Abstract.A new prototype of a portable differential photoacoustic
measurement system was designed and built. The system applies the gas filter
correlation method. A blackbody radiator is used as a radiation source,
which allows measurements of great variety of different IR absorbing gases.
The prototype exploits the high sensitivity of a cantilever pressure sensor.
The differential design allows a selective real time measurement for a
single gas in open air or from the flowing sample. For the first time the
photoacoustic cell geometry for an acceleration noise damping was integrated
to a differential detector. The system has potential applications anywhere
that sensitive in-situ-measurements are required: for example in process
control technology, air quality monitoring, exhale monitoring, alarm devices
and food industry. The measurement of nitric oxide was modeled in the
presence of water vapor. As an example the concentration of water vapor,
that is acceptable if ppb-level nitric oxide measurements are to be done,
was calculated. With a typical band pass filter and a blackbody radiator the
amount is 20 ppm with 1.3 s integration time. The concentration was also
calculated to diode laser operating at 2.663 μm and quantum cascade
laser at 5.263 μm. The respective concentrations were 2 ppm and 6600 ppm.
[1]
A. Cho,et al.
Spectroscopic detection of biological NO with a quantum cascade laser
,
2001,
Applied physics. B, Lasers and optics.
[2]
K. Roth,et al.
Cantilever enhanced photoacoustic detection of carbon dioxide using a tunable diode laser source
,
2007
.
[3]
Jyrki Kauppinen,et al.
Selective differential photoacoustic method for trace gas analysis
,
2005
.
[4]
V. Koskinen,et al.
High sensitivity in gas analysis with photoacoustic detection
,
2004
.
[5]
Jyrki Kauppinen,et al.
Extremely sensitive trace gas analysis with modern photoacoustic spectroscopy
,
2006
.
[6]
Gaetano Scamarcio,et al.
Photoacoustic Spectroscopy with Quantum Cascade Lasers for Trace Gas Detection
,
2006,
Sensors (Basel, Switzerland).
[7]
Tao Chen,et al.
In situ gas filter correlation: photoacoustic CO detection method for fire warning
,
2005
.