Developing and characterizing MR/CT‐visible materials used in QA phantoms for MRgRT systems

PURPOSE Synthetic tissue equivalent (STE) materials currently used to simulate tumor and surrounding tissues for IROC-Houston's anthropomorphic head and thorax QA phantoms cannot be visualized using magnetic resonance (MR) imaging. The purpose of this study was to characterize dual MR/CT-visible STE materials that can be used in an end-to-end QA phantom for MR-guided radiotherapy (MRgRT) modalities. METHODS Over 80 materials' MR, CT, and dosimetric STE properties were investigated for use in MRgRT QA phantoms. The materials tested included homogeneous and heterogeneous materials to simulate soft tissue/tumor and lung tissues. Materials were scanned on a Siemens' Magnetom Espree 1.5 T using four sequences, which showed the materials visual contrast between T1- and T2-weighted images. Each material's Hounsfield number and electron density data was collected using a GE's CT Lightspeed Simulator. Dosimetric properties were examined by constructing a 10 × 10 × 20 cm3 phantom of the selected STE materials that was divided into three sections: anterior, middle, and posterior. Anterior and posterior pieces were composed of polystyrene, whereas the middle section was substituted with the selected STE materials. EBT3 film was inserted into the phantom's midline and was irradiated using an Elekta's Versa 6 MV beam with a prescription of 6 Gy at 1.5 cm and varying field size of: 10 × 10 cm2 , 6 × 6 cm2 , and 3 × 3 cm2 . Measured film PDD curves were compared to planning system calculations and conventional STE materials' percent depth dose (PDD) curves. RESULTS The majority of the tested materials showed comparable CT attenuation properties to their respective organ site; however, most of the tested materials were not visible on either T1- or T2-weighted MR images. Silicone, hydrocarbon, synthetic gelatin, and liquid PVC plastic-based materials showed good MR image contrast. In-house lung equivalent materials made with either silicone- or hydrocarbon-based materials had HUs ranging from: -978 to -117 and -667 to -593, respectively. Synthetic gelatin and PVC plastic-based materials resembled soft tissue/tumor equivalent materials and had HUs of: -175 to -170 and -29 to 32, respectively. PDD curves of the selected MR/CT-visible materials were comparable to IROC-Houston's conventional phantom STE materials. The smallest field size showed the largest disagreements, where the average discrepancies between calculated and measured PDD curves were 1.8% and 5.9% for homogeneous and heterogeneous testing materials, respectively. CONCLUSIONS Gelatin, liquid plastic, and hydrocarbon-based materials were determined as alternative STE substitutes for MRgRT QA phantoms.

[1]  Ananth Ravi,et al.  Technical Note: Multipurpose CT, ultrasound, and MRI breast phantom for use in radiotherapy and minimally invasive interventions. , 2016, Medical physics.

[2]  M. Moerland,et al.  Magnetic resonance imaging acquisition techniques for radiotherapy planning. , 2014, Seminars in radiation oncology.

[3]  Steffen Greilich,et al.  An anthropomorphic multimodality (CT/MRI) head phantom prototype for end-to-end tests in ion radiotherapy. , 2015, Zeitschrift fur medizinische Physik.

[4]  Jay Singh,et al.  Measurement and Analysis of Physical and Climatic Distribution Environment for Air Package Shipment , 2015 .

[5]  Steffen Greilich,et al.  Technical Note: Radiological properties of tissue surrogates used in a multimodality deformable pelvic phantom for MR-guided radiotherapy. , 2016, Medical physics.

[6]  Roberto Orecchia,et al.  Current concepts on imaging in radiotherapy , 2008, European Journal of Nuclear Medicine and Molecular Imaging.

[7]  Jan J W Lagendijk,et al.  The magnetic resonance imaging-linac system. , 2014, Seminars in radiation oncology.

[8]  X Allen Li,et al.  MRI-based IMRT planning for MR-linac: comparison between CT- and MRI-based plans for pancreatic and prostate cancers , 2016, Physics in medicine and biology.

[9]  Kristy K Brock,et al.  Point: Principles of magnetic resonance imaging integration in a computed tomography-based radiotherapy workflow. , 2014, Seminars in radiation oncology.

[10]  Alberto Torresin,et al.  Review of potential improvements using MRI in the radiotherapy workflow. , 2015, Zeitschrift fur medizinische Physik.

[11]  J. Jonsson,et al.  Counterpoint: Opportunities and challenges of a magnetic resonance imaging-only radiotherapy work flow. , 2014, Seminars in radiation oncology.

[12]  B. Erickson,et al.  Comprehensive MRI simulation methodology using a dedicated MRI scanner in radiation oncology for external beam radiation treatment planning. , 2014, Medical physics.

[13]  B. Fallone,et al.  The rotating biplanar linac-magnetic resonance imaging system. , 2014, Seminars in radiation oncology.

[14]  Sasa Mutic,et al.  The ViewRay system: magnetic resonance-guided and controlled radiotherapy. , 2014, Seminars in radiation oncology.

[15]  David A Jaffray,et al.  A facility for magnetic resonance-guided radiation therapy. , 2014, Seminars in radiation oncology.

[16]  David S Followill,et al.  Design, development, and implementation of the radiological physics center's pelvis and thorax anthropomorphic quality assurance phantoms. , 2007, Medical physics.

[17]  Avraham Eisbruch,et al.  Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the Radiation Therapy Oncology Group. , 2005, International journal of radiation oncology, biology, physics.

[18]  B W Raaymakers,et al.  Compensating for the impact of non-stationary spherical air cavities on IMRT dose delivery in transverse magnetic fields , 2015, Physics in medicine and biology.

[19]  Aitang Xing,et al.  Variable angle stereo imaging for rapid patient position correction in an in-house real-time position monitoring system. , 2017, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.