Evaluation of two methods for continuous cardiac output assessment during exercise in chronic heart failure patients.

The purpose of this study was to evaluate the accuracy of two techniques for the continuous assessment of cardiac output in patients with chronic heart failure (CHF): a radial artery pulse contour analysis method that uses an indicator dilution method for calibration (LiDCO) and an impedance cardiography technique (Physioflow), using the Fick method as a reference. Ten male CHF patients (New York Heart Association class II-III) were included. At rest, cardiac output values obtained by LiDCO and Physioflow were compared with those of the direct Fick method. During exercise, the continuous Fick method was used as a reference. Exercise, performed on a cycle ergometer in upright position, consisted of two constant-load tests at 30% and 80% of the ventilatory threshold and a symptom-limited maximal test. Both at rest and during exercise LiDCO showed good agreement with reference values [bias +/- limits of agreement (LOA), -1% +/- 28% and 2% +/- 28%, respectively]. In contrast, Physioflow overestimated reference values both at rest and during exercise (bias +/- LOA, 48% +/- 60% and 48% +/- 52%, respectively). Exercise-related within-patient changes of cardiac output, expressed as a percent change, showed for both techniques clinically acceptable agreement with reference values (bias +/- LOA: 2% +/- 26% for LiDCO, and -2% +/- 36% for Physioflow, respectively). In conclusion, although the limits of agreement with the Fick method are pretty broad, LiDCO provides accurate measurements of cardiac output during rest and exercise in CHF patients. Although Physioflow overestimates cardiac output, this method may still be useful to estimate relative changes during exercise.

[1]  Z. Peng,et al.  An investigation to show the effect of lung fluid on impedance cardiac output in the anaesthetized dog. , 2005, British journal of anaesthesia.

[2]  M. Jessen,et al.  PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. , 2002, The Annals of thoracic surgery.

[3]  Precisión y exactitud de un oxímetro de pulso "de bolsillo" en la ciudad de México , 2006 .

[4]  T. Aoyagi,et al.  Kinetics of oxygen uptake at onset of exercise related to cardiac output, but not to arteriovenous oxygen difference in patients with chronic heart failure. , 1999, The American journal of cardiology.

[5]  F. Marumo,et al.  Beat-to-beat evaluation of cardiac function during recovery from upright bicycle exercise in patients with coronary artery disease. , 1990, American heart journal.

[6]  D. Webb,et al.  Non-invasive measurement of cardiac output in patients with chronic heart failure , 2004, Blood pressure monitoring.

[7]  D P Bernstein,et al.  A new stroke volume equation for thoracic electrical bioimpedance: Theory and rationale , 1986, Critical care medicine.

[8]  Arthur Selzer,et al.  CIRCULATORY PHYSIOLOGY: Cardiac Output and Its Regulation , 1964 .

[9]  Luigi Tavazzi,et al.  Recommendations for exercise testing in chronic heart failure patients. , 2001, European heart journal.

[10]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[11]  A Noordergraaf,et al.  Estimation of total systemic arterial compliance in humans. , 1990, Journal of applied physiology.

[12]  B. Whipp,et al.  A new method for detecting anaerobic threshold by gas exchange. , 1986, Journal of applied physiology.

[13]  A. Charloux,et al.  Does thoracic bioimpedance accurately determine cardiac output in COPD patients during maximal or intermittent exercise? , 2005, Chest.

[14]  G. Ferretti,et al.  Correction of cardiac output obtained by Modelflow from finger pulse pressure profiles with a respiratory method in humans. , 2004, Clinical science.

[15]  K. Ikeda,et al.  Comparison of the accuracy of the lithium dilution technique with the thermodilution technique for measurement of cardiac output. , 1997, British journal of anaesthesia.

[16]  M. Jonas,et al.  Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. , 2002, Current opinion in critical care.

[17]  H. Kemps,et al.  The reliability of continuous measurement of mixed venous oxygen saturation during exercise in patients with chronic heart failure , 2008, European Journal of Applied Physiology.

[18]  Martin Buchheit,et al.  Non-invasive cardiac output evaluation during a maximal progressive exercise test, using a new impedance cardiograph device , 2001, European Journal of Applied Physiology.

