Finite-element analysis of temperature rise and lesion formation from catheter ultrasound ablation transducers

A model using finite-element analysis (FEA) has been developed to calculate the temperature rise in tissue from intracardiac ultrasound ablation catheters and to predict if this temperature rise is adequate for producing a lesion in the tissue. In the model, acoustic fields are simulated with Field II, and heat transfer is modeled with an FEA program. To validate the model, we compare its results to experimental results from an integrated, real-time three-dimensional (3-D) ultrasound imaging and ultrasound ablation catheter. The ultrasound ablation transducer is a ring transmitting at 10 MHz capable of producing an acoustic intensity of 16 W/cm/sup 2/. It was used to ablate four lesions in tissue, and temperature rise as a function of time was monitored by embedded thermocouples. The average absolute difference between final temperatures predicted by FEA and those measured is 1.95 /spl plusmn/ 0.72/spl deg/C. Additionally, model and experimental lesion size are in good agreement. The model then is used to design a new ultrasound catheter with a 7.5 MHz linear phased array for ablation. Eight designs are modeled, and acoustic intensity, temperature rise, and ablation ability are compared.

[1]  H. H. Penns Analysis of tissue and arterial blood temperatures in the resting human forearm , 1948 .

[2]  P. Tchou,et al.  First Human Experience With Pulmonary Vein Isolation Using a Through-the-Balloon Circumferential Ultrasound Ablation System for Recurrent Atrial Fibrillation , 2000 .

[3]  S.W. Smith,et al.  Integrated catheter for 3-D intracardiac echocardiography and ultrasound ablation , 2003, IEEE Symposium on Ultrasonics, 2003.

[4]  D. Haines,et al.  Ablation for the treatment of arrhythmias , 1996, Proc. IEEE.

[5]  D. R. Bacon,et al.  Increased heating by diagnostic ultrasound due to nonlinear propagation. , 1990, The Journal of the Acoustical Society of America.

[6]  P D Wolf,et al.  Real-time three-dimensional intracardiac echocardiography. , 2001, Ultrasound in medicine & biology.

[7]  J. Jensen,et al.  Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[8]  David E Haines,et al.  Intracardiac echocardiography-guided, anatomically based radiofrequency ablation of focal atrial fibrillation originating from pulmonary veins. , 2002, Journal of the American College of Cardiology.

[9]  K. Hynynen,et al.  Cylindrical ultrasonic transducers for cardiac catheter ablation , 1997, IEEE Transactions on Biomedical Engineering.

[10]  F. Duck Physical properties of tissue , 1990 .

[11]  William H. Nau,et al.  Novel Catheter Technology for Ablative Cure of Atrial Fibrillation , 2004, Journal of Interventional Cardiac Electrophysiology.

[12]  A W Guy,et al.  Calculation by the method of finite differences of the temperature distribution in layered tissues. , 1973, IEEE transactions on bio-medical engineering.

[13]  Patrick D Wolf,et al.  Dual lumen transducer probes for real-time 3-D interventional cardiac ultrasound. , 2003, Ultrasound in medicine & biology.

[14]  P. Wolf,et al.  A miniaturized catheter 2-D array for real-time, 3-D intracardiac echocardiography , 2004, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[15]  S.W. Smith,et al.  High-speed ultrasound volumetric imaging system. II. Parallel processing and image display , 1991, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[16]  F. Marcus,et al.  The feasibility of using ultrasound for cardiac ablation , 1993, IEEE Transactions on Biomedical Engineering.

[17]  C. Ibukiyama,et al.  Experimental study of catheter ablation using ultrasound energy in canine and porcine hearts. , 1998, Japanese heart journal.

[18]  J. Wu,et al.  Temperature elevation generated by a focused Gaussian beam of ultrasound. , 1990, Ultrasound in medicine & biology.

[19]  P. Sanders,et al.  Phased‐Array Intracardiac Echocardiography to Guide Radiofrequency Ablation in the Left Atrium and at the Pulmonary Vein Ostium , 2001, Journal of cardiovascular electrophysiology.

[20]  K. Nightingale,et al.  On the thermal effects associated with radiation force imaging of soft tissue , 2004, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[21]  W. Nyborg Solutions of the bio-heat transfer equation. , 1988, Physics in medicine and biology.

[22]  S.W. Smith,et al.  High-speed ultrasound volumetric imaging system. I. Transducer design and beam steering , 1991, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[23]  W. Dewey,et al.  Thermal dose determination in cancer therapy. , 1984, International journal of radiation oncology, biology, physics.

[24]  R. Krimholtz,et al.  New equivalent circuits for elementary piezoelectric transducers , 1970 .

[25]  Xiaobing Fan,et al.  Evaluation of accuracy of a theoretical model for predicting the necrosed tissue volume during focused ultrasound surgery , 1995, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[26]  Stephen W. Smith,et al.  Combined 3D intracardiac echo and ultrasound ablation , 2003, SPIE Medical Imaging.

[27]  Patrick D Wolf,et al.  Feasibility Study of Real‐Time Three‐Dimensional Intracardiac Echocardiography for Guidance of Interventional Electrophysiology , 2002, Pacing and clinical electrophysiology : PACE.