Micro gas analysis system for measurement of atmospheric hydrogen sulfide and sulfur dioxide.

A honeycomb structure microchannel scrubber was developed to achieve efficient and stable gas collection. A thin porous membrane was pasted on a microchannel by the adhesive force of a fresh polydimethylsiloxane surface. The microchannel scrubber achieved much more efficient gas collection than conventional impingers and diffusion scrubbers. Two sets of the microchannel scrubbers and detectors were integrated in a 10 cm x 9 cm plastic board to create a micro gas analysis system (microGAS) for simultaneous measurements of H2S and SO2. The whole system including a battery was incorporated in a carrying case 34 cm W x 16 cm D x 17 cm H for use in the field. Liquid flows at 30 microl min(-1) were obtained by bimetal micropumps. The estimated detection limits were 0.1 ppbv for H2S and 1 ppbv for SO2. The system was demonstrated for real atmospheric gas analysis, and the results agreed well with data concurrently obtained by ion chromatography coupled with a cylindrical diffusion scrubber. The system we developed allowed automated continuous analyses in the field and achieved a much higher time resolution compared to those by ion chromatographic analysis.

[1]  Fluorescent derivatization of nitrite ions with 2,3-diaminonaphthalene utilizing a pH gradient in a Y-shaped microchannel. , 2001, Analytical sciences : the international journal of the Japan Society for Analytical Chemistry.

[2]  T. Korenaga,et al.  Analytical Chemistry for Environmental and Human Health. Development of a gas absorption microchip using porous glass and its evaluation for analytical applications. , 2000 .

[3]  K. Toda Trends in Atmospheric Trace Gas Measurement Instruments with Membrane-based Gas Diffusion Scrubbers , 2004, Analytical sciences : the international journal of the Japan Society for Analytical Chemistry.

[4]  Abul Kalam Azad,et al.  On-site measurements of hydrogen sulfide and sulfur dioxide emissions from tidal flat sediments of Ariake Sea, Japan , 2005 .

[5]  M. Miró,et al.  Automated membrane-based sampling and sample preparation exploiting flow-injection analysis , 2004 .

[6]  K. Toda,et al.  Field instrument for simultaneous large dynamic range measurement of atmospheric hydrogen sulfide, methanethiol, and sulfur dioxide. , 2004, Environmental Science and Technology.

[7]  K. Toda,et al.  Micro-gas analysis system μGAS comprising a microchannel scrubber and a micro-fluorescence detector for measurement of hydrogen sulfide , 2004 .

[8]  Wouter Olthuis,et al.  Miniaturized measurement system for ammonia in air , 2004 .

[9]  A. Manz,et al.  Miniaturized total chemical analysis systems: A novel concept for chemical sensing , 1990 .

[10]  Air-cooled cold trap channel integrated in a microfluidic device for monitoring airborne BTEX with an improved detection limit. , 2002, Analytical chemistry.

[11]  A. Manz,et al.  Micro total analysis systems. Recent developments. , 2004, Analytical chemistry.

[12]  K. Toda,et al.  Fluorometric field instrument for continuous measurement of atmospheric hydrogen sulfide. , 2001, Analytical Chemistry.

[13]  Keisuke Toda Development of miniature key devices for flow analysis and their applications , 2004 .

[14]  K. Toda,et al.  Hybrid microfabricated device for field measurement of atmospheric sulfur dioxide. , 2002, Analytical chemistry.

[15]  O. Niwa,et al.  Microfluidic device for airborne BTEX detection. , 2001, Analytical chemistry.

[16]  Purnendu K. Dasgupta,et al.  A diffusion scrubber for the collection of atmospheric gases , 1984 .

[17]  Wouter Olthuis,et al.  Sampling small volumes of ambient ammonia using a miniaturized gas sampler. , 2004, Lab on a chip.