Ocular aberrations with ray tracing and Shack-Hartmann wave-front sensors: does polarization play a role?

Ocular aberrations were measured in 71 eyes by using two reflectometric aberrometers, employing laser ray tracing (LRT) (60 eyes) and a Shack-Hartmann wave-front sensor (S-H) (11 eyes). In both techniques a point source is imaged on the retina (through different pupil positions in the LRT or a single position in the S-H). The aberrations are estimated by measuring the deviations of the retinal spot from the reference as the pupil is sampled (in LRT) or the deviations of a wave front as it emerges from the eye by means of a lenslet array (in the S-H). In this paper we studied the effect of different polarization configurations in the aberration measurements, including linearly polarized light and circularly polarized light in the illuminating channel and sampling light in the crossed or parallel orientations. In addition, completely depolarized light in the imaging channel was obtained from retinal lipofuscin autofluorescence. The intensity distribution of the retinal spots as a function of entry (for LRT) or exit pupil (for S-H) depends on the polarization configuration. These intensity patterns show bright corners and a dark area at the pupil center for crossed polarization, an approximately Gaussian distribution for parallel polarization and a homogeneous distribution for the autofluorescence case. However, the measured aberrations are independent of the polarization states. These results indicate that the differences in retardation across the pupil imposed by corneal birefringence do not produce significant phase delays compared with those produced by aberrations, at least within the accuracy of these techniques. In addition, differences in the recorded aerial images due to changes in polarization do not affect the aberration measurements in these reflectometric aberrometers.

[1]  F. Delori,et al.  Spectral reflectance of the human ocular fundus. , 1989, Applied optics.

[2]  P Artal,et al.  Double-pass imaging polarimetry in the human eye. , 1999, Optics letters.

[3]  P Artal,et al.  Polarization and retinal image quality estimates in the human eye. , 2001, Journal of the Optical Society of America. A, Optics, image science, and vision.

[4]  J. Y. Wang,et al.  Wave-front interpretation with Zernike polynomials. , 1980, Applied optics.

[5]  D R Williams,et al.  Supernormal vision and high-resolution retinal imaging through adaptive optics. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[6]  G. V. van Blokland,et al.  Corneal polarization in the living human eye explained with a biaxial model. , 1987, Journal of the Optical Society of America. A, Optics and image science.

[7]  Stephen A. Burns,et al.  Comparing Laser Ray Tracing, the Spatially Resolved Refractometer, and the Hartmann-Shack Sensor to Measure the Ocular Wave Aberration , 2001, Optometry and vision science : official publication of the American Academy of Optometry.

[8]  I Iglesias,et al.  Closed-loop adaptive optics in the human eye. , 2001, Optics letters.

[9]  W N Charman,et al.  Objective technique for the determination of monochromatic aberrations of the human eye. , 1984, Journal of the Optical Society of America. A, Optics and image science.

[10]  M. Losada,et al.  Aberrations and Relative Efficiency of Light Pencils in the Living Human Eye , 1997, Optometry and vision science : official publication of the American Academy of Optometry.

[11]  S A Burns,et al.  Direct measurement of human-cone-photoreceptor alignment. , 1995, Journal of the Optical Society of America. A, Optics, image science, and vision.

[12]  R. Navarro,et al.  Odd aberrations and double-pass measurements of retinal image quality. , 1995, Journal of the Optical Society of America. A, Optics, image science, and vision.

[13]  Stephen A. Burns,et al.  A new approach to the study of ocular chromatic aberrations , 1999, Vision Research.

[14]  A. Stanworth,et al.  The Polarization Optics of the Isolated Cornea * , 1950, The British journal of ophthalmology.

[15]  S A Burns,et al.  Cone spacing and waveguide properties from cone directionality measurements. , 1999, Journal of the Optical Society of America. A, Optics, image science, and vision.

[16]  R Navarro,et al.  Laser ray-tracing method for optical testing. , 1999, Optics letters.

[17]  Theo Seiler,et al.  Meßplatz zur Bestimmung der monochromatischen Aberration des menschlichen Auges , 1997, Der Ophthalmologe.

[18]  B. Hochheimer,et al.  Retinal polarization effects. , 1982, Applied optics.

