Turbulence in surgical suction heads as detected by MRI

Background: Blood loss is common during surgical procedures, especially in open cardiac surgery. Allogenic blood transfusion is associated with increased morbidity and mortality. Blood conservation programs in cardiac surgery recommend re-transfusion of shed blood directly or after processing, as this decreases transfusion rates of allogenic blood. But aspiration of blood from the wound area is often associated with increased hemolysis, due to flow induced forces, mainly through development of turbulence. Methods: We evaluated magnetic resonance imaging (MRI) as a qualitative tool for detection of turbulence. MRI is sensitive to flow; this study uses velocity-compensated T1-weighted 3D MRI for turbulence detection in four geometrically different cardiotomy suction heads under comparable flow conditions (0–1250 mL/min). Results: Our standard control suction head Model A showed pronounced signs of turbulence at all flow rates measured, while turbulence was only detectable in our modified Models 1–3 at higher flow rates (Models 1 and 3) or not at all (Model 2). Conclusions: The comparison of flow performance of surgical suction heads with different geometries via acceleration-sensitized 3D MRI revealed significant differences in turbulence development between our standard control Model A and the modified alternatives (Models 1–3). As flow conditions during measurement have been comparable, the specific geometry of the respective suction heads must have been the main factor responsible. The underlying mechanisms and causative factors can only be speculated about, but as other investigations have shown, hemolytic activity is positively associated with degree of turbulence. The turbulence data measured in this study correlate with data from other investigations about hemolysis induced by surgical suction heads. The experimental MRI technique used showed added value for further elucidating the underlying physical phenomena causing blood damage due to non-physiological flow.

[1]  I. Kutschka,et al.  A New Evaluation Q-Factor to Be Calculated for Suction Geometries as a Basis for Smooth Suction in the Operating Field to Ensure the Highest Possible Blood Integrity for Retransfusion Systems. , 2022, The journal of extra-corporeal technology.

[2]  M. Mazzeffi,et al.  Coagulation and Transfusion Updates From 2021 , 2022, Journal of Cardiothoracic and Vascular Anesthesia.

[3]  R. Malinauskas,et al.  The Effect of Blood Viscosity on Shear-Induced Hemolysis using a Magnetically Levitated Shearing Device. , 2022, Artificial organs.

[4]  Jiang Li,et al.  Slug Flow Hydrodynamics Modeling for Gas–Liquid Two-Phase Flow in a Pipe , 2022, Energies.

[5]  H. Ullah,et al.  Cardiac Surgery and Blood-Saving Techniques: An Update , 2022, Cureus.

[6]  M. Trivella,et al.  Transfusion thresholds for guiding red blood cell transfusion. , 2021, The Cochrane database of systematic reviews.

[7]  J. Fraser,et al.  Erythrocyte morphological symmetry analysis to detect sublethal trauma in shear flow , 2021, Scientific Reports.

[8]  M. Zhang,et al.  Risk and Safety Perceptions Contribute to Transfusion Decisions in Coronary Artery Bypass Grafting , 2021, The Journal of ExtraCorporeal Technology.

[9]  Hongyue Wang,et al.  Cell Saver Blood Reinfusion Up to 24 Hours Post Collection in Pediatric Cardiac Surgical Patients Does Not Increase Incidence of Hospital-Acquired Infections or Mortality. , 2021, The journal of extra-corporeal technology.

[10]  R. Malinauskas,et al.  The Effect of Blood Viscosity on Shear-Induced Hemolysis using a Magnetically Levitated Shearing Device , 2021, bioRxiv.

[11]  P. Meybohm,et al.  The association between intraoperative cell salvage and red blood cell transfusion in cardiac surgery – an observational study in a patient blood management centre , 2021, Anaesthesiology intensive therapy.

[12]  M. Picot,et al.  Restrictive Transfusion Strategy after Cardiac Surgery. , 2021, Anesthesiology.

[13]  D. Fergusson,et al.  Intraoperative Red Blood Cell Transfusion Decision-making , 2020, Annals of surgery.

