Dose calibration of Gafchromic EBT3 film for Ir-192 brachytherapy source using 3D-printed PLA and ABS plastics

Abstract3D printing technology has allowed the creation of custom applicators for high dose rate (HDR) brachytherapy, especially for complex anatomy. With conformal therapy comes the need for advanced dosimetric verification. It is important to demonstrate how dose to 3D printed materials can be related to dose to water. This study aimed to determine dose differences and uncertainties using 3D printed PLA and ABS plastics for Radiochromic film calibration in HDR brachytherapy.Gafchromic EBT3 film pieces were irradiated in water with an Ir-192 source at calculated dose levels ranging from 0 to 800 cGy, to create the control calibration curve. Similarly, film was placed below 3D printed PLA and ABS blocks and irradiated at the same dose levels calculated for water, ranging from 0 to 800 cGy. After a 72-h development time, film pieces were scanned on a flatbed scanner and the median pixel value was recorded in the region of highest dose. This value was converted to net optical density (NOD). A rational function was used to fit a calibration curve in water that relates NOD to dose for red, green, and blue color channels. Based on this fitted curve, ABS and PLA NOD values were used to estimate dose in 3D printed plastics.From the fitted calibration curve, mean residual error between measured and planned dose to water was less than 1% for each color channel at high dose levels. At high dose levels, ABS and PLA mean residual errors were about 6.9 and 7.8% in the red channel, while 5.2 and 5.7% in the green channel. Combined uncertainties measured to be about 6.9% at high dose levels. This study demonstrated dose differences and uncertainties using 3D printed applicators for HDR Ir-192 brachytherapy.

[1]  Tom Burrows,et al.  Introduction of novel 3D-printed superficial applicators for high-dose-rate skin brachytherapy. , 2017, Brachytherapy.

[2]  Zoubir Ouhib,et al.  Aspects of dosimetry and clinical practice of skin brachytherapy: The American Brachytherapy Society working group report. , 2015, Brachytherapy.

[3]  N. Ayoobian,et al.  Gafchromic film dosimetry of a new HDR  192Ir brachytherapy source , 2016, Journal of applied clinical medical physics.

[4]  Peter Balter,et al.  Material matters: Analysis of density uncertainty in 3D printing and its consequences for radiation oncology , 2018, Medical physics.

[5]  Mark J Rivard,et al.  Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources. , 2006, Medical physics.

[6]  Jose Perez-Calatayud,et al.  GEC-ESTRO ACROP recommendations in skin brachytherapy. , 2018, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[7]  J J Wilkens,et al.  Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. , 2012, Medical physics.

[8]  W. Butler,et al.  Erratum: Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations (Medical Physics (2004) 31 (633-674)) , 2004 .

[9]  D Granero,et al.  Phantom size in brachytherapy source dosimetric studies. , 2004, Medical physics.

[10]  J. Galvin,et al.  Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine. , 1998, Medical physics.

[11]  L. Anderson,et al.  Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43 , 1995 .

[12]  J. Williamson,et al.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. , 2004 .

[13]  A. Micke,et al.  Multichannel film dosimetry with nonuniformity correction. , 2011, Medical physics.

[14]  S. Chiu‐Tsao,et al.  Energy Dependence of the New Gafchromic EBT3 Film: Dose Response Curves for 50 kV, 6 and 15 MV X-Ray Beams * , 2012 .

[15]  J. Williamson,et al.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. , 2003, Medical physics.

[16]  Santi Tofani,et al.  Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification , 2013, Journal of applied clinical medical physics.

[17]  O. A. García-Garduño,et al.  Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density , 2016, Journal of applied clinical medical physics.

[18]  O. A. García-Garduño,et al.  Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density. , 2016, Journal of Applied Clinical Medical Physics.

[19]  M Mathot,et al.  Gafchromic film dosimetry: four years experience using FilmQA Pro software and Epson flatbed scanners. , 2014, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[20]  Technical Note: Impact of region of interest size and location in Gafchromic film dosimetry. , 2018, Medical physics.

[21]  I. Kawrakow,et al.  On the characterization and uncertainty analysis of radiochromic film dosimetry. , 2009, Medical physics.