[19]  R N Pierson,et al.  Hemodynamic exercise testing. A valuable tool in the selection of cardiac transplantation candidates. , 1996, Circulation.

[20]  D. S. Rustad Precision and Accuracy (3) , 1998 .

[21]  L. Critchley,et al.  A Meta-Analysis of Studies Using Bias and Precision Statistics to Compare Cardiac Output Measurement Techniques , 1999, Journal of Clinical Monitoring and Computing.

[22]  Jean Lonsdorfer,et al.  A new impedance cardiograph device for the non-invasive evaluation of cardiac output at rest and during exercise: comparison with the “direct” Fick method , 2000, European Journal of Applied Physiology.

[23]  R. Leach,et al.  Lithium dilution cardiac output measurement: a comparison with thermodilution. , 1997, Critical care medicine.

[24]  N. Linton,et al.  Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb. , 2001, British journal of anaesthesia.

[25]  C. Lang,et al.  Ease of noninvasive measurement of cardiac output coupled with peak VO2 determination at rest and during exercise in patients with heart failure. , 2007, The American journal of cardiology.

[26]  H. Yamabe,et al.  Clinical application of cardiac output during ramp exercise calculated using the Fick equation--comparison with the 2-stage bicycle ergometer exercise protocol in the supine position. , 1997, Japanese circulation journal.

[27]  K. Beck,et al.  Use of acetylene breathing to determine cardiac output in young and older adults. , 2003, Medicine and science in sports and exercise.

[28]  N. Linton,et al.  Is comparison of changes in cardiac output, assessed by different methods, better than only comparing cardiac output to the reference method? , 2002, British journal of anaesthesia.

[29]  K. Teo,et al.  Reliability and Validity of Measures of Cardiac Output During Incremental to Maximal Aerobic Exercise , 1999, Sports medicine.

[30]  P. Ponikowski,et al.  Origin of oscillatory kinetics of respiratory gas exchange in chronic heart failure. , 1999, Circulation.

[31]  C. Young,et al.  Lithium dilution cardiac output measurement: A clinical assessment of central venous and peripheral venous indicator injection* , 2002, Critical care medicine.

[32]  A. Nogami,et al.  Oxygen uptake kinetics are determined by cardiac function at onset of exercise rather than peak exercise in patients with prior myocardial infarction. , 1994, Circulation.

[33]  P. Jebson,et al.  Continuous Fick cardiac output compared to continuous pulmonary artery electromagnetic flow measurement in pigs. , 1987, Anesthesiology.

[34]  K. Wasserman,et al.  Noninvasive measurement of cardiac output during exercise by inert gas rebreathing technique: a new tool for heart failure evaluation. , 2005, Journal of the American College of Cardiology.

[35]  Pedro Amorim,et al.  Online outlier detection and removal from ECG heart rate and invasive arterial pressure: A-143 , 2006 .

[36]  R. Belardinelli,et al.  Comparison of impedance cardiography with thermodilution and direct Fick methods for noninvasive measurement of stroke volume and cardiac output during incremental exercise in patients with ischemic cardiomyopathy. , 1996, The American journal of cardiology.

[37]  W. Laskey,et al.  Characteristics of vascular hydraulic load in patients with heart failure. , 1985, Circulation.

[38]  P. Poole‐Wilson,et al.  Measurement of cardiac output during exercise by the thermodilution and direct Fick techniques in patients with chronic congestive heart failure. , 1985, The American journal of cardiology.

[39]  S. Rubin,et al.  Accuracy of cardiac output, oxygen uptake, and arteriovenous oxygen difference at rest, during exercise, and after vasodilator therapy in patients with severe, chronic heart failure. , 1982, The American journal of cardiology.

[40]  R. Jansen,et al.  Finapres arterial pulse wave analysis with Modelflow is not a reliable non-invasive method for assessment of cardiac output. , 2002, Clinical science.

[41]  C. K. Mahutte,et al.  Effect of measurement errors on cardiac output calculated with O2 and modified CO2 Fick methods , 1995, Journal of Clinical Monitoring.

[42]  T. Rowland,et al.  Doppler Echocardiography for the Estimation of Cardiac Output with Exercise , 2002, Sports medicine.