Automation of routine elements for spot‐scanning proton patient‐specific quality assurance

PURPOSE At our institution, all proton patient plans undergo patient-specific quality assurance (PSQA) prior to treatment delivery. For intensity-modulated proton beam therapy, quality assurance is complex and time consuming, and it may involve multiple measurements per field. We reviewed our PSQA workflow and identified the steps that could be automated and developed solutions to improve efficiency. METHODS We used the treatment planning system's (TPS) capability to support C# scripts to develop an Eclipse scripting application programming interface (ESAPI) script and automate the preparation of the verification phantom plan for measurements. A local area network (LAN) connection between our measurement equipment and shared database was established to facilitate equipment control, measurement data transfer, and storage. To improve the analysis of the measurement data, a Python script was developed to automatically perform a 2D-3D γ-index analysis comparing measurements in the plane of a two-dimensional detector array with TPS predictions in a water phantom for each acquired measurement. RESULTS Device connection via LAN granted immediate access to the plan and measurement information for downstream analysis using an online software suite. Automated scripts applied to verification plans reduced time from preparation steps by at least 50%; time reduction from automating γ-index analysis was even more pronounced, dropping by a factor of 10. On average, we observed an overall time savings of 55% in completion of the PSQA per patient plan. CONCLUSIONS The automation of the routine tasks in the PSQA workflow significantly reduced the time required per patient, reduced user fatigue, and frees up system users from routine and repetitive workflow steps allowing increased focus on evaluating key quality metrics.

[1]  D. Low,et al.  A technique for the quantitative evaluation of dose distributions. , 1998, Medical physics.

[2]  Jan-Jakob Sonke,et al.  A fast algorithm for gamma evaluation in 3D. , 2007, Medical physics.

[3]  Michael Gillin,et al.  Verification of patient-specific dose distributions in proton therapy using a commercial two-dimensional ion chamber array. , 2010, Medical physics.

[4]  Bijan Arjomandy,et al.  Patient-specific quality assurance for prostate cancer patients receiving spot scanning proton therapy using single-field uniform dose. , 2011, International journal of radiation oncology, biology, physics.

[5]  Heng Li,et al.  Improving spot-scanning proton therapy patient specific quality assurance with HPlusQA, a second-check dose calculation engine. , 2013, Medical physics.

[6]  Heng Li,et al.  Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance. , 2013, Medical physics.

[7]  Charles Holmes,et al.  Spot-Scanning Proton Therapy Patient- Specific Quality Assurance: Results from 309 Treatment Plans , 2014 .

[8]  David Followill,et al.  Comparison of 2D and 3D gamma analyses. , 2014, Medical physics.

[9]  T. Solberg,et al.  Use of a novel two-dimensional ionization chamber array for pencil beam scanning proton therapy beam quality assurance. , 2015, Journal of applied clinical medical physics.

[10]  K. L. Poole,et al.  Three-dimensional gamma criterion for patient-specific quality assurance of spot scanning proton beams. , 2015, Journal of applied clinical medical physics.

[11]  Xiaodong Zhang,et al.  Towards Effective and Efficient Patient-Specific Quality Assurance for Spot Scanning Proton Therapy , 2015, Cancers.

[12]  S Giordanengo,et al.  Quality assurance of carbon ion and proton beams: A feasibility study for using the 2D MatriXX detector. , 2016, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[13]  D C Weber,et al.  Assessing the quality of proton PBS treatment delivery using machine log files: comprehensive analysis of clinical treatments delivered at PSI Gantry 2 , 2016, Physics in medicine and biology.

[14]  Martin Bues,et al.  Clinical Implementation of a Proton Dose Verification System Utilizing a GPU Accelerated Monte Carlo Engine. , 2016, International journal of particle therapy.

[15]  Alessandra Bolsi,et al.  Treatment log files as a tool to identify treatment plan sensitivity to inaccuracies in scanned proton beam delivery. , 2017, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.