[14]  N. Secher,et al.  Effect of low vs high haemoglobin transfusion trigger on cardiac output in patients undergoing elective vascular surgery: Post‐hoc analysis of a randomized trial , 2020, Acta anaesthesiologica Scandinavica.

[15]  M. Karlsson,et al.  Characterization of anisotropic turbulence behavior in pulsatile blood flow , 2020, Biomechanics and Modeling in Mechanobiology.

[16]  K. Kırali,et al.  The effect of vacuum-assisted venous drainage on hemolysis during cardiopulmonary bypass. , 2020, American journal of cardiovascular disease.

[17]  S. Okhota,et al.  Shear Stress-Induced Activation of von Willebrand Factor and Cardiovascular Pathology , 2020, International journal of molecular sciences.

[18]  T. Carrel,et al.  The impact of roller pump-assisted cardiotomy suction unit on hemolysis. , 2020, Artificial organs.

[19]  B. Griffith,et al.  The Role of a Disintegrin and Metalloproteinase Proteolysis and Mechanical Damage in Nonphysiological Shear Stress-Induced Platelet Receptor Shedding , 2020, ASAIO journal.

[20]  J. R. Grace Slug Flow , 2020, Essentials of Fluidization Technology.

[21]  F. Zeman,et al.  Factors associated with hemolysis during extracorporeal membrane oxygenation (ECMO)—Comparison of VA- versus VV ECMO , 2020, PloS one.

[22]  Jiafeng Zhang,et al.  Impact of High Mechanical Shear Stress and Oxygenator Membrane Surface on Blood Damage Relevant to Thrombosis and Bleeding in a Pediatric ECMO Circuit. , 2020, Artificial organs.

[23]  R. Bartlett,et al.  Inflammatory Effects of Blood–Air Interface in a Porcine Cardiopulmonary Bypass Model , 2020, ASAIO journal.

[24]  C. Tam A phenomenological approach to jet noise: the two-source model , 2019, Philosophical Transactions of the Royal Society A.

[25]  T. Pawlik,et al.  Clinical Utility of Autologous Salvaged Blood: a Review , 2019, Journal of Gastrointestinal Surgery.

[26]  A. Driessen,et al.  Patient blood management in the cardiac surgical setting: An updated overview. , 2019, Transfusion and apheresis science : official journal of the World Apheresis Association : official journal of the European Society for Haemapheresis.

[27]  Nathan C. Crawford,et al.  Turbulent Flow Promotes Cleavage of VWF (von Willebrand Factor) by ADAMTS13 (A Disintegrin and Metalloproteinase With a Thrombospondin Type-1 Motif, Member 13). , 2019, Arteriosclerosis, thrombosis, and vascular biology.

[28]  Binghuan Huang,et al.  Experimental and Numerical Investigations on the Flow Characteristics within Hydrodynamic Entrance Regions in Microchannels , 2019, Micromachines.

[29]  S. Manabe,et al.  Safety and efficacy of a simple cardiotomy suction system as a blood salvage procedure during off-pump coronary artery bypass surgery , 2019, Journal of Artificial Organs.

[30]  M. Sharp,et al.  Modeling and prediction of flow-induced hemolysis: a review , 2019, Biomechanics and Modeling in Mechanobiology.

[31]  T. Seyfried,et al.  [Cell salvage : Scientific evidence, clinical practice and legal framework]. , 2019, Der Anaesthesist.

[32]  Min A Kwon,et al.  Korean clinical practice guideline for perioperative red blood cell transfusion from Korean Society of Anesthesiologists , 2018, Korean journal of anesthesiology.

[33]  M G Friedrich,et al.  Modifications of Surgical Suction Tip Geometry for Flow Optimisation: Influence on Suction-Induced Noise Pollution , 2018, Surgery research and practice.

[34]  Raymond K. Newswanger,et al.  Stress and Exposure Time on von Willebrand Factor Degradation , 2018, Artificial organs.

[35]  L. Goubergrits,et al.  Past and future of blood damage modelling in a view of translational research , 2018, The International journal of artificial organs.

[36]  J. Riggert,et al.  The effect of a novel turbulence-controlled suction system in the prevention of hemolysis and platelet dysfunction in autologous surgery blood , 2018, Perfusion.

