Micro MRI of the mouse brain using a novel 400 MHz cryogenic quadrature RF probe

The increasing number of mouse models of human disease used in biomedical research applications has led to an enhanced interest in non‐invasive imaging of mice, e.g. using MRI for phenotyping. However, MRI of small rodents puts high demands on the sensitivity of data acquisition. This requirement can be addressed by using cryogenic radio‐frequency (RF) detection devices. The aim of this work was to investigate the in vivo performance of a 400 MHz cryogenic transmit/receive RF probe (CryoProbe) designed for MRI of the mouse brain. To characterize this novel probe, MR data sets were acquired with both the CryoProbe and a matched conventional receive‐only surface coil operating at room temperature (RT) using conventional acquisition protocols (gradient and spin echo) with identical parameter settings. Quantitative comparisons in phantom and in vivo experiments revealed gains in the signal‐to‐noise ratio (SNR) of 2.4 and 2.5, respectively. The increased sensitivity of the CryoProbe was invested to enhance the image quality of high resolution structural images acquired in scan times compatible with routine operation (< 45 min). In high resolution (30 × 30 × 300 µm3) structural images of the mouse cerebellum, anatomical details such as Purkinje cell and molecular layers could be identified. Similarly, isotropic (60 × 60 × 60 µm3) imaging of mouse cortical and subcortical areas revealed anatomical structures smaller than 100 µm. Finally, 3D MR angiography (52 × 80 × 80 µm3) of the brain vasculature enabled the detailed reconstruction of intracranial vessels (anterior and middle cerebral artery). In conclusion, this low temperature detection device represents an attractive option to increase the performance of small animal MR systems operating at 9.4 Tesla. Copyright © 2009 John Wiley & Sons, Ltd.

[1]  Daniel Marek,et al.  Performance of a 200‐MHz cryogenic RF probe designed for MRI and MRS of the murine brain , 2008, Magnetic resonance in medicine.

[2]  A. S. Hall,et al.  Use of high temperature superconductor in a receiver coil for magnetic resonance imaging , 1991, Magnetic resonance in medicine.

[3]  M. Rudin,et al.  Quantitative and qualitative assessment of articular cartilage in the goat knee with magnetization transfer imaging. , 2001, Magnetic resonance imaging.

[4]  J Hennig,et al.  RARE imaging: A fast imaging method for clinical MR , 1986, Magnetic resonance in medicine.

[5]  L Darrasse,et al.  Perspectives with cryogenic RF probes in biomedical MRI. , 2003, Biochimie.

[6]  N. Radzwill,et al.  Benefits of cryogenic coils for routine in-vivo mouse brain imaging and spectroscopy at 9.4 T , 2007 .

[7]  J. Pauly,et al.  MR angiography by selective inversion recovery , 1987, Magnetic resonance in medicine.

[8]  M. C. Cheng,et al.  Performance of large-size Superconducting coil in 0.21T MRI system , 2004, IEEE Transactions on Biomedical Engineering.

[9]  O. Girard,et al.  Technical aspects: development, manufacture and installation of a cryo-cooled HTS coil system for high-resolution in-vivo imaging of the mouse at 1.5 T. , 2007, Methods.

[10]  Jean-Christophe Ginefri,et al.  Performance of a miniature high‐temperature superconducting (HTS) surface coil for in vivo microimaging of the mouse in a standard 1.5T clinical whole‐body scanner , 2008, Magnetic resonance in medicine.

[11]  A. Haase,et al.  Rapid NMR imaging of dynamic processes using the FLASII technique , 1986, Magnetic resonance in medicine.

[12]  A Mogro-Campero,et al.  A high-temperature superconducting receiver for nuclear magnetic resonance microscopy. , 1993, Science.

[13]  F W Wehrli,et al.  In vivo MR micro imaging with conventional radiofrequency coils cooled to 77 degrees K. , 2000, Magnetic resonance in medicine.

[14]  Nuclear magnetic resonance microscopy , 1999 .

[15]  Michael Garwood,et al.  In vivo visualization of Alzheimer's amyloid plaques by magnetic resonance imaging in transgenic mice without a contrast agent , 2004, Magnetic resonance in medicine.

[16]  G. Johnson,et al.  High-resolution magnetic resonance histology of the embryonic and neonatal mouse: A 4D atlas and morphologic database , 2008, Proceedings of the National Academy of Sciences.

[17]  George Paxinos,et al.  The Mouse Brain in Stereotaxic Coordinates , 2001 .

[18]  Stephen Pickup,et al.  Enhanced delineation of white matter structures of the fixed mouse brain using Gd‐DTPA in microscopic MRI , 2009, NMR in biomedicine.

[19]  J R MacFall,et al.  Performance of a high‐temperature superconducting probe for in vivo microscopy at 2.0 T , 1999, Magnetic resonance in medicine.

[20]  R. Friede The relationship of body size, nerve cell size, axon length, and glial density in the cerebellum. , 1963, Proceedings of the National Academy of Sciences of the United States of America.

[21]  J. Frahm,et al.  Magnetization transfer MRI of mouse brain reveals areas of high neural density. , 2003, Magnetic resonance imaging.

[22]  G A Johnson,et al.  A high‐temperature superconducting Helmholtz probe for microscopy at 9.4 T , 1999, Magnetic resonance in medicine.

[23]  F David Doty,et al.  Radio frequency coil technology for small‐animal MRI , 2007, NMR in biomedicine.

[24]  Anders M. Dale,et al.  Automated segmentation of the actively stained mouse brain using multi-spectral MR microscopy , 2008, NeuroImage.

[25]  S. Crozier,et al.  Progress in high field MRI at the University of Florida , 2001, Magnetic Resonance Materials in Physics, Biology and Medicine.

[26]  P Wach,et al.  Imaging of the active B1 field in vivo , 1996, Magnetic resonance in medicine.

[27]  Clifford R Jack,et al.  In Vivo Magnetic Resonance Microimaging of Individual Amyloid Plaques in Alzheimer's Transgenic Mice , 2005, The Journal of Neuroscience.

[28]  A Van der Linden,et al.  Noninvasive in vivo MRI detection of neuritic plaques associated with iron in APP[V717I] transgenic mice, a model for Alzheimer's disease , 2005, Magnetic resonance in medicine.

[29]  G Allan Johnson,et al.  Design of a superconducting volume coil for magnetic resonance microscopy of the mouse brain. , 2008, Journal of magnetic resonance.

[30]  D Matthaei,et al.  Rapid three-dimensional MR imaging using the FLASH technique. , 1986, Journal of computer assisted tomography.

[31]  A. J. Morton,et al.  Use of magnetic resonance imaging for anatomical phenotyping of the R6/2 mouse model of Huntington's disease , 2009, Neurobiology of Disease.

[32]  P. Styles,et al.  A high-resolution NMR probe in which the coil and preamplifier are cooled with liquid helium. 1984. , 1984, Journal of magnetic resonance.

[33]  Reinhard Schliebs,et al.  Longitudinal assessment of Alzheimer's β‐amyloid plaque development in transgenic mice monitored by in vivo magnetic resonance microimaging , 2006, Journal of magnetic resonance imaging : JMRI.