A novel fiber-optic measurement system for the evaluation of performances of neonatal pulmonary ventilators

Published standards for the performance evaluation of pulmonary ventilators are mainly directed to manufacturers rather than to end-users and often considered inadequate or not comprehensive. In order to contribute to overcome the problems above, a novel measurement system was proposed and tested with waveforms of mechanical ventilation by means of experimental trials carried out with infant ventilators typically used in neonatal intensive care units: the main quantities of mechanical ventilation in newborns are monitored, i.e. air flow rate, differential pressure and volume from infant ventilator are measured by means of two novel fiber-optic sensors (OFSs) developed and characterized by the authors, while temperature and relative humidity of air mass are obtained by two commercial transducers. The proposed fiber-optic sensors (flow sensor Q-OFS, pressure sensor P-OFS) showed measurement ranges of air flow and pressure typically encountered in neonatal mechanical ventilation, i.e. the air flow rate Q ranged from 3 l min−1 to 18 l min−1 (inspiratory) and from −3 l min−1 to −18 l min−1 (expiratory), the differential pressure ΔP ranged from −15 cmH2O to 15 cmH2O. In each experimental trial carried out with different settings of the ventilator, outputs of the OFSs are compared with data from two reference sensors (reference flow sensor RF, reference pressure sensor RP) and results are found consistent: flow rate Q showed a maximum error between Q-OFS and RF up to 13 percent, with an output ratio Q RF/Q OFS of not more than 1.06 ± 0.09 (least square estimation, 95 percent confidence level, R 2 between 0.9822 and 0.9931). On the other hand the maximum error between P-OFS and RP on differential pressure ΔP was lower than 10 percent, with an output ratio ΔP RP/ΔP OFS between 0.977 ± 0.022 and 1.0 ± 0.8 (least square estimation, 95 percent confidence level, R 2 between 0.9864 and 0.9876). Despite the possible improvements, results were encouraging and suggested the proposed measurement system can be considered suitable for performances evaluation of neonatal ventilators and useful for both end-users and manufacturers.

[1]  Andrea Scorza,et al.  Experimental characterization of a novel fiber-optic accelerometer for the quantitative assessment of rest tremor in Parkinsonian patients , 2012, BioMed 2012.

[2]  S. Essouri,et al.  Noninvasive positive pressure ventilation: Five years of experience in a pediatric intensive care unit* , 2006, Pediatric critical care medicine : a journal of the Society of Critical Care Medicine and the World Federation of Pediatric Intensive and Critical Care Societies.

[3]  A. Harf,et al.  Assessment of neonatal ventilator performances , 1995, Intensive Care Medicine.

[4]  Andrea Scorza,et al.  Preliminary evaluation of a fiber-optic sensor for flow measurements in pulmonary ventilators , 2011, 2011 IEEE International Symposium on Medical Measurements and Applications.

[5]  R. Branson Understanding and implementing advances in ventilator capabilities , 2004, Current opinion in critical care.

[6]  Byoungho Lee,et al.  Review of the present status of optical fiber sensors , 2003 .

[7]  R. Chatburn,et al.  Alternative modes of mechanical ventilation: A review for the hospitalist , 2009, Cleveland Clinic Journal of Medicine.

[8]  J G Webster,et al.  60-HZ interference in electrocardiography. , 1973, IEEE transactions on bio-medical engineering.

[9]  P. Cappa,et al.  Experimental evaluation of errors in the measurement of respiratory parameters of the newborn performed by a continuous flow neonatal ventilator , 2006, Journal of medical engineering & technology.

[10]  E. Klein What’s new in prostate cancer screening and prevention? , 2009, Cleveland Clinic Journal of Medicine.

[11]  Sergio Silvestri,et al.  A novel methodology for the experimental evaluation of pulmonary ventilator performance drift , 1997 .

[12]  L Battista,et al.  An air flow sensor for neonatal mechanical ventilation applications based on a novel fiber-optic sensing technique. , 2013, The Review of scientific instruments.

[13]  Sergio Silvestri,et al.  On the robust utilization of non-parametric tests for evaluation of combined cyclical and monotonic drift , 2001 .

[14]  Andrea Scorza,et al.  Fiber-Optic Flow Sensor for the Measurement of Inspiratory Efforts in Mechanical Neonatal Ventilation , 2014 .

[15]  T. Chai,et al.  Root mean square error (RMSE) or mean absolute error (MAE)? – Arguments against avoiding RMSE in the literature , 2014 .

[16]  M Schmid,et al.  Novel formulation of a double threshold algorithm for the estimation of muscle activation intervals designed for variable SNR environments. , 2012, Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology.

[17]  T. Larkin,et al.  Performance evaluation and testing of adult critical care ventilators , 1997 .

[18]  C. Carbery Basic Concepts in Mechanical Ventilation , 2008, Journal of perioperative practice.

[19]  F. P. Wieringa,et al.  Optical fiber sensors for medical applications: practical engineering considerations , 2009 .

[20]  Sergio Silvestri,et al.  Linear Model and Algorithm to Automatically Estimate the Pressure Limit of Pressure Controlled Ventilation for Delivering a Target Tidal Volume , 2006, Journal of Clinical Monitoring and Computing.

[21]  Hao Su,et al.  Real-time MRI-guided needle placement robot with integrated fiber optic force sensing , 2011, 2011 IEEE International Conference on Robotics and Automation.

[22]  Andrea Scorza,et al.  Preliminary Evaluation of a Simple Optical Fiber Measurement System for Monitoring Respiratory Pressure in Mechanically Ventilated Infants , 2012, BioMed 2012.

[23]  Sergio Silvestri,et al.  A high sensitivity fiber optic macro-bend based gas flow rate transducer for low flow rates: theory, working principle, and static calibration. , 2013, The Review of scientific instruments.

[24]  P. Boisvert,et al.  Preliminary survey of the electromagnetic interference environment in hospitals , 1991, IEEE 1991 International Symposium on Electromagnetic Compatibility.

[25]  S. Donn,et al.  Invasive and noninvasive neonatal mechanical ventilation. , 2003, Respiratory care.

[26]  Sergio Silvestri,et al.  A novel preterm respiratory mechanics active simulator to test the performances of neonatal pulmonary ventilators , 2002 .

[27]  M. Dojat,et al.  Mechanical ventilation: changing concepts , 2005 .