ON THE WAY TO A FULLY INTEGRATED DNA-PURIFICATION SYSTEM ON A STANDARD LABORATORY CENTRIFUGE

For the first time, we combine liquid reagent storage in glass capillaries [1], time-controlled reagent release [2], and a novel solution for unidirectional centrifugal routing, paving the way to a fully integrated DNA extraction on a standard laboratory centrifuge. The average yield of the extraction is 192 ± 30 ng DNA out of 32 μl blood, corresponding to 53 ± 8 % of a reference extraction. The overall processing time of ~66 minutes can be reduced to ~8 minutes after optimization of the burst-valves. This novel approach demonstrates a convenient way for fully automated DNAextraction on a standard laboratory centrifuge.