A gamma and neutron phoswich read out with SiPM for SPRD

Abstract A gamma and neutron phoswich was developed for spectroscopic personal radiation detectors (SPRDs). It consisted of a Φ 25  ×  25 mm NaI(Tl) crystal for gamma detection and a Φ 25  ×  3 mm LiI(Eu) crystal for neutron detection. The phoswich was read out by 8  ×  8 ch SiPM array (24  ×  24 mm 2 ). Radiations in NaI(Tl) and LiI(Eu) were discriminated by pulse shape, while gammas and neutrons in LiI(Eu) were separated by pulse amplitude. For the LiI(Eu), the gamma equivalent energy for thermal neutrons was measured as 3.6 ± 0.1 MeV, providing satisfactory gamma rejection. For NaI(Tl), the response of SiPM array was well linear in the energy range up to 1408 keV, at which a deviation less than 2% was measured. Digital pulse shape discrimination (PSD) was implemented with an 8-bit digitizer running at 50 MSPS sampling rate and offline analysis. The signal pulses from NaI(Tl) and LiI(Eu) showed significant difference in falling edge allowing effective PSD. The best figure of merit (FOM) was measured as 4.4 ± 0.2 with optimized parameters, providing excellent PSD performance. The energy resolutions for 661.6 keV gamma rays in NaI(Tl) and thermal neutrons in LiI(Eu) were measured as 7.0 ± 0.2% and 11.2 ± 0.2% respectively, with selected PSD threshold.

[1]  Benjamin S. McDonald,et al.  A wearable sensor based on CLYC scintillators , 2016 .

[2]  D. Ramsden,et al.  A compact neutron detector based on the use of a SiPM detector , 2008, 2008 IEEE Nuclear Science Symposium Conference Record.

[3]  Characterization of 4×4ch MPPC array in scintillation spectrometry , 2013, 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record (NSS/MIC).

[4]  A. Alhawsawi,et al.  A compton-suppressed phoswich detector for gamma spectroscopy , 2013, Journal of Radioanalytical and Nuclear Chemistry.

[5]  Kanai S. Shah,et al.  Development of Cs2LiYCl6 scintillator , 2013 .

[6]  M. Grodzicka-Kobylka,et al.  Comparison of SensL and Hamamatsu 4×4 channel SiPM arrays in gamma spectrometry with scintillators , 2017 .

[7]  Nicholas Dallmann,et al.  A Cs 2 LiYCl 6 :Ce-based advanced radiation monitoring device , 2015 .

[8]  S. Pozzi,et al.  Pulse shape discrimination performance of stilbene coupled to low-noise silicon photomultipliers , 2015 .

[10]  M. Nakhostin Recursive algorithms for digital implementation of neutron/gamma discrimination in liquid scintillation detectors , 2012 .

[11]  M. Zuffa,et al.  Comparison between a silicon PIN diode and a CsI(Tl) coupled to a silicon PIN diode for dosimetric purpose in radiology , 2014 .

[12]  M. Moszynski,et al.  Energy resolution of small scintillation detectors with SiPM light readout , 2013 .

[13]  H. Ishibashi,et al.  Development of a three-layer phoswich alpha–beta–gamma imaging detector , 2015 .

[14]  Yanguo Li,et al.  A novel analog power supply for gain control of the Multi-Pixel Photon Counter (MPPC) , 2016, 1606.03727.

[15]  M. Moszynski,et al.  /sup 6/LiI(Eu) in neutron and /spl gamma/-ray spectrometry-a highly sensitive thermal neutron detector , 2005, IEEE Transactions on Nuclear Science.

[16]  Rhys M. Preston,et al.  Neutron-Gamma Pulse Shape Discrimination Using Organic Scintillators With Silicon Photomultiplier Readout , 2014, IEEE Transactions on Nuclear Science.

[17]  Biao Wang,et al.  NaI(Tl) scintillator read out with SiPM array for gamma spectrometer , 2017 .

[18]  F. Licciulli,et al.  A Novel Technique for the Stabilization of SiPM Gain Against Temperature Variations , 2013, IEEE Transactions on Nuclear Science.

[19]  G. Nebbia,et al.  A novel detector assembly for detecting thermal neutrons, fast neutrons and gamma rays , 2016 .

[20]  A. Alhawsawi,et al.  Real-time pulse-shape discrimination and beta-gamma coincidence detection in field-programmable gate array , 2013 .

[21]  General purpose pulse shape analysis for fast scintillators implemented in digital readout electronics , 2016 .

[22]  M. Moszynski,et al.  Characterization of large TSV MPPC arrays (4×4ch and 8×8 ch) in scintillation spectrometry , 2017 .