[37]  D. Holt,et al.  Improving hemolysis levels associated with cardiotomy suction , 2018, Perfusion.

[38]  P. Meybohm,et al.  Washed cell salvage in surgical patients: A review and meta-analysis of prospective randomized trials under PRISMA: Erratum , 2018, Medicine.

[39]  P. Wenig,et al.  New device for intraoperative blood suction avoiding turbulences , 2018, BMJ Innovations.

[40]  S. Sabapathy,et al.  Repetitive Supra‐Physiological Shear Stress Impairs Red Blood Cell Deformability and Induces Hemolysis , 2017, Artificial organs.

[41]  H. Meiselman,et al.  Prediction of the level and duration of shear stress exposure that induces subhemolytic damage to erythrocytes. , 2017, Biorheology.

[42]  Jens Frahm,et al.  Image denoising for real‐time MRI , 2017, Magnetic resonance in medicine.

[43]  Jinkwon Lee,et al.  Hemolysis During Open-Heart Surgery With Vacuum-Assisted Venous Drainage at Different Negative Pressures in Pediatric Patients Weighing Less Than 10 kilograms , 2017, World journal for pediatric & congenital heart surgery.

[44]  M. D. de Bruijne,et al.  Benchmarking the use of blood products in cardiac surgery to stimulate awareness of transfusion behaviour , 2016, Netherlands Heart Journal.

[45]  Ansar Hassan,et al.  Efficacy of intraoperative cell salvage in decreasing perioperative blood transfusion rates in first-time cardiac surgery patients: a retrospective study. , 2016, Canadian journal of surgery. Journal canadien de chirurgie.

[46]  P. Meybohm,et al.  Washed cell salvage in surgical patients , 2016, Medicine.

[47]  M. Alam,et al.  Turbulent flow analysis on bend and downstream of the bend for different curvature ratio , 2016 .

[48]  Zhongjun J. Wu,et al.  Paradoxical Effect of Nonphysiological Shear Stress on Platelets and von Willebrand Factor. , 2016, Artificial organs.

[49]  Mihailo R. Jovanovic,et al.  Input-output analysis of high-speed axisymmetric isothermal jet noise , 2016 .

[50]  H. Hege,et al.  Turbulence in Blood Damage Modeling , 2016, The International journal of artificial organs.

[51]  K. Träger,et al.  Treatment of Post-Cardiopulmonary Bypass SIRS by Hemoadsorption: A Case Series , 2016, The International journal of artificial organs.

[52]  Dwight Barkley,et al.  The rise of fully turbulent flow , 2015, Nature.

[53]  D. Papavassiliou,et al.  Hemolysis Related to Turbulent Eddy Size Distributions Using Comparisons of Experiments to Computations. , 2015, Artificial organs.

[54]  Jinwei Xie,et al.  Is postoperative cell salvage necessary in total hip or knee replacement? A meta-analysis of randomized controlled trials. , 2015, International journal of surgery.

[55]  B. Griffith,et al.  Shear-Induced Hemolysis: Species Differences. , 2015, Artificial organs.

[56]  M. Maddali,et al.  Cell salvage during coronary artery bypass surgery and allogenic blood exposure , 2015, Asian cardiovascular & thoracic annals.

[57]  T. Schmitz-Rode,et al.  Hemolysis in a laminar flow-through Couette shearing device: an experimental study. , 2014, Artificial organs.

[58]  Khandakar Niaz Morshed,et al.  Theory to Predict Shear Stress on Cells in Turbulent Blood Flow , 2014, PloS one.

[59]  Hai‐feng Liu,et al.  The deterministic chaos and random noise in turbulent jet. , 2014, Chaos.

[60]  G. Ferrari,et al.  Blood transfusion in cardiac surgery does increase the risk of 5‐year mortality: results from a contemporary series of 1714 propensity‐matched patients , 2014, Transfusion.

[61]  M. Chern,et al.  The effect of turbulent viscous shear stress on red blood cell hemolysis , 2014, Journal of Artificial Organs.

