Design, simulation and evaluation of uniform magnetic field systems for head-free eye movement recordings with scleral search coils

The precise measurement of eye movements is important for investigating vision, oculomotor control and vestibular function. The magnetic scleral search coil technique is one of the most precise measurement techniques for recording eye movements with very high spatial (≈ 1 arcmin) and temporal (>kHz) resolution. The technique is based on measuring voltage induced in a search coil through a large magnetic field. This search coil is embedded in a contact lens worn by a human subject. The measured voltage is in direct relationship to the orientation of the eye in space. This requires a magnetic field with a high homogeneity in the center, since otherwise the field inhomogeneity would give the false impression of a rotation of the eye due to a translational movement of the head. To circumvent this problem, a bite bar typically restricts head movement to a minimum. However, the need often emerges to precisely record eye movements under natural viewing conditions. To this end, one needs a uniform magnetic field that is uniform over a large area. In this paper, we present the numerical and finite element simulations of the magnetic flux density of different coil geometries that could be used for search coil recordings. Based on the results, we built a 2.2 × 2.2 × 2.2 meter coil frame with a set of 3 × 4 coils to generate a 3D magnetic field and compared the measured flux density with our simulation results. In agreement with simulation results, the system yields a highly uniform field enabling high-resolution recordings of eye movements.

[1]  G. Stroink,et al.  Uniform magnetic field produced by three, four, and five square coils , 1983 .

[2]  D. Robinson,et al.  A METHOD OF MEASURING EYE MOVEMENT USING A SCLERAL SEARCH COIL IN A MAGNETIC FIELD. , 1963, IEEE transactions on bio-medical engineering.

[3]  H. Collewijn,et al.  Precise recording of human eye movements , 1975, Vision Research.

[4]  Vernon Barger,et al.  Classical Electricity and Magnetism: A Contemporary Perspective , 1987 .

[5]  Sidney M. Rubens,et al.  Cube‐Surface Coil for Producing a Uniform Magnetic Field , 1945 .

[6]  Thomas Eggert,et al.  Improving the homogeneity of the magnetic field in the magnetic search coil technique , 2001, IEEE Transactions on Biomedical Engineering.

[7]  Michele Rucci,et al.  Decorrelation of neural activity during fixational instability: Possible implications for the refinement of V1 receptive fields , 2004, Visual Neuroscience.

[8]  Martina Poletti,et al.  Head-Eye Coordination at a Microscopic Scale , 2015, Current Biology.

[9]  Martina Poletti,et al.  Control and Functions of Fixational Eye Movements. , 2015, Annual review of vision science.

[10]  Marco Boi,et al.  Contributions of Eye Movement Transients to Spatial Vision. , 2015, Journal of vision.

[11]  J. Victor,et al.  Temporal Encoding of Spatial Information during Active Visual Fixation , 2012, Current Biology.

[12]  Giacomo Di Benedetto,et al.  Decoding 3D search coil signals in a non-homogeneous magnetic field , 2010, Vision Research.

[13]  Jorge Otero-Millan,et al.  Knowing what the brain is seeing in three dimensions: A novel, noninvasive, sensitive, accurate, and low-noise technique for measuring ocular torsion. , 2015, Journal of vision.

[14]  M. Rucci,et al.  Microsaccades Precisely Relocate Gaze in a High Visual Acuity Task , 2010, Nature Neuroscience.

[15]  Eileen Kowler,et al.  New directions for oculomotor research , 1990, Vision Research.