Durable microfabricated high-speed humidity sensors.
暂无分享,去创建一个
We describe a durable microfabricated humidity sensor made of interdigitated rhodium electrodes on a silicon substrate covered with a sensing film of Nafion perfluorosulfonate ionomer. Rhodium electrodes are much less prone to oxidative degradation compared to previously described gold electrode-based sensors. Even with dc excitation, Rh electrode sensors exhibit excellent long-term response stability. It has been found that low-amplitude (+/-1 V) square wave excitation can prolong the usability of gold electrode-based sensors to at least several months; however, this mode of interrogation cannot provide subsecond response times. Rhodium deposition on the microsensors is much more difficult than that of gold. We were able to attain crack-free Rh deposits by adaptation of pulsed electroplating techniques. At excitation voltages of >2 V dc, the Rh sensors respond to moisture with 10 <--> 90% rise and fall times of 30-50 ms. These are the fastest microfabricated water vapor sensors reported to date. We demonstrate applications as a breath monitor. Such sensors should also be of utility in atmospheric eddy measurements. Short-term repeatability is better than 0.6% RSD (n = 7).