[62]  Guerlain Ulysse,et al.  Effect of Air Exposure and Suction on Blood Cell Activation and Hemolysis in an In Vitro Cardiotomy Suction Model , 2013, ASAIO journal.

[63]  M. Poullis,et al.  Is single-unit blood transfusion bad post-coronary artery bypass surgery? , 2013, Interactive Cardiovascular and Thoracic Surgery.

[64]  V. Ferraris,et al.  The relationship between intraoperative blood transfusion and postoperative systemic inflammatory response syndrome. , 2013, American journal of surgery.

[65]  Dionne A. Graham,et al.  Pediatric ECMO Outcomes: Comparison of Centrifugal Versus Roller Blood Pumps Using Propensity Score Matching , 2013, ASAIO journal.

[66]  M. Ruel,et al.  Centrifugal pump and roller pump in adult cardiac surgery: a meta-analysis of randomized controlled trials. , 2012, Artificial organs.

[67]  J. Riley Pump sucker discipline. , 2012, The journal of extra-corporeal technology.

[68]  A. Bhutta,et al.  Hemolysis During Cardiac Extracorporeal Membrane Oxygenation: A Case-Control Comparison of Roller Pumps and Centrifugal Pumps in a Pediatric Population , 2011, ASAIO journal.

[69]  Howard K. Song,et al.  2011 update to the Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists blood conservation clinical practice guidelines. , 2011, The Annals of thoracic surgery.

[70]  L. Anderson Blood Conservation in Cardiac Surgery , 2010, Journal of perioperative practice.

[71]  Mehrdad Massoudi,et al.  On the Representation of Turbulent Stresses for Computing Blood Damage. , 2010, International journal of engineering science.

[72]  G. Biondi-Zoccai,et al.  Miniaturized cardiopulmonary bypass improves short-term outcome in cardiac surgery: a meta-analysis of randomized controlled studies. , 2010, The Journal of thoracic and cardiovascular surgery.

[73]  V. Lonský,et al.  Pathophysiological aspects of cardiotomy suction usage , 2010, Perfusion.

[74]  A. Moxey,et al.  Cell salvage for minimising perioperative allogeneic blood transfusion. , 2010, The Cochrane database of systematic reviews.

[75]  N. Quinlan,et al.  Effect of Eddy Length Scale on Mechanical Loading of Blood Cells in Turbulent Flow , 2009, Annals of Biomedical Engineering.

[76]  Petter Dyverfeldt,et al.  On MRI turbulence quantification. , 2009, Magnetic resonance imaging.

[77]  R. Bartlett,et al.  The Relationships Between Air Exposure, Negative Pressure, and Hemolysis , 2009, ASAIO journal.

[78]  D. Bainbridge,et al.  The Efficacy of an Intraoperative Cell Saver During Cardiac Surgery: A Meta-Analysis of Randomized Trials , 2009, Anesthesia and analgesia.

[79]  F. Sellke,et al.  The Association of Perioperative Red Blood Cell Transfusions and Decreased Long-Term Survival After Cardiac Surgery , 2009, Anesthesia and analgesia.

[80]  F. Broccolo,et al.  Myocardial revascularization with miniaturized extracorporeal circulation versus off pump: Evaluation of systemic and myocardial inflammatory response in a prospective randomized study. , 2009, The Journal of thoracic and cardiovascular surgery.

[81]  R. Emery,et al.  Coronary artery bypass grafting with a minimized cardiopulmonary bypass circuit: a prospective, randomized trial. , 2009, The Journal of thoracic and cardiovascular surgery.

[82]  S. Takatani,et al.  Mechanical damage of red blood cells by rotary blood pumps: selective destruction of aged red blood cells and subhemolytic trauma. , 2008, Artificial organs.

[83]  N. Moat,et al.  Coronary artery surgery: cardiotomy suction or cell salvage? , 2007, Journal of cardiothoracic surgery.

[84]  A. Anselmi,et al.  Prospective Randomized Comparison of Coronary Bypass Grafting With Minimal Extracorporeal Circulation System (MECC) Versus Off-Pump Coronary Surgery , 2007, Circulation.

