A novel technique for VMAT QA with EPID in cine mode on a Varian TrueBeam linac.

Volumetric modulated arc therapy (VMAT) is a relatively new treatment modality for dynamic photon radiation therapy. Pre-treatment quality assurance (QA) is necessary and many efforts have been made to apply electronic portal imaging device (EPID)-based IMRT QA methods to VMAT. It is important to verify the gantry rotation speed during delivery as this is a new variable that is also modulated in VMAT. In this paper, we present a new technique to perform VMAT QA using an EPID. The method utilizes EPID cine mode and was tested on Varian TrueBeam in research mode. The cine images were acquired during delivery and converted to dose matrices after profile correction and dose calibration. A sub-arc corresponding to each cine image was extracted from the original plan and its portal image prediction was calculated. Several analyses were performed including 3D γ analysis (2D images + gantry angle axis), 2D γ analysis, and other statistical analyses. The method was applied to 21 VMAT photon plans of 3 photon energies. The accuracy of the cine image information was investigated. Furthermore, this method's sensitivity to machine delivery errors was studied. The pass rate (92.8 ± 1.4%) for 3D γ analysis was comparable to those from Delta(4) system (99.9 ± 0.1%) under similar criteria (3%, 3 mm, 5% threshold and 2° angle to agreement) at 6 MV. The recorded gantry angle and start/stop MUs were found to have sufficient accuracy for clinical QA. Machine delivery errors can be detected through combined analyses of 3D γ, gantry angle, and percentage dose difference. In summary, we have developed and validated a QA technique that can simultaneously verify the gantry angle and delivered MLC fluence for VMAT treatment.This technique is efficient and its accuracy is comparable to other QA methods.

[1]  Jürgen Hesser,et al.  Breath-hold target localization with simultaneous kilovoltage/megavoltage cone-beam computed tomography and fast reconstruction. , 2010, International journal of radiation oncology, biology, physics.

[2]  Peter B. Greer,et al.  An evaluation of cine-mode 3D portal image dosimetry for Volumetric Modulated Arc Therapy , 2010 .

[3]  Jürgen Hesser,et al.  Fast Kilovoltage/Megavoltage (KVMV) Breathhold Cone-beam-CT for Image-guided-radiotherapy of Lung Cancer , 2010 .

[4]  Tom Depuydt,et al.  The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. , 2004, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  Fang-Fang Yin,et al.  Volumetric arc intensity-modulated therapy for spine body radiotherapy: comparison with static intensity-modulated treatment. , 2009, International journal of radiation oncology, biology, physics.

[6]  Cedric X. Yu,et al.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. , 2003, Medical physics.

[7]  Luca Cozzi,et al.  Volumetric modulated arc radiotherapy for carcinomas of the oro-pharynx, hypo-pharynx and larynx: a treatment planning comparison with fixed field IMRT. , 2009, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[8]  Qiuwen Wu,et al.  Independent verification of gantry angle for pre-treatment VMAT QA using EPID , 2012, Physics in medicine and biology.

[9]  A Monte Carlo investigation of contaminant electrons due to a novel in vivo transmission detector. , 2011, Physics in medicine and biology.

[10]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[11]  L. Feldkamp,et al.  Practical cone-beam algorithm , 1984 .

[12]  C. Ling,et al.  Commissioning and quality assurance of RapidArc radiotherapy delivery system. , 2008, International journal of radiation oncology, biology, physics.

[13]  A Manikandan,et al.  Quality assurance of dynamic parameters in volumetric modulated arc therapy. , 2012, The British journal of radiology.

[14]  Lucie Berger,et al.  Performance optimization of the Varian aS500 EPID system , 2006, Journal of applied clinical medical physics.

[15]  P B Greer,et al.  Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT. , 2009, Medical physics.

[16]  David A Jaffray,et al.  Intensity-modulated arc therapy with dynamic multileaf collimation : an alternative to tomotherapy , 2002 .

[17]  Jan-Jakob Sonke,et al.  3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry. , 2010, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  L Kumaraswamy,et al.  Using an EPID for patient-specific VMAT quality assurance. , 2011, Medical physics.

[19]  Cedric X. Yu,et al.  Clinical implementation of intensity-modulated arc therapy. , 2002, International journal of radiation oncology, biology, physics.

[20]  Peter B Greer,et al.  Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry. , 2005, Medical physics.

[21]  Clinical implementation of intensity-modulated arc therapy , 2000 .

[22]  Peter B. Greer,et al.  Gantry angle determination during arc IMRT: evaluation of a simple EPID‐based technique and two commercial inclinometers , 2012, Journal of applied clinical medical physics.

[23]  I. Rosen,et al.  Detection of IMRT delivery errors using a quantitative 2D dosimetric verification system. , 2004, Medical physics.

[24]  Karl Otto,et al.  Volumetric modulated arc therapy: IMRT in a single gantry arc. , 2007, Medical physics.

[25]  A. Nahum,et al.  IMAT-SIM: a new method for the clinical dosimetry of intensity-modulated arc therapy (IMAT). , 2007, Medical physics.

[26]  Stine Korreman,et al.  The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy , 2008, Radiation oncology.

[27]  P. Lambin,et al.  A literature review of electronic portal imaging for radiotherapy dosimetry. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[28]  Daliang Cao,et al.  Clinical implementation of intensity-modulated arc therapy. , 2011, Frontiers of radiation therapy and oncology.