Force and Position Control System for Freehand Ultrasound

A hand-held force-controlled ultrasound probe has been developed for medical imaging applications. The probe-patient contact force can be held constant to improve image stability, swept through a range, or cycled. The mechanical portion of the device consists of a ball screw linear actuator driven by a servo motor, along with a load cell, accelerometer, and limit switches. The performance of the system was assessed in terms of the frequency response to simulated sonographer hand motion and in hand-held image feature tracking during simulated patient motion. The system was found to attenuate contact force variation by 97% at 0.1 Hz, 83% at 1 Hz, and 33% 10 Hz, a range that spans the typical human hand tremor frequency spectrum. In studies with 15 human operators, the device applied the target contact force with ten times less variation than in conventional ultrasound imaging. An ergonomic, human-in-the-loop, imaging-workflow enhancing control scheme, which combines both force- and position-control, permits smooth making and breaking of probe-patient contact, and helps the operator keep the probe centered within its range of motion. By controlling ultrasound probe contact force and consequently the amount of tissue deformation, the system enhances the repeatability, usability, and diagnostic capabilities of ultrasound imaging.

[1]  N Magnavita,et al.  [Frequency of musculoskeletal symptoms in diagnostic medical sonographers. Results of a pilot survey]. , 1999, La Radiologia medica.

[2]  Cameron N. Riviere,et al.  Micron: An Actively Stabilized Handheld Tool for Microsurgery , 2012, IEEE Transactions on Robotics.

[3]  Subhash Rakheja,et al.  a Comparison of Biodynamic Models of the Human HAND-ARM System for Applications to Hand-Held Power Tools , 2002 .

[4]  Richard G Barr,et al.  Effects of Precompression on Elasticity Imaging of the Breast , 2012, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[5]  Neville Hogan,et al.  Impedance Control: An Approach to Manipulation: Part I—Theory , 1985 .

[6]  Magda Osman,et al.  Control Systems Engineering , 2010 .

[7]  Jin Young Kwak,et al.  Interobserver variability of ultrasound elastography: how it affects the diagnosis of breast lesions. , 2011, AJR. American journal of roentgenology.

[8]  R Willinger,et al.  Copolymer-in-oil phantom materials for elastography. , 2009, Ultrasound in medicine & biology.

[9]  M. Gooding,et al.  A system for simultaneously measuring contact force, ultrasound, and position information for use in force-based correction of freehand scanning , 2005, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[10]  P. Asbach,et al.  Noninvasive assessment of the rheological behavior of human organs using multifrequency MR elastography: a study of brain and liver viscoelasticity , 2007, Physics in medicine and biology.

[11]  T A Krouskop,et al.  Nonlinear stress-strain relationships in tissue and their effect on the contrast-to-noise ratio in elastograms. , 2000, Ultrasound in medicine & biology.

[12]  D. Rubens,et al.  Imaging the elastic properties of tissue: the 20 year perspective , 2011, Physics in medicine and biology.

[13]  Alexandre Krupa,et al.  Active stabilization of ultrasound image for robotically-assisted medical procedures , 2012 .

[14]  R. Elble,et al.  Mechanistic components of normal hand tremor. , 1978, Electroencephalography and clinical neurophysiology.

[15]  M. Tobin,et al.  Validity and reliability of rectus femoris ultrasound measurements: Comparison of curved-array and linear-array transducers. , 2014, Journal of rehabilitation research and development.

[16]  Matthew W. Gilbertson,et al.  Assessing duchenne muscular dystrophy with force-controlled ultrasound , 2014, 2014 IEEE 11th International Symposium on Biomedical Imaging (ISBI).

[17]  Matthew W. Gilbertson,et al.  Handheld Force-Controlled Ultrasound Probe , 2013 .

[18]  Charles W. Suggs,et al.  Hand-arm vibration part III: A distributed parameter dynamic model of the human hand-arm system , 1978 .

[19]  Gregory D. Hager,et al.  Needle Guidance Using Handheld Stereo Vision and Projection for Ultrasound-Based Interventions , 2014, MICCAI.

[20]  Vilchis-Gonzalez,et al.  TERMI Robot , 2007, Electronics, Robotics and Automotive Mechanics Conference (CERMA 2007).

[21]  N. Hogan,et al.  Impedance Control:An Approach to Manipulation,Parts I,II,III , 1985 .

[22]  Robert D. Howe,et al.  Force tracking with feed-forward motion estimation for beating heart surgery , 2010, IEEE Transactions on Robotics.

[23]  J Alison Noble,et al.  Slip imaging: reducing ambiguity in breast lesion assessment. , 2010, Ultrasound in medicine & biology.

[24]  Matthew W. Gilbertson,et al.  Impedance-controlled ultrasound probe , 2011, Medical Imaging.