[85]  I. Komuro,et al.  Predictive indicators of deep venous thrombosis and pulmonary arterial thromboembolism in 54 subjects after total knee arthroplasty using multislice computed tomography in logistic regression models. , 2007, International journal of cardiology.

[86]  J. Yorke,et al.  Turbulence transition and the edge of chaos in pipe flow. , 2007, Physical review letters.

[87]  J. Waters,et al.  Modification of Suction-Induced Hemolysis During Cell Salvage , 2007, Anesthesia and analgesia.

[88]  T. Ebbers,et al.  Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase‐contrast MRI , 2006, Magnetic resonance in medicine.

[89]  J. Sellgren,et al.  Hemodynamic effects of cardiotomy suction blood. , 2006, The Journal of thoracic and cardiovascular surgery.

[90]  G. Nuttall,et al.  Excessive Bleeding and Transfusion in a Prior Cardiac Surgery is Associated with Excessive Bleeding and Transfusion in the Next Surgery , 2006, Anesthesia and analgesia.

[91]  Prof. Dr. med. Gustav Steinhoff,et al.  Pericardial suction blood separation attenuates inflammatory response and hemolysis after cardiopulmonary bypass , 2006, Scandinavian cardiovascular journal : SCJ.

[92]  Giacomo Di Benedetto,et al.  A novel formulation for blood trauma prediction by a modified power-law mathematical model , 2005, Biomechanics and modeling in mechanobiology.

[93]  R. Kerswell,et al.  Recent progress in understanding the transition to turbulence in a pipe , 2005 .

[94]  Mitsuo Umezu,et al.  Effects of Turbulent Stresses upon Mechanical Hemolysis: Experimental and Computational Analysis , 2004, ASAIO journal.

[95]  Amir A Amini,et al.  Factors affecting the accuracy of pressure measurements in vascular stenoses from phase‐contrast MRI , 2004, Magnetic resonance in medicine.

[96]  A. Bengtsson,et al.  Coronary surgery without cardiotomy suction and autotransfusion reduces the postoperative systemic inflammatory response. , 2004, The Annals of thoracic surgery.

[97]  Klaus Affeld,et al.  Numerical estimation of blood damage in artificial organs. , 2004, Artificial organs.

[98]  R. Paul,et al.  Shear stress related blood damage in laminar couette flow. , 2003, Artificial organs.

[99]  K. Engström,et al.  The inflammatory response to recycled pericardial suction blood and the influence of cell-saving , 2003, Scandinavian cardiovascular journal : SCJ.

[100]  Pascal Verdonck,et al.  Numerical calculation of hemolysis levels in peripheral hemodialysis cannulas. , 2002, Artificial organs.

[101]  C. Triggs,et al.  Limitation of thrombin generation, platelet activation, and inflammation by elimination of cardiotomy suction in patients undergoing coronary artery bypass grafting treated with heparin-bonded circuits. , 2002, The Journal of thoracic and cardiovascular surgery.

[102]  J. Pincemail,et al.  Levels of inflammatory markers in the blood processed by autotransfusion devices during cardiac surgery associated with cardiopulmonary bypass circuit , 2002, Perfusion.

[103]  P. Lu,et al.  A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow. , 2001, Journal of biomechanics.

[104]  G. Wright Haemolysis during cardiopulmonary bypass: update , 2001, Perfusion.

[105]  L. V. von Segesser,et al.  Smart suction device for less blood trauma: a comparison with Cell Saver. , 2001, European Journal of Cardio-Thoracic Surgery.

[106]  J. Hammon,et al.  Processing scavenged blood with a cell saver reduces cerebral lipid microembolization. , 2000, The Annals of thoracic surgery.

[107]  D. Sharpe,et al.  Haemolysis during cardiopulmonary bypass: an in vivo comparison of standard roller pumps, nonocclusive roller pumps and centrifugal pumps , 1999, Perfusion.

[108]  L. V. von Segesser,et al.  In situ control of cardiotomy suction reduces blood trauma. , 1998, ASAIO journal.

[109]  E. Gams,et al.  Centrifugal pumping during routine open heart surgery improves clinical outcome. , 1998, Artificial organs.

[110]  R. Pettigrew,et al.  Turbulent Fluctuation Velocity: The Most Significant Determinant of Signal Loss in Stenotic Vessels , 1995, Magnetic resonance in medicine.

[111]  E. Papoutsakis,et al.  Damage mechanisms of suspended animal cells in agitated bioreactors with and without bubble entrainment , 1990, Biotechnology and bioengineering.

[112]  H. Reul,et al.  Estimation of Shear Stress-related Blood Damage in Heart Valve Prostheses - in Vitro Comparison of 25 Aortic Valves , 1990, The International journal of artificial organs.

[113]  P. Sperryn,et al.  Blood. , 1989, British journal of sports medicine.

[114]  W. van Oeveren,et al.  Effects of aprotinin on hemostatic mechanisms during cardiopulmonary bypass. , 1987, The Annals of thoracic surgery.

[115]  D. Firmin,et al.  Measurement of Flow with NMR Imaging Using a Gradient Pulse and Phase Difference Technique , 1984, Journal of computer assisted tomography.

[116]  G. Wright,et al.  Cellular aggregation and trauma in cardiotomy suction systems. , 1979, Thorax.

[117]  M. H. Wagner,et al.  Developing flow in circular conduits: transition from plug flow to tube flow , 1975, Journal of Fluid Mechanics.

[118]  J. Wielogorski,et al.  The effects of subatmospheric pressure on the haemolysis of blood. , 1975, Journal of biomechanics.

[119]  R. Aster,et al.  Red Blood Cell Transfusion , 1973, Transfusion.

[120]  S P Sutera,et al.  HEMOLYSIS AND SUBHEMOLYTIC ALTERATIONS OF HUMAN RBC INDUCED BY TURBULENT SHEAR FLOW , 1972, Transactions - American Society for Artificial Internal Organs.

[121]  E. B. Christiansen,et al.  Entrance region flow , 1965 .

[122]  Ephraim M Sparrow,et al.  Flow Development in the Hydrodynamic Entrance Region of Tubes and Ducts , 1964 .

[123]  A. Stammers,et al.  The Influence of Intraoperative Autotransfusion on Postoperative Hematocrit after Cardiac Surgery: A Cross-Sectional Study. , 2017, The journal of extra-corporeal technology.

[124]  C. Bottura,et al.  Key Points for Curbing Cardiopulmonary Bypass Inflammation. , 2016, Acta cirurgica brasileira.

[125]  Danny Bluestein,et al.  Evaluation of Shear-Induced Platelet Activation Models Under Constant and Dynamic Shear Stress Loading Conditions Relevant to Devices , 2013, Annals of biomedical engineering.

[126]  A. Smits,et al.  TURBULENT PIPE FLOW THROUGH A 90° BEND , 2011, Proceeding of Seventh International Symposium on Turbulence and Shear Flow Phenomena.

[127]  Howard K. Song,et al.  Update to The Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists Blood Conservation Clinical Practice Guidelines , 2011 .

[128]  L. V. von Segesser,et al.  A novel device for reducing hemolysis provoked by cardiotomy suction during open heart cardiopulmonary bypass surgery: a randomized prospective study. , 2007, Artificial organs.

[129]  H. Nishida,et al.  Less blood damage in the impeller centrifugal pump: a comparative study with the roller pump in open heart surgery. , 1996, Artificial organs.

[130]  P. Blackshear,et al.  A low-hemolysis blood aspirator conserves blood during surgery. , 1995, Biomedical Instrumentation & Technology.

[131]  Philippe Grenier,et al.  EVOLUTION OF SLUG FLOW IN LONG PIPE , 1993 .

[132]  R. E. Collingham,et al.  Shear, wall interaction and hemolysis. , 1966, Transactions - American Society for Artificial Internal Organs.

[133]  P. Blackshear,et al.  SOME MECHANICAL EFFECTS THAT INFLUENCE HEMOLYSIS. , 1965, Transactions - American Society for Artificial Internal Organs.

[134]  Kh. I. Amirkhanov,et al.  the Influence , 2022 .