Investigation of PVA cryogel Young's modulus stability with time, controlled by a simple reliable technique.

We describe a quasistatic method for mechanical characterization of tissue-mimicking material used in elastography. We demonstrate that it is possible to assess the elasticity modulus with a reasonable reproducibility using simple and easy tools and methods. Possessing a simple relevant technique with evaluated relative error to assess Young's modulus of these phantoms could deeply improve the quality of the research in the field of elastography. The method was tested and validated with four samples of polyvinyl alcohol (PVA) cryogel with different elasticity values corresponding to those of stiffer soft biological tissues. Young's moduli, varying from 70 to 180 kPa depending on the number of freeze-thaw cycles (two to five), were measured within strict measurement conditions and found to have a reproducibility varying from 4% to 8%. Relative error, estimated as the ratio between observed and reference values, varied from 16% to 32%. Besides, measurement stability over 4 months was evaluated. The method demonstrated good feasibility and acceptable reproducibility for mechanically characterizing and controlled over time phantoms used for validating new potential ultrasound imaging techniques in the field of elastography. Nevertheless, in this study, investigation was performed on gel possessing young's modulus values ranging from 80 to 215 kPa. Some tissue values of Young'modulus were reported to be lower, ranging from 0.6 to 28 kPa as liver or glandular values. Consequently, further validation of this static method for mechanical characterization of phantom gels should be performed using softer PVA cryogel.

[1]  J. Ophir,et al.  Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues , 1991, Ultrasonic imaging.

[2]  D. Vray,et al.  Characterization of PVA cryogel for intravascular ultrasound elasticity imaging , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[3]  T. Krouskop,et al.  Elastic Moduli of Breast and Prostate Tissues under Compression , 1998, Ultrasonic imaging.

[4]  B. Rutt,et al.  Polyvinyl alcohol cryogel: An ideal phantom material for MR studies of arterial flow and elasticity , 1997, Magnetic resonance in medicine.

[5]  J. Gennisson,et al.  Estimation of polyvinyl alcohol cryogel mechanical properties with four ultrasound elastography methods and comparison with gold standard testings , 2007, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[6]  T. Peters,et al.  Poly(vinyl alcohol) cryogel phantoms for use in ultrasound and MR imaging , 2004, Physics in medicine and biology.

[7]  I Mano,et al.  New polyvinyl alcohol gel material for MRI phantoms , 1986, Magnetic resonance in medicine.

[8]  R H Selzer,et al.  Improved common carotid elasticity and intima-media thickness measurements from computer analysis of sequential ultrasound frames. , 2001, Atherosclerosis.

[9]  T. Varghese,et al.  Tissue-mimicking agar/gelatin materials for use in heterogeneous elastography phantoms , 2005, Physics in medicine and biology.

[10]  Wen-Chun Yeh,et al.  Elastic modulus measurements of human liver and correlation with pathology. , 2002, Ultrasound in medicine & biology.