Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch.

We have evaluated the fluoroscopic on-line portal imaging (OPI) system developed by Siemens (Beamview-1, Concord, CA, U.S.A.) in routine clinical radiotherapy, involving the treatment of 883 fields (559 patient set-ups for treatment) on 21 patients. The image was typically generated by delivering 10 monitor units when used in single exposure or 1-2 monitor units on a large open field followed by 8-10 monitor units on the actual field when double exposure was used. Comparison between the portal image and the simulator film was done by eye. A region of tolerance was drawn on the simulator film and the field edges on the portal image had to project within this region. If this criterion was not met, adjustments followed by verification portal images were done before the remaining field dose was delivered. If possible, these adjustments were performed by moving the patient couch by remote control. The image quality was insufficient for evaluation in 75/883 (8.5%) fields. The abovementioned criterion was not met in 95/808 (11.8%) of the evaluable fields (26/559 patient set-ups were not evaluable). Of the 533 evaluable patient set-ups, 92 had to be adjusted (17.2%) including three (pelvic irradiations) set-ups that were adjusted on both field irradiated during the same radiotherapy session. In one case an incorrect tray (with wrong blocks) was detected and replaced. In one case (a 5.5 x 6.0 cm rectangular larynx field) the x and y axis of the field were interswitched. In one case incorrect focusing of a block was shown by the portal image. To make adjustments, the couch longitudinal position was changed 20 times (range -10 to +15 mm). The lateral position was changed 73 times (range -15 to +16 mm). The height position was changes 6 times (range -7 to +6 mm). Diaphragma rotation changes were performed 5 times (1 degree). The fraction of treatment time that was related to the use of OPI was 30.7% median (mean 32.4%, S.D. 14.1%). The range was 4.1 to 78.6%. On the basis of calculations assuming no OPI would have been used, field treatment time was increased by a median of 44.2% (mean 55.8%; S.D. 41.2%) by using OPI. The fraction of monitor units (fraction of the dose) to generate a satisfactory image was 10% median.(ABSTRACT TRUNCATED AT 400 WORDS)

[1]  L. Reinstein,et al.  Assessment of geometric treatment accuracy using time-lapse on-line portal imaging , 1990 .

[2]  Shlomo Shalev,et al.  Digital video system for on-line portal verification , 1990, Medical Imaging.

[3]  P. Munro,et al.  Clinical evaluation of a fluoroscopic imaging device for radiotherapy localizaton , 1990 .

[4]  M W Vannier,et al.  A method to analyze 2-dimensional daily radiotherapy portal images from an on-line fiber-optic imaging system. , 1991, International journal of radiation oncology, biology, physics.

[5]  K W Leszczynski,et al.  An adaptive technique for digital noise suppression in on-line portal imaging. , 1990, Physics in medicine and biology.

[6]  J. W. Epstein,et al.  On-line radiotherapy imaging with an array of fiber-optic image reducers. , 1990, International journal of radiation oncology, biology, physics.

[7]  S. Shalev,et al.  Digital contrast enhancement for online portal imaging , 1989, Medical and Biological Engineering and Computing.

[8]  M van Herk,et al.  First clinical experience with a newly developed electronic portal imaging device. , 1990, International journal of radiation oncology, biology, physics.

[9]  Konrad W. Leszczynski,et al.  A robust algorithm for contrast enhancement by local histogram modification , 1989, Image Vis. Comput..

[10]  S Shalev,et al.  Video techniques for on-line portal imaging. , 1989, Computerized medical imaging and graphics : the official journal of the Computerized Medical Imaging Society.

[11]  A. Dutreix When and how can we improve precision in radiotherapy? , 1984, Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology.