. Dielectric and thermal models were generated, based on a hyperthermia planning computerised tomography (CT), at a resolution of 2.5 × 2.5 × 5.0 mm3. Calculation of the electric fields with the finite-difference time-domain method was followed by SAR- and temperature-based optimisation. The ability to correct for axial shifts of the patient by phase/amplitude steering was investigated for both systems. Finally, it was investigated whether adjusting the ring-to-ring distance of the AMC-8 system can be used for further optimisation. Results: An average increase in T90 of ∼0.5°C (0.2°–0.8°C) was found for the AMC-8 system compared to the AMC-4 system. The gain in T50 and T10 was also 0.5°C on average. The additional power required to achieve this gain was 36% to 71% of the power required for the AMC-4 system. The AMC-8 system has the capability of correcting changes in axial position (−8 cm, +8 cm), contrary to the AMC-4 system. For both systems the axial position should be known within 1–2 cm. Conclusions: Hyperthermia treatment with the AMC-8 system can lead to a clinically relevant increase of the target temperature compared to treatment with the AMC-4 system. The AMC-8 system provides large freedom in the axial positioning of the patient.

[1]  H P Kok,et al.  Optimization in hyperthermia treatment planning: the impact of tissue perfusion uncertainty. , 2010, Medical physics.

[2]  Yvette van Norden,et al.  Hyperthermia dose-effect relationship in 420 patients with cervical cancer treated with combined radiotherapy and hyperthermia. , 2009, European journal of cancer.

[3]  M M Paulides,et al.  Patient positioning in deep hyperthermia: influences of inaccuracies, signal correction possibilities and optimization potential , 2009, Physics in medicine and biology.

[4]  J F Bakker,et al.  A literature survey on indicators for characterisation and optimisation of SAR distributions in deep hyperthermia, a plea for standardisation , 2009, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[5]  J. Crezee,et al.  Improving locoregional hyperthermia delivery using the 3-D controlled AMC-8 phased array hyperthermia system: A preclinical study , 2009, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[6]  R. Issels Hyperthermia adds to chemotherapy. , 2008, European journal of cancer.

[7]  E. D. Geijsen,et al.  Relation between body size and temperatures during locoregional hyperthermia of oesophageal cancer patients , 2008, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[8]  Peter Wust,et al.  Simulation of different applicator positions for treatment of a presacral tumour , 2007, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[9]  Margarethus M. Paulides,et al.  A head and neck hyperthermia applicator: Theoretical antenna array design , 2007, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[10]  J. Lagendijk,et al.  Towards patient specific thermal modelling of the prostate , 2006, Physics in Medicine and Biology.

[11]  H P Kok,et al.  High-resolution temperature-based optimization for hyperthermia treatment planning , 2005, Physics in medicine and biology.

[12]  J. van der Zee,et al.  Characterization of the SAR-distribution of the Sigma-60 applicator for regional hyperthermia using a Schottky diode sheet , 2003, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[13]  J. Lagendijk,et al.  Regional hyperthermia applicator design using FDTD modelling. , 2001, Physics in medicine and biology.

[14]  P Wust,et al.  Electromagnetic phased arrays for regional hyperthermia: optimal frequency and antenna arrangement. , 2001, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[15]  A. Hart,et al.  Comparison of radiotherapy alone with radiotherapy plus hyperthermia in locally advanced pelvic tumours: a prospective, randomised, multicentre trial , 2000, The Lancet.

[16]  S.,et al.  Numerical Solution of Initial Boundary Value Problems Involving Maxwell’s Equations in Isotropic Media , 1966 .

[17]  S K Das,et al.  Computational techniques for fast hyperthermia temperature optimization. , 1999, Medical physics.

[18]  K D Paulsen,et al.  Optimization of pelvic heating rate distributions with electromagnetic phased arrays. , 1999, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[19]  H. H. Pennes Analysis of tissue and arterial blood temperatures in the resting human forearm. 1948. , 1948, Journal of applied physiology.

[20]  P Wust,et al.  ESHO quality assurance guidelines for regional hyperthermia. , 1998, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[21]  C Gabriel,et al.  The dielectric properties of biological tissues: I. Literature survey. , 1996, Physics in medicine and biology.

[22]  F. Bardati,et al.  SAR optimization in a phased array radiofrequency hyperthermia system , 1995, IEEE Transactions on Biomedical Engineering.

[23]  J. Overgaard,et al.  Randomised trial of hyperthermia as adjuvant to radiotherapy for recurrent or metastatic malignant melanoma , 1995, The Lancet.

[24]  Jean-Pierre Berenger,et al.  A perfectly matched layer for the absorption of electromagnetic waves , 1994 .

[25]  Jian L. Zhou,et al.  User's Guide for CFSQP Version 2.0: A C Code for Solving (Large Scale) Constrained Nonlinear (Minimax) Optimization Problems, Generating Iterates Satisfying All Inequality Constraints , 1994 .

[26]  B.J. James,et al.  Creation of three-dimensional patient models for hyperthermia treatment planning , 1992, IEEE Transactions on Biomedical Engineering.

[27]  L. Blank,et al.  Results of deep body hyperthermia with large waveguide radiators. , 1990, Advances in experimental medicine and biology.

[28]  J W Hand,et al.  Quality assurance guidelines for ESHO protocols. , 1989, International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group.

[29]  J J Lagendijk,et al.  Hyperthermia treatment planning , 2000, Physics in medicine and biology.

[30]  W. Dewey,et al.  Thermal dose determination in cancer therapy. , 1984, International journal of radiation oncology, biology, physics.

[31]  H. H. Penns Analysis of tissue and arterial blood temperatures in the resting human forearm , 1948 .