A novel flex-rigid and soft-release ECoG array

This article addresses a novel fabrication process for an electrocorticogram (ECoG) electrode array. It consists of three regions: a flexible recording area, a flexible cable, and a rigid field for soldering the connectors. The flexible components can adapt to the curved shape of the cerebral cortex. Furthermore, the entire structure is a free-standing membrane, attached by removable polyimide straps to its carrier substrate. This configuration allows for a high level of control during soldering, electrode characterization and sterilization, as well as a soft release of the array off its carrier just before implantation. The array contains 128 gold electrodes, each 300 nm thick, sandwiched between two 5 μm thick polyimide films. The measuring area of the device is a regular hexagon with a side length of 7.2 mm, designed for implantation on the primary visual cortex of a Rhesus monkey. The flexible cable is 4 cm long. The rigid soldering area was designed for 4×32 OMNETICS connectors. The line resistance from an electrode site to the corresponding electrical connector pin is 540 Ω.

[1]  José del R. Millán,et al.  Non-Invasive Brain-Actuated Control of a Mobile Robot , 2003, IJCAI.

[2]  J. Carmena,et al.  Brain-machine interfaces and transcranial stimulation: future implications for directing functional movement and improving function after spinal injury in humans. , 2012, Handbook of clinical neurology.

[3]  T. Stieglitz,et al.  Polymers for neural implants , 2011 .

[4]  R. Fisher,et al.  Brain stimulation for epilepsy , 2005, Nature Clinical Practice Neurology.

[5]  Jonathan R Wolpaw,et al.  Control of a two-dimensional movement signal by a noninvasive brain-computer interface in humans. , 2004, Proceedings of the National Academy of Sciences of the United States of America.

[6]  Patrick Ruther,et al.  A Novel Assembly Method for Silicon-Based Neural Devices , 2009 .

[7]  Jon A. Mukand,et al.  Neuronal ensemble control of prosthetic devices by a human with tetraplegia , 2006, Nature.

[8]  W. H. Dobelle Artificial vision for the blind by connecting a television camera to the visual cortex. , 2000, ASAIO journal.

[9]  O. Paul,et al.  Microprobe Array with Low Impedance Electrodes and Highly Flexible Polyimide Cables for Acute Neural Recording , 2007, 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[10]  E. Fetz,et al.  Direct control of paralyzed muscles by cortical neurons , 2008, Nature.

[11]  R. Oostenveld,et al.  A MEMS-based flexible multichannel ECoG-electrode array , 2009, Journal of neural engineering.

[12]  M. Carandini,et al.  Local Origin of Field Potentials in Visual Cortex , 2009, Neuron.

[13]  Gerwin Schalk,et al.  A brain–computer interface using electrocorticographic signals in humans , 2004, Journal of neural engineering.

[14]  Justin A. Blanco,et al.  Dissolvable films of silk fibroin for ultrathin conformal bio-integrated electronics. , 2010, Nature materials.

[15]  Takafumi Suzuki,et al.  Simultaneous recording of ECoG and intracortical neuronal activity using a flexible multichannel electrode-mesh in visual cortex , 2011, NeuroImage.

[16]  A. Schulze-Bonhage,et al.  First long term in vivo study on subdurally implanted Micro-ECoG electrodes, manufactured with a novel laser technology , 2011, Biomedical microdevices.