Goniometer with continuously rotating gratings for use as an angle standard
暂无分享,去创建一个
Abstract The paper describes an angle measuring system for the calibration of reference angle standards used by standardizing laboratories and industry. The machine has a range of 360° with a measurement uncertainty of 0.05 are sec. Measurements of angle gauges and a polygon are tabulated and discussed.
[1] A H McIlraith,et al. A radial grating dividing engine of high accuracy , 1967 .