Achromatically filtered diamond photoconductive detectors for high power soft x-ray flux measurements

A 1 mm square diamond photoconductive detector (PCD) has been installed on the LLNL Nova laser system, for use as a broad band soft x-ray power diagnostic. The PCD is installed behind an array of pinholes, which cast multiple, overlapping images of the source onto the diamond. This allows reduction of the x-ray intensity, to avoid saturation problems, while avoiding the spectral dependency of thin film filters. The diode current is read out on a 5 GHz bandwidth scope. The system is calibrated by comparison to an absolutely calibrated array of filtered vacuum x-ray photodiodes (XRD` s) (``dante``). The time response of the PCD and its bias electronics have been characterized using the 5th harmonic (210 nm) of a short pulse (< 1 ps) Ti:sapphire laser. The data show a fast rise, limited by the 5 GHz scope bandwidth, and a slower fall off, characterized by an RC time of order 200 ps.