Provocative Testing for the Assessment of the Electromagnetic Interference of RFID and NFC Readers on Implantable Pacemaker

The purpose of this study is to perform provocative testing for the assessment of the electromagnetic interference (EMI) of radiofrequency identification (RFID) and near field communication (NFC) readers on implantable pacemakers (PMs). Ten PMs were exposed to the electromagnetic field generated by an RFID/NFC reader emulator, in the LF (125 kHz), HF and NCF (13.56 MHz), and UHF (900 MHz) bands. Provocative tests were performed by increasing the field strength beyond the maximum levels allowed for commercial devices. PMs were affected by EMI for all three frequency bands. For the LF RFID, 2/10 PMs when exposed to magnetic field levels typical of commercial devices (100 A/m) and 9/10 PMs at a magnetic field of 500 A/m. For the HF RFID and NFC, 3/10 PMs when exposed to output power levels of magnetic field typical of commercial devices (4 W), and 7/10 PMs at 10 W. For the UHF RFID, 2/10 at output power lever of 2 W, and increasing the output power level up to 20 W did not caused EMI in the other PMs. For the PMs evaluated, the provocative tests performed in this study revealed that the safety margin is narrow in the LF and HF/NFC bands, whereas a fair margin exists in the UHF band.

[1]  N AamiPacemakerCommittee-EMCtaskforcePac,et al.  Active implantable medical devices-Electromagnetic compatibility-EMC test protocols for implantable cardiac pacemakers and implantable cardioverter defibrillators , 1999 .

[2]  R. Van der Togt,et al.  Electromagnetic interference from radio frequency identification inducing potentially hazardous incidents in critical care medical equipment. , 2008, JAMA.

[3]  David Parry,et al.  Interference with the operation of medical devices resulting from the use of radio frequency identification technology. , 2009, The New Zealand medical journal.

[5]  B. Christe,et al.  Testing potential interference with RFID usage in the patient care environment. , 2008, Biomedical instrumentation & technology.

[6]  Giovanni Calcagnini,et al.  Radiofrequency identification and medical devices: the regulatory framework on electromagnetic compatibility. Part II: active implantable medical devices , 2012, Expert review of medical devices.

[7]  Toshio Nojima,et al.  In vitro Assessment of Electromagnetic Interference due to Low‐Band RFID reader/writers on Active Implantable Medical Devices , 2009 .

[8]  Seth J. Seidman,et al.  Electromagnetic compatibility of pacemakers and implantable cardiac defibrillators exposed to RFID readers , 2007, Int. J. Radio Freq. Identif. Technol. Appl..

[9]  V Barbaro,et al.  On the mechanisms of interference between mobile phones and pacemakers: parasitic demodulation of GSM signal by the sensing amplifier. , 2003, Physics in medicine and biology.

[10]  Joshua W Guag,et al.  Electromagnetic compatibility of implantable neurostimulators to RFID emitters , 2011, Biomedical engineering online.

[11]  Joshua W. Guag,et al.  In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators. , 2010, Heart rhythm.

[12]  G Calcagnini,et al.  Setups for in vitro assessment of RFID interference on pacemakers , 2013, Physics in medicine and biology.

[13]  Seth J. Seidman,et al.  Design of unique simulators to evaluate medical device susceptibility to radio frequency identification exposure , 2014, IEEE Electromagnetic Compatibility Magazine.