[19]  F. James Rohlf,et al.  Biometry: The Principles and Practice of Statistics in Biological Research , 1969 .

[20]  A. Bradley,et al.  The chromatic eye: a new reduced-eye model of ocular chromatic aberration in humans. , 1992, Applied optics.

[21]  Larry N. Thibos,et al.  Standards for reporting the optical aberrations of eyes , 2000 .

[22]  K. Meek,et al.  Organization of collagen fibrils in the corneal stroma in relation to mechanical properties and surgical practice. , 1999, Journal of refractive surgery.

[23]  A Bradley,et al.  Comparison of the eye's wave-front aberration measured psychophysically and with the Shack-Hartmann wave-front sensor. , 1998, Journal of the Optical Society of America. A, Optics, image science, and vision.

[24]  M L Wolbarsht,et al.  The corneal polarization cross. , 1978, Journal of the Optical Society of America.

[25]  [Measuring device for determining monochromatic aberration of the human eye]. , 1997, Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft.

[26]  R. Navarro,et al.  Laser Ray Tracing versus Hartmann-Shack sensor for measuring optical aberrations in the human eye. , 2000, Journal of the Optical Society of America. A, Optics, image science, and vision.

[27]  Junzhong Liang,et al.  Aberrations and retinal image quality of the normal human eye. , 1997, Journal of the Optical Society of America. A, Optics, image science, and vision.

[28]  Wavefront sensing in the human eye , 2002 .

[29]  D. Van Norren,et al.  Intensity and polarization of light scattered at small angles from the human fovea , 1986, Vision Research.

[30]  R Navarro,et al.  Ocular aberrations before and after myopic corneal refractive surgery: LASIK-induced changes measured with laser ray tracing. , 2001, Investigative ophthalmology & visual science.

[31]  C K Dorey,et al.  In vivo fluorescence of the ocular fundus exhibits retinal pigment epithelium lipofuscin characteristics. , 1995, Investigative ophthalmology & visual science.

[32]  J. Schwiegerling,et al.  Customized corneal ablation and super vision. , 2000, Journal of refractive surgery.

[33]  Stephen A Burns,et al.  Effect of the polarization on ocular wave aberration measurements. , 2002, Journal of the Optical Society of America. A, Optics, image science, and vision.

[34]  Junzhong Liang,et al.  Objective measurement of wave aberrations of the human eye with the use of a Hartmann-Shack wave-front sensor. , 1994, Journal of the Optical Society of America. A, Optics, image science, and vision.

[35]  Y Fainman,et al.  Adaptive control of a micromachined continuous-membrane deformable mirror for aberration compensation. , 1999, Applied optics.

[36]  W. Charman,et al.  Objective measurements of the longitudinal chromatic aberration of the human eye , 1976, Vision Research.

[37]  J. Gorrand,et al.  Diffusion of the human retina and quality of the optics of the eye on the fovea and the peripheral retina , 1979, Vision Research.

[38]  Norberto López-Gil,et al.  Measurement of the eye’s near infrared wave-front aberration using the objective crossed-cylinder aberroscope technique , 1999, Vision Research.

[39]  J C Dainty,et al.  Single-pass measurements of the wave-front aberrations of the human eye by use of retinal lipofuscin autofluorescence. , 1999, Optics letters.

[40]  M A Muriel,et al.  Real-time Fourier transformer based on fiber gratings. , 1999, Optics letters.

[41]  R. Alfieri,et al.  Diffusion of the retinal layers of the living human eye , 1984, Vision Research.

[42]  R Birngruber,et al.  Influence of temperature and time on thermally induced forces in corneal collagen and the effect on laser thermokeratoplasty. , 2000, Journal of cataract and refractive surgery.

[43]  R Navarro,et al.  Phase plates for wave-aberration compensation in the human eye. , 2000, Optics letters.

[44]  B. Singer,et al.  Improvement in retinal image quality with dynamic correction of the eye's aberrations. , 2001, Optics express.

[45]  L. Thibos,et al.  Standards for reporting the optical aberrations of eyes. , 2002, Journal of refractive surgery.

[46]  S A Burns,et al.  Measurement of the wave-front aberration of the eye by a fast psychophysical procedure. , 1998, Journal of the Optical Society of America. A, Optics, image science, and vision.