Measurement of astigmatism arising from the internal ocular surfaces.

A method is described for measuring internal ocular surface (posterior cornea, anterior and posterior crystalline lens) astigmatism. This involves the use of videokeratography, A-scan ultrasonography, and autorefractometry along with multi-meridional phakometric measurements of Purkinje images I(anterior corneal surface) II(posterior corneal surface) and IV(posterior lens surface). Data was collected from both eyes of 66 subjects. Right and left eyes exhibited similar mean levels of astigmatism from the posterior corneal surface (R + 0.21 DC axis 82 degrees; L + 0.22 DC axis 80 degrees), anterior lens surface (R + 0.52 DC axis 8 degrees; L + 0.49 DC axis 165 degrees) and posterior lens surface (R + 1.48 DC axis 99 degrees; L + 1.16 DC axis 90 degrees). It was generally found that astigmatism arising from the anterior corneal and lens surfaces in conjunction with intraocular distance effectivity are almost completely compensated for by the posterior corneal and lens surface. Repeatability was assessed on 20 subjects. Although the methods is prone to accumulated experimental errors, these are random in nature so that the difference between repeat group averaged data never exceeded +/- 0.27 DC cylindrical component and +/- 6 degrees cylinder axis.

[1]  A. Kooijman,et al.  Light distribution on the retina of a wide-angle theoretical eye. , 1983, Journal of the Optical Society of America.

[2]  W. Long A MATHEMATICAL ANALYSIS OF MULTI‐MERIDIONAL REFRACTOMETRY* , 1974, American journal of optometry and physiological optics.

[3]  W M Lyle,et al.  CHANGES IN CORNEAL ASTIGMATISM WITH AGE* , 1971, American journal of optometry and archives of American Academy of Optometry.

[4]  T. Olsen,et al.  THE THICKNESS OF THE HUMAN CORNEA AS DETERMINED BY A SPECULAR METHOD , 1984, Acta ophthalmologica.

[5]  D. A. Barnes,et al.  Posterior Corneal Surface Toricity and Total Corneal Astigmatism , 1991, Optometry and vision science : official publication of the American Academy of Optometry.

[6]  M C Dunne,et al.  Normal variations of the posterior corneal surface , 1992, Acta ophthalmologica.

[7]  D. Edgar,et al.  Effects of different ocular fixation conditions on A‐scan ultrasound biometry measurements , 1992, Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians.

[8]  M. Dunne,et al.  A study of the axis of orientation of residual astigmatism , 1994, Acta ophthalmologica.

[9]  S. Wittenberg,et al.  MEASUREMENTS OF THE OCULAR DIOPTRIC ELEMENTS UTILIZING PHOTOGRAPHIC METHODS: PART II. CORNEA—THEORETICAL CONSIDERATIONS , 1966 .

[10]  M Millodot,et al.  Clinical evaluation of the Canon Autoref R-1. , 1985, American journal of optometry and physiological optics.

[11]  M. Dunne,et al.  Measurement of Posterior Corneal Surface Toricity , 1990, Optometry and vision science : official publication of the American Academy of Optometry.

[12]  D A Atchison,et al.  The optical modelling of the human lens , 1991, Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians.