Wide tourniquets eliminate blood flow at low inflation pressures.

Previous recommendations for use of pneumatic tourniquets in extremity surgery suggest parameters for maximum pressure and time limits without regard for optimum cuff width. Wide cuffs produce lower readings of blood pressure relative to narrow cuffs, presumably because the wide cuffs arrest flow at lower inflation pressure. We investigated three tourniquet sizes and the inflation pressure required to eliminate flow to the upper extremity using an ultrasonic Doppler device to monitor blood flow in the radial artery of ten normal subjects (arm circumference 24.5 to 37 cm). Arterial flow was always eliminated at the lowest pressure using the widest tourniquet cuff. Significantly lower inflation pressures will eliminate blood flow if wider tourniquet cuffs are used; therefore, use of a wider tourniquet cuff may result in a reduced incidence of tourniquet complications.

[1]  O. B. Heywood,et al.  Bio-Mechanical Aspects of the Tourniquet , 1973, The Hand.

[2]  A. Shapiro Steady Flow in Collapsible Tubes , 1977 .

[3]  C. Nordborg,et al.  Changes in nerve function and nerve fibre structure induced by acute, graded compression. , 1980, Journal of neurology, neurosurgery, and psychiatry.

[4]  J. Doupe,et al.  Some factors affecting the auscultatory measurement of arterial blood pressures. , 1949, Canadian journal of research.

[5]  A A Sapega,et al.  Optimizing tourniquet application and release times in extremity surgery. A biochemical and ultrastructural study. , 1985, The Journal of bone and joint surgery. American volume.

[6]  R. Balderston,et al.  Pathophysiologic effects distal to a tourniquet in the dog. , 1979, Journal of Trauma.

[7]  D. Denny-Brown,et al.  PARALYSIS OF NERVE INDUCED BY DIRECT PRESSURE AND BY TOURNIQUET , 1944 .

[8]  A. Flatt Tourniquet time in hand surgery. , 1972, Archives of surgery.

[9]  F. H. Bentley,et al.  The effects of pressure on conduction in peripheral nerve , 1943, The Journal of physiology.

[10]  J. Shaw,et al.  The relationship between tourniquet pressure and underlying soft-tissue pressure in the thigh. , 1982, The Journal of bone and joint surgery. American volume.

[11]  J. M. Bruner SAFFTY FACTORS IN THE USE OF THE PNEUMATIC TOURNIQUET FOR HEMOSTASIS IN SURGERY OF THE HAND , 1951 .

[12]  L. Klenerman,et al.  The effect of pneumatic tourniquets on the ultrastructure of skeletal muscle. , 1979, The Journal of bone and joint surgery. British volume.

[13]  W. Akeson,et al.  Local compression patterns beneath pneumatic tourniquets applied to arms and thighs of human cadavera , 1987, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[14]  W. H. Jacob,et al.  Tourniquet hemostasis. A clinical study. , 1983, Clinical orthopaedics and related research.

[15]  G. Lundborg,et al.  Blockage of axonal transport induced by acute, graded compression of the rabbit vagus nerve. , 1980, Journal of neurology, neurosurgery, and psychiatry.

[16]  C. Rorabeck,et al.  The pressure distribution under tourniquets. , 1985, The Journal of bone and joint surgery. American volume.

[17]  I. D. Hall,et al.  Intravenous regional analgesia , 1984, Anaesthesia.

[18]  G Lundborg,et al.  Effects of graded compression on intraneural blood blow. An in vivo study on rabbit tibial nerve. , 1981, The Journal of hand surgery.

[19]  T. J. Fowler,et al.  Recovery of nerve conduction after a pneumatic tourniquet: observations on the hind-limb of the baboon1 , 1972, Journal of neurology, neurosurgery, and psychiatry.

[20]  L. Klenerman Tourniquet Time — how Long? , 1980, The Hand.

[21]  T. J. Fowler,et al.  Anatomical changes in peripheral nerves compressed by a pneumatic tourniquet. , 1972, Journal of anatomy.

[22]  A. C. Burton,et al.  RECOMMENDATIONS for human blood pressure determinations by sphygmomanometers. , 1967, Journal of the American Medical Association.