An innovative method for 192Ir HDR calibration by farmer chamber, V-film, and solid phantom

Abstract A simple, practical and economical technique was proposed to calibrate an 192 Ir HDR brachytherapy source in terms of air kerma strength. This technique makes use of the 0.6 cm 3 Farmer type ion chamber, radiographic film and polystyrene phantom. These tools are commonly used for dosimetry quality assurance of the clinical linear accelerator. In this study, the Exradin A19, PTW N30004 and TM30001 Farmer type ion chambers were used for the calibration of the 192 Ir HDR source. To perform the calibration, a 25.4×30.5 cm 2 radiographic film was taped on a piece of polystyrene plate, and a straight applicator probe of a HDR brachytherapy unit and the Farmer type ion chamber were affixed to the film envelope. The film was irradiated by the 192 Ir source, followed by an exposure in the simulator X-ray beam. The film set with the film removed was then placed on a 5 cm thick polystyrene phantom for calibration measurement. Based on the electrometer reading from the Farmer type ion chamber irradiated by 192 Ir and the measured source-to-chamber distance by means of the images on the developed film, we can calculate the air kerma strength of the 192 Ir using the new technique. Our calibration results were compared to the data provided by the manufacturer and that of five different well type ion chambers, namely, Sun Nuclear cooperation (SNC) 1008, Nucletron SDS 077.091, SDS 077.094, PTW TN33004 and Standard Imaging (SI) HDR-1000 Plus. The differences were all within 1.6%. Relative to the “7-distance measurement technique” by Stump et al., 2002, our method is more efficient if our empirical formula was used. In summary, our method is simpler and cost-effective to calibrate an 192 Ir HDR brachytherapy source for those hospitals without a calibration jig or a well type ion chamber.

[1]  D W Pearson,et al.  Calibration of 192Ir high-dose-rate afterloading systems. , 1991, Medical physics.

[2]  A Flynn,et al.  Calibration of a Microselectron HDR iridium 192 source. , 1991, The British journal of radiology.

[3]  J F Williamson,et al.  Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. American Association of Physicists in Medicine. , 1997, Medical physics.

[4]  B J Mijnheer,et al.  Replacement corrections of a Farmer-type ionization chamber for the calibration of Cs-137 and Ir-192 sources in a solid phantom. , 1994, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[5]  L. Dewerd,et al.  Calibration of new high dose rate 192Ir sources. , 2002, Medical physics.

[6]  Yi-Chun Du,et al.  Room scatter factor modelling and measurement error analysis of 192Ir HDR calibration by a Farmer chamber. , 2007, Physics in medicine and biology.

[7]  P. S. Nagarajan,et al.  Monte Carlo aided room scatter studies in the primary air kerma strength standardization of a remote afterloading 192Ir HDR source. , 2001, Physics in medicine and biology.

[8]  M T Nair,et al.  HDR source calibration methods and discrepancies. , 1997, International journal of radiation oncology, biology, physics.

[9]  N P Patel,et al.  Study of scattered radiation for in-air calibration by a multiple-distance method using ionization chambers and an HDR 192Ir brachytherapy source. , 2006, The British journal of radiology.

[10]  Narayan Prasad Patel,et al.  In-air calibration of an HDR 192Ir brachytherapy source using therapy ion chambers. , 2005, Journal of cancer research and therapeutics.

[11]  C Kolotas,et al.  Comparison of calibration procedures for 192Ir high-dose-rate brachytherapy sources. , 1999, International journal of radiation oncology, biology, physics.

[12]  M. L. Randolph,et al.  EFFECT OF FINITE SIZE OF IONIZATION CHAMBERS ON MEASUREMENTS OF SMALL PHOTON SOURCES , 1960 .

[13]  L. Anderson,et al.  Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43 , 1995 .

[14]  K R Das,et al.  A parallel plate chamber for calibration of 192Ir LDR and HDR sources. , 1999, Medical physics.

[15]  J. Williamson,et al.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. , 2004 .