A New Differential Pressure Flow Meter for Measurement of Human Breath Flow: Simulation and Experimental Investigation.

The development and performance characterization of a new differential pressure-based flow meter for human breath measurements is presented in this article. The device, called a "Confined Pitot Tube," is comprised of a pipe with an elliptically shaped expansion cavity located in the pipe center, and an elliptical disk inside the expansion cavity. The elliptical disk, named Pitot Tube, is exchangeable, and has different diameters, which are smaller than the diameter of the elliptical cavity. The gap between the disk and the cavity allows the flow of human breath to pass through. The disk causes an obstruction in the flow inside the pipe, but the elliptical cavity provides an expansion for the flow to circulate around the disk, decreasing the overall flow resistance. We characterize the new sensor flow experimentally and theoretically, using Comsol Multiphysics® software with laminar and turbulent models. We also validate the sensor, using inhalation and exhalation tests and a reference method.

[1]  Y. Shimada,et al.  Evaluation of a hot-wire respiratory flowmeter for clinical applicability. , 1979, Journal of applied physiology: respiratory, environmental and exercise physiology.

[2]  J. P. Conkle,et al.  Trace composition of human respiratory gas. , 1975, Archives of environmental health.

[3]  D C Muir,et al.  Acceptable levels for breathing resistance of respiratory apparatus. , 1973, Archives of environmental health.

[4]  S. Turney,et al.  A mathematical model for the ultrasonic measurement of respiratory flow , 2006, Medical and biological engineering.

[5]  A. N. Kolmogorov Equations of turbulent motion in an incompressible fluid , 1941 .

[6]  P. Tsilingiris Thermophysical and transport properties of humid air at temperature range between 0 and 100 °C , 2008 .

[7]  Ulrike Tisch,et al.  Chemical sensors for breath gas analysis: the latest developments at the Breath Analysis Summit 2013 , 2014, Journal of breath research.

[8]  F. Sciurba,et al.  The accuracy of pneumotachograph measurements during mechanical ventilation. , 1996, American journal of respiratory and critical care medicine.

[9]  Ki-Hyun Kim,et al.  A review of breath analysis for diagnosis of human health , 2012 .

[10]  Paul J. LaNasa,et al.  Types of Fluid Flow Measurement , 2014 .

[11]  D. Wilcox Formulation of the k-w Turbulence Model Revisited , 2008 .

[12]  H. Haugen,et al.  Indirect calorimetry: a practical guide for clinicians. , 2007, Nutrition in clinical practice : official publication of the American Society for Parenteral and Enteral Nutrition.

[13]  Ranganath Krishnan,et al.  A pocket-sized metabolic analyzer for assessment of resting energy expenditure. , 2014, Clinical nutrition.

[14]  R M Gardner,et al.  Turbine flowmeter vs. Fleisch pneumotachometer: a comparative study for exercise testing. , 1987, Journal of applied physiology.

[15]  R. Withers,et al.  Evaluation of five small turbine-type respirometers used in adult anesthesia , 1993, Journal of Clinical Monitoring.