Pulse oximetry using a disposable finger sensor placed on the forehead in hypoxic patients.

BACKGROUND Disposable pulse oximetry sensors designed for the finger are sometimes placed on the forehead. Although finger sensors have not been designed to accurately measure arterial oxygen saturation when placed on other body parts, they are used on the forehead when an oximetry waveform cannot be obtained from the finger. OBJECTIVES The purpose of this prospective observational study was to measure the agreement between disposable transmittal sensors placed on the finger and the same sensors placed on the forehead. METHODS We used a convenience sample of hypoxic emergency department patients. Patients were attached to three oximeters: 1) a non-disposable finger sensor (reference); 2) a disposable finger sensor placed on the finger; and 3) a disposable finger sensor placed on the forehead. Data were analyzed using the Bland-Altman method. A difference of 5% was considered clinically significant. RESULTS Twenty-five patients were approached, and 20 were enrolled. Eleven of the patients had an absolute difference between the disposable finger sensor placed on the finger compared to the forehead > or = 5%. CONCLUSION Pulse oximetry measurements taken on the forehead using a disposable finger sensor were inaccurate in over half of the subjects. Therefore, disposable finger oximetry sensors should not be placed on the forehead.

[1]  B. Kataria,et al.  Nail polish does not affect pulse oximeter saturation. , 1986, Anesthesia and analgesia.

[2]  N. Frickey,et al.  Forehead SpO2 monitoring compared to finger SpO2 recording in emergency transport , 2004, Anaesthesia.

[3]  D. Altman,et al.  Measuring agreement in method comparison studies , 1999, Statistical methods in medical research.

[4]  D C Hoaglin,et al.  The Effect of Nail Polish on Pulse Oximetry , 1988, Anesthesia and analgesia.

[5]  W B Runciman,et al.  Potential errors in pulse oximetry III: Effects of interference, dyes, dyshaemoglobins and other pigments * , 1991, Anaesthesia.

[6]  W B Runciman,et al.  Potential errors in pulse oximetry , 1991, Anaesthesia.

[7]  E. Cheng,et al.  Forehead pulse oximetry compared with finger pulse oximetry and arterial blood gas measurement , 1988, Journal of Clinical Monitoring.

[8]  W B Runciman,et al.  Pulse oximeter probes A comparison between finger, nose, ear and forehead probes under conditions of poor perfusion * , 1991, Anaesthesia.

[9]  Renate Huch,et al.  Limitations of forehead pulse oximetry , 2005, Journal of Clinical Monitoring.

[10]  A. C. M. Dassel,et al.  Reflectance pulse oximetry at the forehead improves by pressure on the probe , 1995, Journal of Clinical Monitoring.

[11]  Lynn Schallom,et al.  Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusion. , 2007, Heart & lung : the journal of critical care.

[12]  R. K. Webb,et al.  Potential errors in pulse oximetry , 1991, Anaesthesia.

[13]  Richard D Branson,et al.  Forehead oximetry in critically ill patients: the case for a new monitoring site. , 2004, Respiratory care clinics of North America.

[14]  D. Altman,et al.  Applying the right statistics: analyses of measurement studies , 2003, Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology.

[15]  J. Severinghaus,et al.  Errors in 14 pulse oximeters during profound hypoxia , 1989, Journal of Clinical Monitoring.

[16]  S. Sugino,et al.  Forehead is as sensitive as finger pulse oximetry during general anesthesia , 2004, Canadian journal of anaesthesia = Journal canadien d'anesthesie.

[17]  E. Chan,et al.  What is the effect of fingernail polish on pulse oximetry? , 2003, Chest.

[18]  Gregory D. Jay,et al.  Enamel Nail Polish Does Not Interfere With Pulse Oximetry Among Normoxic Volunteers , 2002, Journal of Clinical Monitoring and Computing.

[19]  R. Stout,et al.  The Effect of Venous Pulsation on the Forehead Pulse Oximeter Wave Form as a Possible Source of Error in Spo2 Calculation , 2005, Anesthesia and analgesia.

[20]  P. Andrews,et al.  A comparison of the Invos 3100 and the Critikon 2020 near‐infrared spectrophotometers as monitors of cerebral oxygenation , 1997, Anaesthesia.

[21]  R. K. Webb,et al.  A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion , 1991, Anaesthesia.

[22]  Paul D. Mannheimer,et al.  The Influence of Larger Subcutaneous Blood Vessels on Pulse Oximetry , 2003, Journal of Clinical Monitoring and Computing.