Error corrections for an automated time-domain network analyzer
暂无分享,去创建一个
The errors are considered for an automated time-domain network analyzer based on a Tektronix sampling oscilloscope. Procedures are presented for removing the errors produced by two important sources: unwanted reflections associated with discontinuities and nonlinearities in the time base of the oscilloscope. The effectiveness of the error correction procedures is demonstrated by measurements of standard terminations with known reflection coefficients.
[1] L. Susman,et al. Applications of Time-Domain Metrology to the Automation of Broad-Band Microwave Measurements , 1972 .
[2] P. R. Rigg,et al. Low-cost computer-based time-domain microwave network analyser , 1980 .
[3] W. L. Gans,et al. Continuous and discrete Fourier transforms of steplike waveforms , 1982, IEEE Transactions on Instrumentation and Measurement.
[4] A. M. Nicolson. Forming the fast Fourier transform of a step response in time-domain metrology , 1973 .