The use of automatic noninvasive blood pressure (NIBP) devices has become a common technique to monitor blood pressure intraoperatively. The usual cuff placement for these devices on the upper arm sometimes poses problems. As an alternative, many clinicians place the cuff on the ankle. This practice has not been previously investigated to determine its efficacy. The purpose of our study was to determine whether a noninvasive blood pressure cuff on the arm could be replaced by one on the ankle.We monitored 24 patients intraoperatively with two non-invasive blood pressure cuffs, one on the upper arm and one on the ankle. Systolic, diastolic, and mean pressures were obtained from each cuff placement at intervals of no shorter than 3 minutes. The time necessary to obtain the measurements and the presence of any artifact were also recorded.A total of 404 pairs of data were obtained and the systolic blood pressure ranged from 82 to 196 mm Hg. The mean and diastolic pressure readings were equivalent between the arm and ankle blood pressure readings. The systolic pressures were not equivalent, reflecting the fact that the ankle systolic blood pressure is physiologically higher than the arm systolic blood pressure. The difference between the times necessary to obtain the readings from arm or ankle was not statistically significant. Eight of the paired readings 2.0%) represented artifact, arbitrarily defined as a difference in mean blood pressure readings of 15 mm Hg between the arm and the ankle. Since the mean blood pressure readings obtained at the arm and at the ankle were statistically equivalent, we concluded that the ankle cuff placement provided a reliable alternative to the placement of the cuff on the arm.
[1]
G. Åkerström,et al.
Ankle-brachial index (ABI) and leg blood flow following epidural anaesthesia.
,
1988,
International Angiology.
[2]
A. Nicolaides,et al.
Intermittent sequential pneumatic compression of the legs and thromboembolism-deterrent stockings in the prevention of postoperative deep venous thrombosis.
,
1983,
Surgery.
[3]
Pollock Av,et al.
Intermittent sequential pneumatic compression of the legs in the prevention of venous stasis and postoperative deep venous thrombosis.
,
1980
.
[4]
W. Robertson,et al.
Cyclic sequential compression of the lower limb in prevention of deep venous thrombosis.
,
1982,
The Journal of bone and joint surgery. American volume.
[5]
E. Mühe.
Intermittent sequential high-pressure compression of the leg. A new method of preventing deep vein thrombosis.
,
1984,
American journal of surgery.
[6]
A K Ream,et al.
Indirect measurement of mean blood pressure in the anesthetized patient.
,
1979,
Anesthesiology.
[7]
D. Koh,et al.
Statistical evaluation of agreement between two methods for measuring a quantitative variable.
,
1989,
Computers in biology and medicine.