Density Independent Moisture Meter at X-Band

A new method of density independent moisture determination at one single frequency is developed. It is based on the two-parameter-measurement of the complex dielectric constant being composed to a density independent calibration curve A(¿) which is a function of the moisture content ¿ only. The feasibility of this principle is demonstrated by an industrial moisture meter at 9 GHz for different materials and sensing heads.

[1]  A. Kraszewski,et al.  An Improved Microwave Method of Moisture Content Measurement and Control , 1976, IEEE Transactions on Industrial Electronics and Control Instrumentation.

[2]  B. Schiek,et al.  Microwave Systems for Industrial Measurements , 1981 .