Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial.

OBJECTIVES The purpose of this study was to evaluate the safety and effectiveness of pacemaker lead extraction with the excimer sheath in comparison to nonlaser lead extraction. BACKGROUND Fibrotic attachments that develop between chronically implanted pacemaker leads and to the venous, valvular and cardiac structures are the major obstacles to safe and consistent lead extraction. Locking stylets and telescoping sheaths produce a technically demanding but effective technique of mechanically disrupting the fibrosis. However, ultraviolet excimer laser light dissolves instead of tearing the tissue attachments. METHODS A randomized trial of lead extraction was conducted in 301 patients with 465 chronically implanted pacemaker leads. The laser group patients had the leads removed with identical tools as the nonlaser group with the exception that the inner telescoping sheath was replaced with the 12-F excimer laser sheath. Success for both groups was defined as complete lead removal with the randomized therapy without complications. RESULTS Complete lead removal rate was 94% in the laser group and 64% in the nonlaser group (p = 0.001). Failed nonlaser extraction was completed with the laser tools 88% of the time. The mean time to achieve a successful lead extraction was significantly reduced for patients randomized to the laser tools, 10.1 +/- 11.5 min compared with 12.9 +/- 19.2 min for patients randomized to nonlaser techniques (p < 0.04). Potentially life-threatening complications occurred in none of the nonlaser and three of the laser patients, including one death (p = NS). CONCLUSIONS Laser-assisted pacemaker lead extraction has significant clinical advantages over extraction without laser tools and is associated with significant risks.

[1]  A. Bernstein,et al.  Extraction of implanted transvenous pacing leads: a review of a persistent clinical problem. , 1991, American heart journal.

[2]  C. Byrd,et al.  Lead extraction. Indications and techniques. , 1992, Cardiology clinics.

[3]  C. Reiser,et al.  Large laser sheaths for pacing and defibrillator lead removal , 1998, Lasers in surgery and medicine.

[4]  David W. Hosmer,et al.  Applied Logistic Regression , 1991 .

[5]  T. D. Sellers,et al.  Five‐Years Experience with Intravascular Lead Extraction , 1994 .

[6]  J. Anderson,et al.  The encapsulation of polyurethane-insulated transvenous cardiac pacemaker leads. , 1995, Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology.

[7]  M. V. van Gemert,et al.  Effect of force on ablation depth for a XeCl excimer laser beam delivered by an optical fiber in contact with arterial tissue under saline , 1992, Lasers in surgery and medicine.

[8]  S L Jacques,et al.  XeCl laser ablation of atherosclerotic aorta: Optical properties and energy pathways , 1992, Lasers in surgery and medicine.

[9]  D. Holmes,et al.  Extraction of the Telectronics Accufix 330-801 atrial lead: the Mayo Clinic experience. , 1996, Mayo Clinic proceedings.

[10]  Bittl Ja Clinical results with excimer laser coronary angioplasty. , 1996 .

[11]  T. V. van Leeuwen,et al.  Fundamental laser-tissue interactions. , 1996, Seminars in interventional cardiology : SIIC.

[12]  A. Welch,et al.  Excimer laser induced bubble: Dimensions, theory, and implications for laser angioplasty , 1996, Lasers in surgery and medicine.

[13]  Christopher Reiser,et al.  Excimer laser catheter for pacing lead removal , 1997, Photonics West - Biomedical Optics.

[14]  I. R. Zucker,et al.  The fate of permanent intracardiac electrodes. , 1966, The Journal of surgical research.