Use of UV-C radiation to disinfect non-critical patient care items: a laboratory assessment of the Nanoclave Cabinet

BackgroundThe near-patient environment is often heavily contaminated, yet the decontamination of near-patient surfaces and equipment is often poor. The Nanoclave Cabinet produces large amounts of ultraviolet-C (UV-C) radiation (53 W/m2) and is designed to rapidly disinfect individual items of clinical equipment. Controlled laboratory studies were conducted to assess its ability to eradicate a range of potential pathogens including Clostridium difficile spores and Adenovirus from different types of surface.MethodsEach test surface was inoculated with known levels of vegetative bacteria (106 cfu/cm2), C. difficile spores (102-106 cfu/cm2) or Adenovirus (109 viral genomes), placed in the Nanoclave Cabinet and exposed for up to 6 minutes to the UV-C light source. Survival of bacterial contaminants was determined via conventional cultivation techniques. Degradation of viral DNA was determined via PCR. Results were compared to the number of colonies or level of DNA recovered from non-exposed control surfaces. Experiments were repeated to incorporate organic soils and to compare the efficacy of the Nanoclave Cabinet to that of antimicrobial wipes.ResultsAfter exposing 8 common non-critical patient care items to two 30-second UV-C irradiation cycles, bacterial numbers on 40 of 51 target sites were consistently reduced to below detectable levels (≥ 4.7 log10 reduction). Bacterial load was reduced but still persisted on other sites. Objects that proved difficult to disinfect using the Nanoclave Cabinet (e.g. blood pressure cuff) were also difficult to disinfect using antimicrobial wipes. The efficacy of the Nanoclave Cabinet was not affected by the presence of organic soils. Clostridium difficile spores were more resistant to UV-C irradiation than vegetative bacteria. However, two 60-second irradiation cycles were sufficient to reduce the number of surface-associated spores from 103 cfu/cm2 to below detectable levels. A 3 log10 reduction in detectable Adenovirus DNA was achieved within 3 minutes; after 6 minutes, viral DNA was undetectable.ConclusionThe results of this study suggest that the Nanoclave Cabinet can provide rapid and effective disinfection of some patient-related equipment. However, laboratory studies do not necessarily replicate ‘in-use’ conditions and further tests are required to assess the usability, acceptability and relative performance of the Nanoclave Cabinet when used in situ.

[1]  T. Elliott,et al.  Re-usable ECG electrodes--a vehicle for cross-infection? , 1988, The Journal of hospital infection.

[2]  T. Y. Lee,et al.  Ultraviolet radiation for the sterilization of contact lenses. , 1990, The CLAO journal : official publication of the Contact Lens Association of Ophthalmologists, Inc.

[3]  W. Rutala,et al.  Disinfection and Sterilization of Patient-Care Items , 1996, Infection Control & Hospital Epidemiology.

[4]  F. Manian,et al.  Clostridium difficile Contamination of Blood Pressure Cuffs: A Call for a Closer Look at Gloving Practices in the Era of Universal Precautions , 1996, Infection Control & Hospital Epidemiology.

[5]  M. Zervos,et al.  Epidemiological study of hospital-acquired infection with vancomycin-resistant Enterococcus faecium: possible transmission by an electronic ear-probe thermometer. , 1997, Infection control and hospital epidemiology.

[6]  G. Ridgway,et al.  Evaluation of microbicidal activity of a new disinfectant: Sterilox 2500 against Clostridium difficile spores, Helicobacter pylori, vancomycin resistant Enterococcus species, Candida albicans and several Mycobacterium species. , 1999, The Journal of hospital infection.

[7]  S. Moreau,et al.  Low-temperature sterilization using gas plasmas: a review of the experiments and an analysis of the inactivation mechanisms. , 2001, International journal of pharmaceutics.

[8]  A. Heim,et al.  Rapid and quantitative detection of human adenovirus DNA by real‐time PCR , 2003, Journal of medical virology.

[9]  D. Gould The fall and rise of cleanliness in British healthcare and the nursing contribution , 2005 .

[10]  H. Humphreys,et al.  Environmental reservoirs of methicillin-resistant Staphylococcus aureus in isolation rooms: correlation with patient isolates and implications for hospital hygiene. , 2006, The Journal of hospital infection.

[11]  William A Rutala,et al.  Bacterial Contamination of Keyboards: Efficacy and Functional Impact of Disinfectants , 2006, Infection Control & Hospital Epidemiology.

[12]  F. Drangsholt,et al.  Comparison of UV C Light and Chemicals for Disinfection of Surfaces in Hospital Isolation Units , 2006, Infection Control & Hospital Epidemiology.

[13]  W. Hijnen,et al.  Inactivation credit of UV radiation for viruses, bacteria and protozoan (oo)cysts in water: a review. , 2006, Water research.

[14]  J. Cheesbrough,et al.  Blood pressure cuffs: friend or foe? , 2006, The Journal of hospital infection.

[15]  R. Weinstein,et al.  Reduction in acquisition of vancomycin-resistant enterococcus after enforcement of routine environmental cleaning measures. , 2006, Clinical infectious diseases : an official publication of the Infectious Diseases Society of America.

[16]  J. Otter,et al.  Assessing the biological efficacy and rate of recontamination following hydrogen peroxide vapour decontamination. , 2007, The Journal of hospital infection.

[17]  Lisa S. Smith,et al.  Disinfection of Acinetobacter baumannii-contaminated surfaces relevant to medical treatment facilities with ultraviolet C light. , 2007, Military medicine.

[18]  H Humphreys,et al.  The contribution of beds to healthcare-associated infection: the importance of adequate decontamination. , 2008, The Journal of hospital infection.

[19]  M. Exner,et al.  Comparison of chemo resistance of Clostridium difficile ribotype 027 spores and Bacillus subtilis spores against desinfectants , 2008 .

[20]  A. Mitchell,et al.  Stethoscope or 'Staphoscope'? Infection by auscultation. , 2010, The Journal of hospital infection.

[21]  S. Sattar Promises and pitfalls of recent advances in chemical means of preventing the spread of nosocomial infections by environmental surfaces. , 2010, American journal of infection control.

[22]  S. Shepherd,et al.  Spores of Clostridium difficile in Hospital Air. , 2010, Clinical infectious diseases : an official publication of the Infectious Diseases Society of America.

[23]  C. Donskey,et al.  Evaluation of an automated ultraviolet radiation device for decontamination of Clostridium difficile and other healthcare-associated pathogens in hospital rooms , 2010, BMC infectious diseases.

[24]  D Adams,et al.  Role of copper in reducing hospital environment contamination. , 2010, The Journal of hospital infection.

[25]  W. Rutala,et al.  Room Decontamination with UV Radiation , 2010, Infection Control & Hospital Epidemiology.

[26]  P. Humphreys Testing standards for sporicides. , 2011, The Journal of hospital infection.