Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy.

PURPOSE The purpose of this study is to present and evaluate a dose-verification protocol for pulsed dose-rate (PDR) brachytherapy based on in vivo time-resolved (1 s time resolution) fiber-coupled luminescence dosimetry. METHODS Five cervix cancer patients undergoing PDR brachytherapy (Varian GammaMed Plus with 192Ir) were monitored. The treatments comprised from 10 to 50 pulses (1 pulse/h) delivered by intracavitary/interstitial applicators (tandem-ring systems and/or needles). For each patient, one or two dosimetry probes were placed directly in or close to the tumor region using stainless steel or titanium needles. Each dosimeter probe consisted of a small aluminum oxide crystal attached to an optical fiber cable (1 mm outer diameter) that could guide radioluminescence (RL) and optically stimulated luminescence (OSL) from the crystal to special readout instrumentation. Positioning uncertainty and hypothetical dose-delivery errors (interchanged guide tubes or applicator movements from +/-5 to +/-15 mm) were simulated in software in order to assess the ability of the system to detect errors. RESULTS For three of the patients, the authors found no significant differences (P>0.01) for comparisons between in vivo measurements and calculated reference values at the level of dose per dwell position, dose per applicator, or total dose per pulse. The standard deviations of the dose per pulse were less than 3%, indicating a stable dose delivery and a highly stable geometry of applicators and dosimeter probes during the treatments. For the two other patients, the authors noted significant deviations for three individual pulses and for one dosimeter probe. These deviations could have been due to applicator movement during the treatment and one incorrectly positioned dosimeter probe, respectively. Computer simulations showed that the likelihood of detecting a pair of interchanged guide tubes increased by a factor of 10 or more for the considered patients when going from integrating to time-resolved dose verification. The likelihood of detecting a +/-15 mm displacement error increased by a factor of 1.5 or more. CONCLUSIONS In vivo fiber-coupled RL/OSL dosimetry based on detectors placed in standard brachytherapy needles was demonstrated. The time-resolved dose-rate measurements were found to provide a good way to visualize the progression and stability of PDR brachytherapy dose delivery, and time-resolved dose-rate measurements provided an increased sensitivity for detection of dose-delivery errors compared with time-integrated dosimetry.

[1]  Anatoly Rosenfeld,et al.  Verification of the plan dosimetry for high dose rate brachytherapy using metal-oxide-semiconductor field effect transistor detectors. , 2007, Medical physics.

[2]  Sören Mattsson,et al.  Real-time optical-fibre luminescence dosimetry for radiotherapy: physical characteristics and applications in photon beams , 2004, Physics in medicine and biology.

[3]  A Isambert,et al.  Spectral discrimination of Cerenkov radiation in scintillating dosimeters. , 2005, Medical physics.

[4]  Kari Tanderup,et al.  MRI-guided 3D optimization significantly improves DVH parameters of pulsed-dose-rate brachytherapy in locally advanced cervical cancer. , 2008, International journal of radiation oncology, biology, physics.

[5]  J. Williamson Current brachytherapy quality assurance guidance: does it meet the challenges of emerging image-guided technologies? , 2008, International journal of radiation oncology, biology, physics.

[6]  N Suchowerska,et al.  Real-time verification of HDR brachytherapy source location: implementation of detector redundancy , 2004, Physics in medicine and biology.

[7]  J. Borrás,et al.  Patterns of care for brachytherapy in Europe: facilities and resources in brachytherapy in the European area. , 2008, Brachytherapy.

[8]  D. Georg,et al.  In-vivo dosimetry for gynaecological brachytherapy: physical and clinical considerations. , 2005, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[9]  L. Anderson,et al.  Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43 , 1995 .

[10]  Steffen Greilich,et al.  Temperature coefficients for in vivo RL and OSL dosimetry using Al2O3:C , 2008 .

[11]  D Baltas,et al.  In vivo thermoluminescence dosimetry dose verification of transperineal 192Ir high-dose-rate brachytherapy using CT-based planning for the treatment of prostate cancer. , 2003, International journal of radiation oncology, biology, physics.

[12]  Steffen Greilich,et al.  Characterization of a fiber-coupled Al2O3:C luminescence dosimetry system for online in vivo dose verification during 192Ir brachytherapy. , 2009, Medical physics.

