Enhancements of NMC'S Josephson array voltage standard system
暂无分享,去创建一个
A series of system enhancements for the Josephson array voltage standard (JAVS) have been implemented to improve the efficiency and quality of dc voltage dissemination in the National Metrology Centre, A*STAR. Periodic direct comparison between two JAVS systems has been adopted as a quality assurance procedure to ensure accurate reproduction of the volt for calibration of secondary standards. A miniature millimeter-wave synthesizer is employed for better control of microwave power and locking frequency. Low thermal terminal reversing switches are introduced to minimize zero-offset measurement uncertainty. Monte Carlo method and ANOVA method are implemented in the data analysis and reporting.
[1] Yi-hua Tang,et al. Evaluating the uncertainty of Josephson voltage standards , 1999 .
[2] S. W. Chua,et al. A direct comparison of Josephson Array Voltage Standards , 2008, 2008 Conference on Precision Electromagnetic Measurements Digest.