Development of a cuffless blood pressure measurement system

This study constructs a novel blood pressure measurement device without the air cuff to overcome the problem of discomfort and portability. The proposed device measures the blood pressure through a mechanism that is made of silicon rubber and pressure transducer. The system uses a microcontroller to control the measurement procedure and to perform the necessary computation. To verify the feasibility of the constructed device, ten young volunteers were recruited. Ten blood pressure readings were obtained using the new system and were compared with ten blood pressure readings from bedside monitor (Spacelabs Medical, model 90367). The results indicated that, when all the readings were included, the mean pressure, systolic pressure and diastolic pressure from the new system were all higher than those from bedside monitor. The correlation coefficients between these two were 0.15, 0.18 and 0.29, for mean, systolic and diastolic pressures, respectively. After excluding irregular apparatus utilization, the correlation coefficient increased to 0.71, 0.60 and 0.41 for diastolic pressure, mean pressure and systolic pressure, respectively. We can conclude from these results that the accuracy can be improved effectively by defining the user regulation more precisely. The above mentioned irregular apparatus utilization factors can be identified and eliminated by the microprocessor to provide a reliable blood pressure measurement in practical applications in the future.