[25]  Eric C Ford,et al.  Use of respiratory-correlated four-dimensional computed tomography to determine acceptable treatment margins for locally advanced pancreatic adenocarcinoma. , 2010, International journal of radiation oncology, biology, physics.

[26]  Matthew W. Gilbertson,et al.  Electromechanical systems to enhance the usability and diagnostic capabilities of ultrasound imaging , 2014 .

[27]  Kevin Cleary,et al.  Robotically assisted ultrasound interventions , 2008, SPIE Medical Imaging.

[28]  Wen-Hong Zhu,et al.  Robot-Assisted Diagnostic Ultrasound - Design and Feasibility Experiments , 1999, MICCAI.

[29]  M. Kadour,et al.  Assisted-freehand ultrasound elasticity imaging , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[30]  Ivo A. M. J. Broeders,et al.  Robotics revolutionizing surgery: the Intuitive Surgical “Da Vinci” system , 2001 .

[31]  Philippe Cinquin,et al.  A new robot architecture for tele-echography , 2003, IEEE Trans. Robotics Autom..

[32]  Matthew W. Gilbertson,et al.  An ergonomic, instrumented ultrasound probe for 6-axis force/torque measurement , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[33]  Michael Burcher,et al.  A novel ultrasound indentation system for measuring biomechanical properties of in vivo soft tissue. , 2003, Ultrasound in medicine & biology.

[34]  Matthew W. Gilbertson,et al.  Evaluating the clinical relevance of force-correlated ultrasound , 2014, 2014 IEEE 11th International Symposium on Biomedical Imaging (ISBI).

[35]  M. Fink,et al.  Shear elasticity probe for soft tissues with 1-D transient elastography , 2002, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[36]  A. Ahuja,et al.  Ultrasound elastography in the head and neck. Part I. Basic principles and practical aspects , 2013, Cancer imaging : the official publication of the International Cancer Imaging Society.

[37]  Nariman Sepehri,et al.  A novel hand-controller for remote ultrasound imaging , 2008 .

[38]  Jocelyne Troccaz,et al.  A one-DOF freehand haptic device for robotic tele-echography. , 2004, Studies in health technology and informatics.

[39]  Michael Goldfarb,et al.  Modeling the human hand as it interacts with a telemanipulation system , 2005 .

[40]  H. Akima,et al.  Estimation of muscle atrophy based on muscle thickness in knee surgery patients , 2014, Clinical physiology and functional imaging.

[41]  Matthew W. Gilbertson,et al.  6-DOF probe tracking via skin mapping for freehand 3D ultrasound , 2013, 2013 IEEE 10th International Symposium on Biomedical Imaging.

[42]  Matthew W. Gilbertson,et al.  Ergonomic control strategies for a handheld force-controlled ultrasound probe , 2012, 2012 IEEE/RSJ International Conference on Intelligent Robots and Systems.

[43]  Robert Rohling,et al.  A needle tracking device for ultrasound guided percutaneous procedures. , 2005, Ultrasound in medicine & biology.

[44]  S. Wojcinski,et al.  Acoustic radiation force impulse imaging with virtual touch tissue quantification: measurements of normal breast tissue and dependence on the degree of pre-compression. , 2013, Ultrasound in medicine & biology.

[45]  Akio Morita,et al.  Robot-assisted vitreoretinal surgery: development of a prototype and feasibility studies in an animal model. , 2009, Ophthalmology.

[46]  Matthew M. Williamson,et al.  Series elastic actuators , 1995, Proceedings 1995 IEEE/RSJ International Conference on Intelligent Robots and Systems. Human Robot Interaction and Cooperative Robots.

[47]  Robert Rohling,et al.  An Active Dynamic Vibration Absorber for a Hand-Held Vibro-Elastography Probe , 2007 .

[48]  N. Djelal,et al.  A removable device for axial force and orientation measurement on medical ultrasound probe , 2012, International Multi-Conference on Systems, Sygnals & Devices.

[49]  Etienne Dombre,et al.  HIPPOCRATE: an intrinsically safe robot for medical applications , 1998, Proceedings. 1998 IEEE/RSJ International Conference on Intelligent Robots and Systems. Innovations in Theory, Practice and Applications (Cat. No.98CH36190).

[50]  Dale A. Lawrence Stability and transparency in bilateral teleoperation , 1993, IEEE Trans. Robotics Autom..

[51]  Peter Kazanzides,et al.  Development and Application of a New Steady-Hand Manipulator for Retinal Surgery , 2007, Proceedings 2007 IEEE International Conference on Robotics and Automation.

[52]  Neville Hogan,et al.  Impedance Control: An Approach to Manipulation , 1984, 1984 American Control Conference.

[53]  J. Ophir,et al.  Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues , 1991, Ultrasonic imaging.