Image based evaluation of mediastinal constraints for the development of a pulsatile total artificial heart

BackgroundGood anatomical compatibility is an important aspect in the development of cardiovascular implants. This work analyzes the interaction of the pump unit of an electrically driven pulsatile Total Artificial Heart (TAH) and the mediastinum. For an adequate compliance, both overall dimensions and alignment of inlets and outlets must be matched.MethodsCross-sectional medical image data of 27 individuals, including male and female patients suffering from end stage heart failure, was segmented and reconstructed to three dimensional (3D) surface models. Dimensions and orientations of relevant structures were identified and analyzed. The TAH surface model was virtually placed in orthotopic position and aligned with atrioventricular valves and big vessels. Additionally seven conventional cadaver studies were performed to validate different pump chamber designs based on virtual findings. Thereby 3D-coordinates were captured and introduced to the virtual environment to allow quantitative comparison between different individuals.ResultsSpatial parameters varied more in male patients with higher values if heart failure persists. Good correlation of the virtual analysis both to literature data and conventional cadaver studies could be shown. The full data of the 27 individuals as well as the summarized values found in literature are enclosed in the appendix. By superimposing the TAH-volume model to the anatomy, various misalignments were found and the TAH-design was adjusted.ConclusionsVirtual fitting allows implant design adjustments in realistic anatomy which has not been influenced by thoracotomy. Higher numbers of relevant individuals can be reasonably investigated in the virtual environment and quantitatively correlated. Using this approach, conventional cadaver studies can be significantly reduced but not obviated, due to the unavailable haptic feedback and immobility of potentially compressed structures.

[1]  Y Nosé,et al.  Intraoperative determination of mediastinal constraints for a total artificial heart. , 1991, ASAIO transactions.

[2]  Y Nosé,et al.  Anatomic fit study for development of a one piece total artificial heart. , 1991, ASAIO transactions.

[3]  T. Komoda,et al.  Three-dimensional imaging of mitral and tricuspid annuli for total artificial heart implantation. , 1992, Artificial organs.

[4]  C Uyama,et al.  Study of anatomic constraints using three dimensionally reconstructed images for total artificial heart implantation. , 1992, ASAIO journal.

[5]  D. Chatel,et al.  Criteria for anatomical compatibility of the total artificial heart: computerized three-dimensional modeling of the cardiovascular anatomy. , 1993, Artificial organs.

[6]  T. Komoda,et al.  Study of cardiac structures in relation to the pericardial cavity for total artificial heart implantation. , 1995, Artificial organs.

[7]  R. Dowling,et al.  Operative Techniques for Implantation of the AbioCor Total Artificial Heart , 2002 .

[8]  Jenny Pang,et al.  Prospective randomized trial of high thoracic epidural analgesia for coronary artery bypass surgery. , 2003, The Annals of thoracic surgery.

[9]  O H Frazier,et al.  The AbioCor implantable replacement heart. , 2003, The Annals of thoracic surgery.

[10]  O H Frazier,et al.  Surgery for Acquired Cardiovascular Disease Initial experience with the AbioCor Implantable Replacement Heart System , 2004 .

[11]  Nils Reiss,et al.  Total artificial heart-implantation technique using the CardioWest or the Thoratec system. , 2007, Multimedia manual of cardiothoracic surgery : MMCTS.

[12]  A. Fritschi,et al.  Analysis of flow patterns in the Total Artificial Heart “ReinHeart” , 2009 .

[13]  Mark S Slaughter,et al.  Transcutaneous Energy Transmission for Mechanical Circulatory Support Systems: History, Current Status, and Future Prospects , 2010, Journal of cardiac surgery.

[14]  Marvin J Slepian,et al.  The SynCardia temporary total artificial heart-evolving clinical role and future status , 2011 .

[15]  Kay Hameyer,et al.  Drive optimisation of a pulsatile Total Artificial Heart , 2011 .

[16]  J. Copeland,et al.  Experience with more than 100 total artificial heart implants. , 2012, The Journal of thoracic and cardiovascular surgery.

[17]  Kay Hameyer,et al.  Experimental validation of the linear drive train for a total artificial heart system , 2013 .