Robot-assisted retinal vein cannulation with force-based puncture detection: Micron vs. the steady-hand eye robot

Retinal vein cannulation is a demanding procedure where therapeutic agents are injected into occluded retina veins. The feasibility of this treatment is limited due to challenges in identifying the moment of venous puncture, achieving cannulation and maintaining it throughout the drug delivery period. In this study, we integrate a force-sensing microneedle with two distinct robotic systems: the handheld micromanipulator Micron, and the cooperatively controlled Steady-Hand Eye Robot (SHER). The sensed tool-to-tissue interaction forces are used to detect venous puncture and extend the robots' standard control schemes with a new position holding mode (PHM) that assists the operator hold the needle position fixed and maintain cannulation for a longer time with less trauma on the vasculature. We evaluate the resulting systems comparatively in a dry phantom, stretched vinyl membranes. Results have shown that modulating the admittance control gain of SHER alone is not a very effective solution for preventing the undesired tool motion after puncture. However, after using puncture detection and PHM the deviation from the puncture point is significantly reduced, by 65% with Micron, and by 95% with SHER representing a potential advantage over freehand for both.

[1]  Tien Yin Wong,et al.  The prevalence of retinal vein occlusion: pooled data from population studies from the United States, Europe, Asia, and Australia. , 2010, Ophthalmology.

[2]  J. Weiss,et al.  Injection of tissue plasminogen activator into a branch retinal vein in eyes with central retinal vein occlusion. , 2001, Ophthalmology.

[3]  Ingrid U. Scott,et al.  Retinal vein occlusion. , 2010, British medical journal.

[4]  Jung Kim,et al.  Design of a novel tremor suppression device using a linear delta manipulator for micromanipulation , 2013, 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[5]  B. Nelson,et al.  Characterization of puncture forces for retinal vein cannulation , 2011 .

[6]  Russell H. Taylor,et al.  Force-based puncture detection and active position holding for assisted retinal vein cannulation , 2015, 2015 IEEE International Conference on Multisensor Fusion and Integration for Intelligent Systems (MFI).

[7]  Sadaf Hamid,et al.  Branch Retinal Vein Occlusion , 2009 .

[8]  Russell H. Taylor,et al.  Effects of micro-vibratory modulation during robot-assisted membrane peeling , 2015, 2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS).

[9]  Fumihito Arai,et al.  Quantitative assessment of manual and robotic microcannulation for eye surgery using new eye model , 2015, The international journal of medical robotics + computer assisted surgery : MRCAS.

[10]  Guang-Zhong Yang,et al.  An ungrounded hand-held surgical device incorporating active constraints with force-feedback , 2013, 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[11]  Russell H. Taylor,et al.  A comparative study for robot assisted vitreoretinal surgery: Micron vs. the Steady-Hand Robot , 2013, 2013 IEEE International Conference on Robotics and Automation.

[12]  Russell H. Taylor,et al.  Force sensing micro-forceps for robot assisted retinal surgery , 2012, 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[13]  Dominiek Reynaerts,et al.  Design and realisation of a novel robotic manipulator for retinal surgery , 2013, 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[14]  Russell H. Taylor,et al.  A sub-millimetric, 0.25 mN resolution fully integrated fiber-optic force-sensing tool for retinal microsurgery , 2009, International Journal of Computer Assisted Radiology and Surgery.

[15]  Nabil Simaan,et al.  Enabling Technology for Microvascular Stenting in Ophthalmic Surgery , 2010 .

[16]  Russell H. Taylor,et al.  Towards robot-assisted vitreoretinal surgery: Force-sensing micro-forceps integrated with a handheld micromanipulator , 2014, 2014 IEEE International Conference on Robotics and Automation (ICRA).

[17]  Sungwook Yang,et al.  Towards vision-based control of a handheld micromanipulator for retinal cannulation in an eyeball phantom , 2012, 2012 4th IEEE RAS & EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob).

[18]  Matthew R. Glucksberg,et al.  In vivo micropuncture of retinal vessels , 2004, Graefe's Archive for Clinical and Experimental Ophthalmology.

[19]  W. T. Latt,et al.  A compact hand-held active physiological tremor compensation instrument , 2009, 2009 IEEE/ASME International Conference on Advanced Intelligent Mechatronics.

[20]  Akira Arakawa,et al.  An experimental study of retinal endovascular surgery with a microfabricated needle. , 2011, Investigative ophthalmology & visual science.

[21]  Jin U. Kang,et al.  Force sensing micro-forceps with integrated fiber Bragg grating for vitreoretinal surgery , 2012, Other Conferences.

[22]  Cameron N. Riviere,et al.  Physiological tremor amplitude during retinal microsurgery , 2002, Proceedings of the IEEE 28th Annual Northeast Bioengineering Conference (IEEE Cat. No.02CH37342).

[23]  Russell H. Taylor,et al.  Force-sensing microneedle for assisted retinal vein cannulation , 2014, IEEE SENSORS 2014 Proceedings.

[24]  S S Hayreh,et al.  Retinal vein occlusion. , 1976, Indian journal of ophthalmology.

[25]  J. Edward Colgate,et al.  Toward robot-assisted vascular microsurgery in the retina , 2005, Graefe's Archive for Clinical and Experimental Ophthalmology.

[26]  Russell H. Taylor,et al.  Cooperative Robot Assistant for Retinal Microsurgery , 2008, MICCAI.

[27]  Rajnikant V. Patel,et al.  An active handheld device for compensation of physiological tremor using an ionic polymer metallic composite actuator , 2013, 2013 IEEE/RSJ International Conference on Intelligent Robots and Systems.