Annular and Cylindrical Phased Array Geometries for Transrectal High‐Intensity Focused Ultrasound (HIFU) using PZT and Piezocomposite Materials

This paper presents engineering progress and the latest in‐vitro and in‐vivo results obtained with a 4.0 MHz, 20 element, PZT annular transrectal HIFU array and several 4.0 MHz, 211 element, PZT and piezocomposite cylindrical transrectal HIFU arrays for the treatment of prostate cancer. The geometries of both arrays were designed and analyzed to steer the HIFU beams to the desired sites in the prostate volume using multi‐channel electronic drivers, with the intent to increase treatment efficiency and reliability for the next generation of HIFU systems. The annular array is able to focus in depth from 25 mm to 50 mm, generate total acoustic powers in excess of 60W, and has been integrated into a modified Sonablate®500 HIFU system capable of controlling such an applicator through custom treatment planning and execution software. Both PZT‐ and piezocomposite cylindrical arrays were constructed and their characteristics were compared for the transrectal applications. These arrays have been installed into appr...

[1]  R. Hawes,et al.  High-intensity focused ultrasound. , 1994, Gastrointestinal endoscopy clinics of North America.

[2]  J. T. Kouzmanoff,et al.  Ultrasound phased arrays for prostate treatment. , 2001, The Journal of the Acoustical Society of America.