Flow-through coaxial sample holder design for dielectric properties measurements from 1 to 350 MHz

A system for measuring the dielectric properties of cereal grains from 1 to 350 MHz with a coaxial sample holder is presented. A signal-flow graph model was used to determine the permittivity of several polar alcohols from the full two-port S-parameter measurements. At the lowest frequencies 1-25 MHz where the phase measurements are less accurate, a lumped-parameter model was used to predict the dielectric loss factor values. The system was calibrated with measurements on air and decanol and verified with measurements on octanol, hexanol, and pentanol. The standard error for the polar alcohols used for verification was 2.3% for the dielectric constant and 7.6% for the dielectric loss factor. Although measurements were taken on static samples, the sample holder is designed to accommodate flowing grain.

[1]  P. Somlo The Computation of Coaxial Line Step Capacitances , 1967 .

[2]  A. A. Maryott,et al.  Circular of the Bureau of Standards no. 589:: tables of dielectric dispersion data for pure liquids and dilute solutions , 1958 .

[3]  R. J. King,et al.  Microwave moisture measurement of grains , 1991 .

[4]  Yves Dutuit,et al.  On a multiple reflection time domain method in dielectric spectroscopy: Application to the study of some normal primary alcohols , 1977 .

[5]  L. Stetson,et al.  Frequency and moisture dependence of the dielectric properties of hard red winter wheat , 1976 .

[6]  A. Kraszewski,et al.  An Improved Microwave Method of Moisture Content Measurement and Control , 1976, IEEE Transactions on Industrial Electronics and Control Instrumentation.

[7]  Stuart O. Nelson,et al.  DENSITY-INDEPENDENT MOISTURE DETERMINATION IN WHEAT BY MICROWAVE MEASUREMENT , 1991 .

[8]  Weng Cho Chew,et al.  Design and calibration of a large broadband dielectric measurement cell , 1991, IEEE Trans. Geosci. Remote. Sens..

[9]  Graham A. Moore,et al.  Continuous Measurement of Grain Moisture Content During Harvest , 1988 .

[10]  S. O. Nelson,et al.  RADIO-FREQUENCY DENSITY-INDEPENDENT MOISTURE DETERMINATION IN WHEAT , 1993 .

[11]  A. Kraszewski,et al.  A Preliminary Study on Microwave Monitoring of Moisture Content in Wheat , 1977 .

[12]  Stuart O. Nelson,et al.  WHEAT MOISTURE DETERMINATION BY 1- TO 110-MHZ SWEPT-FREQUENCY ADMITTANCE MEASUREMENTS , 1998 .

[13]  Andrzej Kraszewski,et al.  Microwave aquametry : electromagnetic wave interaction with water-containing materials , 1996 .