Improvement of Accuracy in a High-Capacity, Six Degree-of-freedom Load Cell: Application to Robotic Testing of Musculoskeletal Joints

AbstractThis study investigated a previously unaccounted for source of error in a high-capacity, six degree-of-freedom load cell used in multi-degree-of-freedom robotic testing of musculoskeletal joints, an application requiring a load cell with high accuracy in addition to high load capacity. A method of calibration is presented for reducing the error caused by changes in universal force-moment sensor (UFS) orientation within a gravitational field. Uncorrected, this error can exceed a magnitude of 1% of the full-scale load capacity—the manufacturer-stated accuracy of the UFS. Implementation of the calibration protocol reduced this error by approximately 75% for a variety of loading conditions. This improvement in load cell accuracy (while maintaining full load capacity) should improve both the measurement and control of specimen kinetics by robotic/UFS and other biomechanical testing systems. © 1999 Biomedical Engineering Society. PAC99: 8719Rr, 8780Vt, 0620Fn, 0620Dk, 8719Ff

[1]  H Fujie,et al.  Determination of the in situ forces and force distribution within the human anterior cruciate ligament. , 1995, Annals of biomedical engineering.

[2]  Identification of time-varying properties of the human triceps surae stretch reflex: II. rapid imposed movement , 1992, 1992 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[3]  T. Buchanan,et al.  Muscle activation at the human knee during isometric flexion‐extension and varus‐valgus loads , 1997, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[4]  S Arai,et al.  The use of robotics technology to study human joint kinematics: a new methodology. , 1993, Journal of biomechanical engineering.

[5]  G A Livesay,et al.  The use of a universal force-moment sensor to determine in-situ forces in ligaments: a new methodology. , 1995, Journal of biomechanical engineering.

[6]  Harry N. Norton,et al.  Handbook of transducers , 1969 .

[7]  M W Keith,et al.  Measurement of isometric elbow and shoulder moments: position-dependent strength of posterior deltoid-to-triceps muscle tendon transfer in tetraplegia. , 1996, IEEE transactions on rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society.

[8]  G A Livesay,et al.  A combined robotic/universal force sensor approach to determine in situ forces of knee ligaments. , 1996, Journal of biomechanics.

[9]  J. Cusick,et al.  Cyclic Compression‐Flexion Loading of the Human Lumbar Spine , 1994, Spine.

[10]  Richard S. C. Cobbold,et al.  Transducers for biomedical measurements: principles and applications , 1974 .

[11]  R. Cobbold Transducers for Biomedical Measurements , 1978 .

[12]  R. Kearney,et al.  Identification of time-varying dynamics of the human triceps surae stretch reflex , 1992, 1992 14th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.