[13]  E G Yukihara,et al.  Optically stimulated luminescence (OSL) dosimetry in medicine , 2008, Physics in medicine and biology.

[14]  C E Andersen,et al.  An algorithm for real-time dosimetry in intensity-modulated radiation therapy using the radioluminescence signal from Al2O3:C. , 2006, Radiation protection dosimetry.

[15]  F. Ballester,et al.  Technical note: Monte-Carlo dosimetry of the HDR 12i and Plus 192Ir sources. , 2001, Medical physics.

[16]  Sten Nilsson,et al.  Dosimetry of anal radiation in high-dose-rate brachytherapy for prostate cancer. , 2007, Brachytherapy.

[17]  R. Pötter,et al.  Quality control in interstitial brachytherapy of the breast using pulsed dose rate: treatment planning and dose delivery with an Ir-192 afterloading system. , 2001, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[18]  Christian Kirisits,et al.  Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer. , 2008, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[19]  Louis Archambault,et al.  Measurement accuracy and cerenkov removal for high performance, high spatial resolution scintillation dosimetry. , 2005, Medical physics.

[20]  K. Tanderup,et al.  Geometric stability of intracavitary pulsed dose rate brachytherapy monitored by in vivo rectal dosimetry. , 2006, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[21]  L. Blank,et al.  Safety aspects of pulsed dose rate brachytherapy: analysis of errors in 1,300 treatment sessions. , 2008, International journal of radiation oncology, biology, physics.

[22]  A. Isambert,et al.  Multichannel dosemeter and Al2O3:C optically stimulated luminescence fibre sensors for use in radiation therapy: evaluation with electron beams. , 2008, Radiation protection dosimetry.

[23]  A G Dalgleish,et al.  Gene therapy for prostate cancer. , 1996, Gene therapy.

[24]  J. Bucci,et al.  The effect of rectal heterogeneity on wall dose in high dose rate brachytherapy. , 2008, Medical physics.

[25]  J. Lambert,et al.  In vivo dosimeters for HDR brachytherapy: a comparison of a diamond detector, MOSFET, TLD, and scintillation detector. , 2007, Medical physics.

[26]  Mark S. Akselrod,et al.  Optically stimulated luminescence of Al2O3 , 1998 .

[27]  T. Kron,et al.  Thermoluminescence dosimetry for in-vivo verification of high dose rate brachytherapy for prostate cancer , 2007, Australasian Physics & Engineering Sciences in Medicine.

[28]  J. Valentin,et al.  Protecting people against radiation exposure in the event of a radiological attack , 2005, Annals of the ICRP.

[29]  In vivo dosimetry of high-dose-rate interstitial brachytherapy in the pelvic region: use of a radiophotoluminescence glass dosimeter for measurement of 1004 points in 66 patients with pelvic malignancy. , 2008, International journal of radiation oncology, biology, physics.

[30]  J. Valentin,et al.  Abstract: Avoidance of radiation injuries from medical interventional procedures, ICRP Publication 85 , 2000 .

[31]  Ben Mijnheer,et al.  State of the art of in vivo dosimetry. , 2008, Radiation protection dosimetry.

[32]  S. McKeever,et al.  A real-time, high-resolution optical fibre dosemeter based on optically stimulated luminescence (OSL) of KBr:Eu, for potential use during the radiotherapy of cancer. , 2006, Radiation protection dosimetry.

[33]  A. Beddar,et al.  Plastic scintillation dosimetry and its application to radiotherapy , 2006 .

[34]  Abdelhamid Saoudi,et al.  Feasibility study of using MOSFET detectors for in vivo dosimetry during permanent low-dose-rate prostate implants. , 2006, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.

[35]  H Shirato,et al.  A feasibility study of novel plastic scintillation dosimetry with pulse-counting mode. , 2009, Physics in medicine and biology.

[36]  P. Lambin,et al.  In vivo dosimetry using a linear Mosfet-array dosimeter to determine the urethra dose in 125I permanent prostate implants. , 2009, International journal of radiation oncology, biology, physics.

[37]  A. Rosenfeld Electronic dosimetry in radiation therapy , 2006 .

[38]  Norberto Chiodini,et al.  Ce-doped optical fibre as radioluminescent dosimeter in radiotherapy , 2008 .