Autoadaptive phase-correlated (AAPC) reconstruction for 4D CBCT.

PURPOSE Kilovoltage cone-beam computed tomography (CBCT) is widely used in image-guided radiation therapy for exact patient positioning prior to the treatment. However, producing time series of volumetric images (4D CBCT) of moving anatomical structures remains challenging. The presented work introduces a novel method, combining high temporal resolution inside anatomical regions with strong motion and image quality improvement in regions with little motion. METHODS In the proposed method, the projections are divided into regions that are subject to motion and regions at rest. The latter ones will be shared among phase bins, leading thus to an overall reduction in artifacts and noise. An algorithm based on the concept of optical flow was developed to analyze motion-induced changes between projections. The technique was optimized to distinguish patient motion and motion deriving from gantry rotation. The effectiveness of the method is shown in numerical simulations and patient data. RESULTS The images reconstructed from the presented method yield an almost the same temporal resolution in the moving volume segments as a conventional phase-correlated reconstruction, while reducing the noise in the motionless regions down to the level of a standard reconstruction without phase correlation. The proposed simple motion segmentation scheme is yet limited to rotation speeds of less than3°∕s. CONCLUSIONS The method reduces the noise in the reconstruction and increases the image quality. More data are introduced for each phase-correlated reconstruction, and therefore the applied dose is used more efficiently.

[1]  Berthold K. P. Horn,et al.  Determining Optical Flow , 1981, Other Conferences.

[2]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[3]  B. D. Lucas Generalized image matching by the method of differences , 1985 .

[4]  W A Kalender,et al.  Electrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. , 1998, Medical physics.

[5]  Willi A Kalender,et al.  Kymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart. , 2002, Medical physics.

[6]  Willi A Kalender,et al.  Extended parallel backprojection for standard three-dimensional and phase-correlated four-dimensional axial and spiral cone-beam CT with arbitrary pitch, arbitrary cone-angle, and 100% dose usage. , 2004, Medical physics.

[7]  M. V. van Herk,et al.  Respiratory correlated cone beam CT. , 2005, Medical physics.

[8]  J. Weickert,et al.  Lucas/Kanade meets Horn/Schunck: combining local and global optic flow methods , 2005 .

[9]  Ellen Yorke,et al.  Integrating respiratory gating into a megavoltage cone-beam CT system. , 2006, Medical physics.

[10]  L Xing,et al.  Motion correction for improved target localization with on-board cone-beam computed tomography , 2006, Physics in medicine and biology.

[11]  Günter Lauritsch,et al.  Towards cardiac C-arm computed tomography , 2006, IEEE Transactions on Medical Imaging.

[12]  P. Munro,et al.  Four-dimensional cone-beam computed tomography using an on-board imager. , 2006, Medical physics.

[13]  Uwe Oelfke,et al.  Linac-integrated 4D cone beam CT: first experimental results , 2006, Physics in medicine and biology.

[14]  L. Xing,et al.  Optimizing 4D cone-beam CT acquisition protocol for external beam radiotherapy. , 2007, International journal of radiation oncology, biology, physics.

[15]  T. M. Guerrero,et al.  Four-dimensional cone beam CT with adaptive gantry rotation and adaptive data sampling. , 2007, Medical physics.

[16]  Lei Xing,et al.  Enhanced 4D cone-beam CT with inter-phase motion model. , 2007, Medical physics.

[17]  S. Rit,et al.  On-the-fly motion-compensated cone-beam CT using an a priori model of the respiratory motion. , 2009, Medical physics.